Evaluation of a Novel Real-Time Continuous Glucose-Monitoring System for Use in Cats
View through CrossRef
Abstract
Background
The Guardian REAL-Time is a continuous glucose-monitoring system (CGMS) recently developed to provide instantaneous interstitial glucose concentrations; the system does not require a monitor being fixed to the animal.
Hypothesis
The CGMS provides accurate and reproducible real-time readings of glucose concentration in cats.
Animals
Thirty-two diabetic cats, 2 cats with suspected insulinoma, and 5 healthy cats.
Methods
Prospective, observational study. CGMS accuracy was compared with a reference glucose meter at normal, high, and low blood glucose concentrations using error grid analysis. Reading variability of 2 simultaneously used CGMS was determined in diabetic cats by calculating correlation and percentage of concordance of paired data at different glycemic ranges. The time interval between increasing glycemia and a rise in interstitial fluid glucose measured by the CGMS was assessed in healthy cats receiving glucose IV; the time point of maximal increase in interstitial glucose concentrations was calculated.
Results
The CGMS was 100, 96.1, and 91.0% accurate at normal, high, and low blood glucose concentrations. Measurements deviated from reference by −12.7 ± 70.5 mg/dL at normal, −12.1 ± 141.5 mg/dL at high, and −1.9 ± 40.9 mg/dL at low glucose concentrations. Overall, paired CGMS readings correlated significantly (r= 0.95, P < .0001) and concordance was 95.7%. The median delay after IV administration of glucose to an increase in interstitial glucose was 11.4 minutes (range: 8.8–19.7 minutes).
Conclusions and Clinical Importance
Although some readings substantially deviated from reference values, the CGMS yields reproducible results, is clinically accurate in cats with hyperglycemia and euglycemia, and is slightly less accurate if blood glucose concentrations are low. Rapidly increasing interstitial glucose after a glycemic rise suggests that the CGMS is suitable for real-time measurement under clinical conditions.
Oxford University Press (OUP)
Title: Evaluation of a Novel Real-Time Continuous Glucose-Monitoring System for Use in Cats
Description:
Abstract
Background
The Guardian REAL-Time is a continuous glucose-monitoring system (CGMS) recently developed to provide instantaneous interstitial glucose concentrations; the system does not require a monitor being fixed to the animal.
Hypothesis
The CGMS provides accurate and reproducible real-time readings of glucose concentration in cats.
Animals
Thirty-two diabetic cats, 2 cats with suspected insulinoma, and 5 healthy cats.
Methods
Prospective, observational study.
CGMS accuracy was compared with a reference glucose meter at normal, high, and low blood glucose concentrations using error grid analysis.
Reading variability of 2 simultaneously used CGMS was determined in diabetic cats by calculating correlation and percentage of concordance of paired data at different glycemic ranges.
The time interval between increasing glycemia and a rise in interstitial fluid glucose measured by the CGMS was assessed in healthy cats receiving glucose IV; the time point of maximal increase in interstitial glucose concentrations was calculated.
Results
The CGMS was 100, 96.
1, and 91.
0% accurate at normal, high, and low blood glucose concentrations.
Measurements deviated from reference by −12.
7 ± 70.
5 mg/dL at normal, −12.
1 ± 141.
5 mg/dL at high, and −1.
9 ± 40.
9 mg/dL at low glucose concentrations.
Overall, paired CGMS readings correlated significantly (r= 0.
95, P < .
0001) and concordance was 95.
7%.
The median delay after IV administration of glucose to an increase in interstitial glucose was 11.
4 minutes (range: 8.
8–19.
7 minutes).
Conclusions and Clinical Importance
Although some readings substantially deviated from reference values, the CGMS yields reproducible results, is clinically accurate in cats with hyperglycemia and euglycemia, and is slightly less accurate if blood glucose concentrations are low.
Rapidly increasing interstitial glucose after a glycemic rise suggests that the CGMS is suitable for real-time measurement under clinical conditions.
Related Results
Pregnancy and Challenging Transient Anti-GAD65 Positivity: A Case Report with Literature Review
Pregnancy and Challenging Transient Anti-GAD65 Positivity: A Case Report with Literature Review
Abstract
Introduction
During pregnancy, women may develop blood glucose abnormalities like gestational diabetes mellitus (GDM) or, rarely, type 1 diabetes (T1D), which can lead to ...
Prevalence of toxoplasmosis in semi-domesticated and pet cats within and around Bangkok, Thailand
Prevalence of toxoplasmosis in semi-domesticated and pet cats within and around Bangkok, Thailand
Abstract
Background
Toxoplasmosis is one of the most common parasitic zoonoses worldwide. Cats become infected after ingesting infected tissue cysts...
Quantification of defecation frequency in cats with and without chronic kidney disease
Quantification of defecation frequency in cats with and without chronic kidney disease
Objectives
The purpose of this study was to objectively measure defecation frequency in cats with and without chronic kidney disease (CKD). Cats with CKD are at higher ...
Because Neglect Isn't Cute: Tuxedo Stan's Campaign for a Humane World
Because Neglect Isn't Cute: Tuxedo Stan's Campaign for a Humane World
On 10 September 2012, a cat named Tuxedo Stan launched his campaign for mayor of the Halifax Regional Municipality in Nova Scotia, Canada (“Tuxedo Stan for Mayor”). Backed by his h...
Sleeping and resting respiratory rates in healthy adult cats and cats with subclinical heart disease
Sleeping and resting respiratory rates in healthy adult cats and cats with subclinical heart disease
Sleeping and resting respiratory rates are commonly measured variables in patients with cardiac disease. However, little information is available on these variables in healthy clie...
Feline Erythrocytic Osmotic Fragility in Normal and Anemic Cats—A Preliminary Study
Feline Erythrocytic Osmotic Fragility in Normal and Anemic Cats—A Preliminary Study
Erythrocyte osmotic fragility is an excellent parameter for evaluating the red blood cell (RBC) membrane, which may be abnormal in several pathological conditions. The flow cytomet...
Evaluation of plasma antithrombin activity and D‐dimer concentration in populations of healthy cats, clinically ill cats, and cats with cardiomyopathy
Evaluation of plasma antithrombin activity and D‐dimer concentration in populations of healthy cats, clinically ill cats, and cats with cardiomyopathy
Background: Current coagulation tests lack sensitivity and detect disseminated intravascular coagulation (DIC) only when it is severe. Measurement of antithrombin (AT) activity and...
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Abstract
The rapid growth of open access publishing (OAP) has significantly improved the accessibility and dissemination of scientific knowledge. However, this expansion has also c...