Javascript must be enabled to continue!
SU‐C‐211‐04: Evaluation of IMRT and VMAT Treatment Plan Quality Delivered with and Without Flattening Filter Using Pareto Optimal Fronts
View through CrossRef
Purpose: To evaluate treatment plan quality of IMRT and VMAT plans for high energy photon beams delivered with (FF) and without (FFF) flattening filter. Methods: 9‐field IMRT and 360° single arc VMAT plans, based on 10 MV beams with and without flattening filter, were created using Monaco (Elekta/CMS, v.2.04, USA) for three different patients (prescribed PTV dose 78 Gy). This treatment planning systems was commissioned for FF and FFF beams provided by an Elekta linac. The Pareto optimal fronts were created by calculating different treatment plans with varying rectum constraints and evaluating the PTV volume receiving less than 95% of the prescribed dose (V <95%) and the volume of the rectum receiving 70Gy or more (V 70Gy). Treatment plan efficiency was evaluated by recording number of monitor units (MUs) and by measuring the delivery time (T) using an Elekta Precise linear accelerator in FF and FFF mode. Results: The POFs of the rectum for both IMRT and VMAT in FFF mode were similar or even superior to the FF‐ modalities. For two of the three patients the POFs of IMRT and VMAT revealed a systematic difference of about 2% in target coverage. The POFs of the remaining patient showed virtually no difference for all four modalities. The delivery time of IMRT FFF decreased by about 23% compared to IMRT FF. In contrast to the IMRT techniques, delivery time increased by 20% when using VMAT FFF compared to VMAT FF. Conclusions: The evaluation of the POFs confirms that a flattening filter is not necessary for static and rotational IMRT treatments. Similar studies are needed for other IMRT indications, like head‐and‐neck cancer, to draw final conclusions. Sequencing algorithms for FFF‐beams need to be optimized and further improvement of the mechanical MLC properties are desirable for VMAT in FFF mode. Corporate sponsored research by Elekta
Title: SU‐C‐211‐04: Evaluation of IMRT and VMAT Treatment Plan Quality Delivered with and Without Flattening Filter Using Pareto Optimal Fronts
Description:
Purpose: To evaluate treatment plan quality of IMRT and VMAT plans for high energy photon beams delivered with (FF) and without (FFF) flattening filter.
Methods: 9‐field IMRT and 360° single arc VMAT plans, based on 10 MV beams with and without flattening filter, were created using Monaco (Elekta/CMS, v.
2.
04, USA) for three different patients (prescribed PTV dose 78 Gy).
This treatment planning systems was commissioned for FF and FFF beams provided by an Elekta linac.
The Pareto optimal fronts were created by calculating different treatment plans with varying rectum constraints and evaluating the PTV volume receiving less than 95% of the prescribed dose (V <95%) and the volume of the rectum receiving 70Gy or more (V 70Gy).
Treatment plan efficiency was evaluated by recording number of monitor units (MUs) and by measuring the delivery time (T) using an Elekta Precise linear accelerator in FF and FFF mode.
Results: The POFs of the rectum for both IMRT and VMAT in FFF mode were similar or even superior to the FF‐ modalities.
For two of the three patients the POFs of IMRT and VMAT revealed a systematic difference of about 2% in target coverage.
The POFs of the remaining patient showed virtually no difference for all four modalities.
The delivery time of IMRT FFF decreased by about 23% compared to IMRT FF.
In contrast to the IMRT techniques, delivery time increased by 20% when using VMAT FFF compared to VMAT FF.
Conclusions: The evaluation of the POFs confirms that a flattening filter is not necessary for static and rotational IMRT treatments.
Similar studies are needed for other IMRT indications, like head‐and‐neck cancer, to draw final conclusions.
Sequencing algorithms for FFF‐beams need to be optimized and further improvement of the mechanical MLC properties are desirable for VMAT in FFF mode.
Corporate sponsored research by Elekta.
Related Results
Modified VMAT Plans for Locally Advanced Centrally Located Non-Small Cell Lung Cancer (NSCLC)
Modified VMAT Plans for Locally Advanced Centrally Located Non-Small Cell Lung Cancer (NSCLC)
Objectives: This study aimed to find the optimal radiotherapy VMAT plans, that achieved high conformity and homogeneity to the planned target volume (PTV), and minimize the dose to...
Dosimetric comparative study of Volumetric Modulated Arc Therapy (VMAT), Intensity Modulated Radiotherapy (IMRT), and ThreeDimensional Conformal Radiation therapy (3DCRT) for post-mastectomy radiotherapy (PMRT) in Left breast cancer patients
Dosimetric comparative study of Volumetric Modulated Arc Therapy (VMAT), Intensity Modulated Radiotherapy (IMRT), and ThreeDimensional Conformal Radiation therapy (3DCRT) for post-mastectomy radiotherapy (PMRT) in Left breast cancer patients
Purpose: To evaluate the difference in planning target volume (PTV) coverage and dose to the organs at risk
(OAR) by using three different planning methods for the same patient-Vol...
COMPARISON OF COPLANAR AND NON- COPLANARVMAT FOR BRAIN CANCER BY USING THE DOSIMETRICAL AND RADIOBIOLOGICAL INDICES
COMPARISON OF COPLANAR AND NON- COPLANARVMAT FOR BRAIN CANCER BY USING THE DOSIMETRICAL AND RADIOBIOLOGICAL INDICES
Background: Recent techniques of radiotherapy such as volumetric modulated arc therapy (VMAT) that delivered in Coplanar technique or non-coplanar technique allows to deliver high ...
Dose and normal tissue complication probability analysis of various radiotherapy regimens for thymomas
Dose and normal tissue complication probability analysis of various radiotherapy regimens for thymomas
Objective
This study aimed to compare dosimetric differences and normal tissue complication probability (NTCP) among coplanar intensity-modulated radiation ther...
So sánh kỹ thuật xạ trị 3D-CRT, IMRT và VMAT trong điều trị ung thư phổi sử dụng xạ trị lập thể định vị thân
So sánh kỹ thuật xạ trị 3D-CRT, IMRT và VMAT trong điều trị ung thư phổi sử dụng xạ trị lập thể định vị thân
Purpose: To compare the difference of dose distribution and physical characteristics on treatment plans using Three-Dimensional Conformal Radiation Therapy (3D-CRT), Intensity Modu...
Feasibility of 4D VMAT-CT
Feasibility of 4D VMAT-CT
Abstract
Objective. Feasibility of three-dimensional (3D) tracking of volumetric modulated arc therapy (VMAT) based on VMAT–computed tomography (V...
Feasibility of hypofractionated pelvic radiotherapy with parametrial and nodal simultaneous integrated boost for locally advanced cervical cancers: dummy run for the HYACINCT trial
Feasibility of hypofractionated pelvic radiotherapy with parametrial and nodal simultaneous integrated boost for locally advanced cervical cancers: dummy run for the HYACINCT trial
Objectives
The HYACINCT trial will investigate the role of dose-adapted hypofractionated pelvic radiotherapy in patients with locally advanced cervical cancer who are ineligible fo...
Radioterapia Lattice: Um Estudo Comparativo de Viabilidade da técnica IMRT com a VMAT
Radioterapia Lattice: Um Estudo Comparativo de Viabilidade da técnica IMRT com a VMAT
O estudo comparativo incluiu dez casos de neoplasias planejados com IMRT, contrastados com planejamentos com VMAT, em estudo prévio na instituição. A técnica Lattice de Radiote...

