Screening Sleep Disordered Breathing with Noncontact Measurement in a Clinical Site
View through CrossRef
[abstFig src='/00290002/06.jpg' width='300' text='Respiratory rate from simulator and Kinect' ]<span class=”bold”>Background and purpose:</span>It has been considered that sleep-disordered breathing disorders, such as sleep apnea syndrome (SAS), cause an increase in the risk of cardiovascular disease or traffic accident risk, and thus early detection of SAS is important. It has been also important for medical workers at clinical sites to quantitatively evaluate the respiratory condition of hospitalized patients who are asleep in a simple method. A noncontact-type system was proposed to monitor the respiratory condition of sleeping patients and minimized patient-related stress such that medical workers could use the system for SAS screening and perform a preliminary check prior to definite diagnosis.<span class=”bold”>Method:</span>The system included Microsoft Kinect™ for windows® (Kinect), a tripod, and a PC. A depth sensor of Kinect was used to measure movement in the thorax motion. Data obtained from periodic waveforms were divided with the intervals of 1 min, and the number of peaks was used to obtain the respiratory rate. Additionally, a frequency analysis was performed to calculate the respiratory frequency from a frequency at which the maximum amplitude was observed. In Experiment 1), a METI-man® PatientSimulator (CAE healthcare) (simulator) was used to study the respiratory rate and frequency calculated from the Kinect data by gradually changing the designated respiratory rate. In Experiment 2), the respiratory condition of four sleeping subjects was monitored to calculate their respiratory rate and frequencies. Furthermore, a video camera was used to confirm periodic waveforms and spectrum features of body movements during sleep.<span class=”bold”>Results:</span>In Experiment 1), the results indicated that both the respiratory rate and frequency corresponded to the designated respiratory rate in each time zone. In Experiment 2), the results indicated that the respiratory rate of examines 1, 2, 3, and 4 corresponded to 12.79±2.44 times/min (average ± standard deviation), 16.46±4.33 times/min, 28.24±2.79 times/min, and 13.05±2.64 times/min, respectively. The findings also indicated that the frequency of examines 1, 2, 3, and 4 corresponded to 0.20±0.04 Hz, 0.26±0.06 Hz, 0.45±0.12 Hz, and 0.22±0.06 Hz, respectively. The periodic waveforms and amplitude spectra were enhanced with respect to body movements although regular waveform data were obtained after the body movement occurred.<span class=”bold”>Discussions:</span>The results indicated that body movement and posture temporarily affected monitoring of the system. However, the findings also revealed that it was possible to calculate the respiratory rate and frequency, and thus it was considered that the system was useful for monitoring the respiration confirm with the non-contact or SAS screening of patients in clinical site.
Fuji Technology Press Ltd.
Title: Screening Sleep Disordered Breathing with Noncontact Measurement in a Clinical Site
Description:
[abstFig src='/00290002/06.
jpg' width='300' text='Respiratory rate from simulator and Kinect' ]<span class=”bold”>Background and purpose:</span>It has been considered that sleep-disordered breathing disorders, such as sleep apnea syndrome (SAS), cause an increase in the risk of cardiovascular disease or traffic accident risk, and thus early detection of SAS is important.
It has been also important for medical workers at clinical sites to quantitatively evaluate the respiratory condition of hospitalized patients who are asleep in a simple method.
A noncontact-type system was proposed to monitor the respiratory condition of sleeping patients and minimized patient-related stress such that medical workers could use the system for SAS screening and perform a preliminary check prior to definite diagnosis.
<span class=”bold”>Method:</span>The system included Microsoft Kinect™ for windows® (Kinect), a tripod, and a PC.
A depth sensor of Kinect was used to measure movement in the thorax motion.
Data obtained from periodic waveforms were divided with the intervals of 1 min, and the number of peaks was used to obtain the respiratory rate.
Additionally, a frequency analysis was performed to calculate the respiratory frequency from a frequency at which the maximum amplitude was observed.
In Experiment 1), a METI-man® PatientSimulator (CAE healthcare) (simulator) was used to study the respiratory rate and frequency calculated from the Kinect data by gradually changing the designated respiratory rate.
In Experiment 2), the respiratory condition of four sleeping subjects was monitored to calculate their respiratory rate and frequencies.
Furthermore, a video camera was used to confirm periodic waveforms and spectrum features of body movements during sleep.
<span class=”bold”>Results:</span>In Experiment 1), the results indicated that both the respiratory rate and frequency corresponded to the designated respiratory rate in each time zone.
In Experiment 2), the results indicated that the respiratory rate of examines 1, 2, 3, and 4 corresponded to 12.
79±2.
44 times/min (average ± standard deviation), 16.
46±4.
33 times/min, 28.
24±2.
79 times/min, and 13.
05±2.
64 times/min, respectively.
The findings also indicated that the frequency of examines 1, 2, 3, and 4 corresponded to 0.
20±0.
04 Hz, 0.
26±0.
06 Hz, 0.
45±0.
12 Hz, and 0.
22±0.
06 Hz, respectively.
The periodic waveforms and amplitude spectra were enhanced with respect to body movements although regular waveform data were obtained after the body movement occurred.
<span class=”bold”>Discussions:</span>The results indicated that body movement and posture temporarily affected monitoring of the system.
However, the findings also revealed that it was possible to calculate the respiratory rate and frequency, and thus it was considered that the system was useful for monitoring the respiration confirm with the non-contact or SAS screening of patients in clinical site.
Related Results
Sleep apnea in patients undergoing coronary artery bypass grafting: Impact on perioperative outcomes
Sleep apnea in patients undergoing coronary artery bypass grafting: Impact on perioperative outcomes
SummarySleep‐disordered breathing is common in patients with coronary artery disease undergoing coronary artery bypass grafting. Sleep‐disordered breathing is associated with incre...
Acupuncture as therapeutic resource in patient with bruxism
Acupuncture as therapeutic resource in patient with bruxism
Bruxism is the harmful habit of clenching or grinding the teeth during the day and / or night, with unconscious pattern, with particular intensity and frequency, outside the functi...
Sleep Disordered Breathing in Children with Autism Spectrum Disorder: An In-depth Review of Correlations and Complexities
Sleep Disordered Breathing in Children with Autism Spectrum Disorder: An In-depth Review of Correlations and Complexities
Sleep-disordered breathing is a significant problem affecting the pediatric population. These conditions can affect sleep quality and significantly affect children's overa...
0279 Sleep Hygiene for Sleep Health in the General Population: What Does Data From Consumer Sleep Technology Tell Us?
0279 Sleep Hygiene for Sleep Health in the General Population: What Does Data From Consumer Sleep Technology Tell Us?
Abstract
Introduction
Despite being used and widely recommended since the 1970s, few studies have examined whether adherence to ...
Sleep Disorders in Patients with Breast Cancer Based on Polysomnographic Data.
Sleep Disorders in Patients with Breast Cancer Based on Polysomnographic Data.
Abstract
Background. Complaints of sleep disturbances are common in patients with breast cancer and may affect their quality of life. Insomnia, fatigue, and sleep fr...
0864 Severe Central Sleep Apnea
0864 Severe Central Sleep Apnea
Abstract
Introduction
Central sleep apnea (CSA) is a rare form of sleep disordered breathing with repeated apneic episodes with ...
0202 Predicting Sleep Inertia in a Biomathematical Model of Fatigue and Performance: A Novel Approach
0202 Predicting Sleep Inertia in a Biomathematical Model of Fatigue and Performance: A Novel Approach
Abstract
Introduction
Biomathematical models of fatigue typically include sleep inertia as an additive process during wakefulnes...
Is anemia in children with sleep disordered breathing actually a consequence of chronic disease?
Is anemia in children with sleep disordered breathing actually a consequence of chronic disease?
Objective. We aimed to investigate iron metabolism and laboratory findings of anemia in children undergoing adenotonsillectomy for sleep disordered breathing.Methods. Consecutive c...