Thoracic Endovascular Aortic Repair Using the C-TAG® Device with ACTIVE CONTROL System
View through CrossRef
Objective: To describe our experience with the Gore® C-TAG® endograft with ACTIVE CONTROL System (ACS) (W.L. Gore & Associates, Inc., Flagstaff, AZ, USA) in thoracic aortic repair, focusing on deployment accuracy and aortic wall apposition. Methods: All patients who underwent thoracic endovascular aortic repair (TEVAR) using the Gore® C-TAG® endograft with ACS from September 2017 to September 2021 were enrolled in a dedicated database and retrospectively analysed. We collected anatomic data on aortic arch angulation and tortuosity, proximal and distal landing zones, and the target for deployment accuracy. Proximal and distal deployment accuracies (PDA and DDA) were measured through intraoperative digital subtraction angiography (DSA), and postoperative computed tomography angiography (CTA) was required to define endograft apposition to the aortic wall. Results: Twenty-eight patients who underwent TEVAR with the Gore® C-TAG® with ACS at our institution were selected for this study: 46% presented with a type 3 aortic arch and a proximal landing zone < 3 was used in 53% of cases. Mean PDA and DDA were 1.89 ± 3.5 mm and 0.6 ± 1.4 mm, and were obtained in 93% and 100% of procedures, respectively. Mean proximal and distal wall apposition were 91 ± 17% and 98 ± 5.9%. Fifteen patients required an associated planned procedure, either to revascularize supra-aortic vessels when PLZ was < 3 or to assure optimal distal fixation with EndoAnchors™ (Medtronic, Minneapolis, MN, USA) delivery in selected cases. Two patients required reintervention during the same hospitalisation because of type 1a endoleak onset. No further reinterventions were needed during follow-up. Conclusions: Our single-centre analysis found promising results using the Gore® C-TAG® with ACS, with an optimal accuracy in deployment and wall apposition at both proximal and distal landing zones.
Surgical Technology Online
Title: Thoracic Endovascular Aortic Repair Using the C-TAG® Device with ACTIVE CONTROL System
Description:
Objective: To describe our experience with the Gore® C-TAG® endograft with ACTIVE CONTROL System (ACS) (W.
L.
Gore & Associates, Inc.
, Flagstaff, AZ, USA) in thoracic aortic repair, focusing on deployment accuracy and aortic wall apposition.
Methods: All patients who underwent thoracic endovascular aortic repair (TEVAR) using the Gore® C-TAG® endograft with ACS from September 2017 to September 2021 were enrolled in a dedicated database and retrospectively analysed.
We collected anatomic data on aortic arch angulation and tortuosity, proximal and distal landing zones, and the target for deployment accuracy.
Proximal and distal deployment accuracies (PDA and DDA) were measured through intraoperative digital subtraction angiography (DSA), and postoperative computed tomography angiography (CTA) was required to define endograft apposition to the aortic wall.
Results: Twenty-eight patients who underwent TEVAR with the Gore® C-TAG® with ACS at our institution were selected for this study: 46% presented with a type 3 aortic arch and a proximal landing zone < 3 was used in 53% of cases.
Mean PDA and DDA were 1.
89 ± 3.
5 mm and 0.
6 ± 1.
4 mm, and were obtained in 93% and 100% of procedures, respectively.
Mean proximal and distal wall apposition were 91 ± 17% and 98 ± 5.
9%.
Fifteen patients required an associated planned procedure, either to revascularize supra-aortic vessels when PLZ was < 3 or to assure optimal distal fixation with EndoAnchors™ (Medtronic, Minneapolis, MN, USA) delivery in selected cases.
Two patients required reintervention during the same hospitalisation because of type 1a endoleak onset.
No further reinterventions were needed during follow-up.
Conclusions: Our single-centre analysis found promising results using the Gore® C-TAG® with ACS, with an optimal accuracy in deployment and wall apposition at both proximal and distal landing zones.
Related Results
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract
Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
Safety and efficacy of physician-modified thoracic endovascular aortic repair using a low-profile device for aortic arch lesions
Safety and efficacy of physician-modified thoracic endovascular aortic repair using a low-profile device for aortic arch lesions
Background
We investigated the safety and efficacy of physician-modified thoracic endovascular aortic repair using a low-profile device for aortic arch lesions....
Presentation and Management of Cervical Thoracic Duct Cyst: A Systematic Review of the Literature
Presentation and Management of Cervical Thoracic Duct Cyst: A Systematic Review of the Literature
Abstract
Introduction
Thoracic duct cysts are an uncommon phenomenon, especially within the cervical region. Due to its limited reported cases, very little is known about its etiol...
Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Abstract
Background
Although an association between elevated blood pressure and risk of aortic aneurysm is established, f...
A Matched Case-Control Study on Open and Endovascular Treatment of Popliteal Artery Aneurysms
A Matched Case-Control Study on Open and Endovascular Treatment of Popliteal Artery Aneurysms
Objective:
To compare early and late results of open and endovascular management of popliteal artery aneurysm in a retrospective single-center matched case-cont...
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Abstract
Introduction
Thoracic outlet syndrome is a group of disorders arising from compressive forces on the neurovascular bundle in that region due to different etiologies. This...
Effect of stent graft fixation types on pararenal aortic diameter and renal function after endovascular aortic repair
Effect of stent graft fixation types on pararenal aortic diameter and renal function after endovascular aortic repair
Background:
This study aims to identify pararenal aortic diameter changes following endovascular aortic repair and to investigate the effects of suprarenal fixa...
RELAY™ Branched device: The story of an aortic arch device
RELAY™ Branched device: The story of an aortic arch device
Background: The management of aortic arch pathologies represents a great
challenge and is associated with high rates of mortality and morbidity.
A superior endovascular approach vi...