Comparison of LA and PVC mapping using OCTARAY and OPTRELL catheters
View through CrossRef
Abstract
Background
Multielectrode mapping catheters, such as the OCTARAY and OPTRELL, are essential in creating myocardial electroanatomical mapping in arrhythmias. The OCTARAY is a multi-spline mapping catheter with 48 closely spaced multielectrodes that enables high-resolution electroanatomical mapping, while the OPTRELL is a multi-electrode catheter with 36 electrodes arranged on eight radiating splines. However, only a few studies have compared their performance. In this study, we aimed to compare the OCTARAY and OPTRELL catheters in two areas: left atrial (LA) mapping during atrial fibrillation (AF) ablation; and premature ventricular contraction (PVC) mapping.
Methods
Twenty patients (Ten patients for LA mapping and ten for PVC mapping) were enrolled. LA voltage mapping was performed twice, alternating between catheters post-AF ablation. Parameters compared included mapping time, mapping points, catheter-induced premature atrial contraction (PACs), tissue proximity indication, low voltage area, and fluoroscopy time. For PVC mapping, comparisons included mapping time, catheter-induced PVCs, earliest activation time measured from the onset of PVC QRS, earliest activation point, and fluoroscopy time.
Results
Compared with THE OCTARAY, mean voltage using the OPTRELL was higher (0.192 mV[0.072, 0.48] vs. 0.126 mV[0.042, 0.378];
P
= .001) and the percentage of tissue proximity indication positive was also higher (14.97% vs. 11.45%;
P
< .001). However, there were no significant differences in low voltage area between the two groups (39.5 m
2
[16.5, 66.8] vs. 40[23.6, 61];
P
= .861), and in other LA parameters. In PVC mapping, catheter-induced PVCs using OPTRELL were significantly fewer than the OCTARAY (100 [32, 337] vs. 247 [110, 745],
P
= .039), with fewer catheter induced PVCs per minute (15 [6, 23] vs. 35 [20, 71],
P
= .039). However, no significant differences were observed in other PVCs mapping parameters.
Conclusion
The OPTRELL catheter demonstrated higher voltage recordings in LA mapping and fewer catheter-induced PVCs compared with the OCTARAY catheter. However, no significant difference was observed in other mapping parameters.
Title: Comparison of LA and PVC mapping using OCTARAY and OPTRELL catheters
Description:
Abstract
Background
Multielectrode mapping catheters, such as the OCTARAY and OPTRELL, are essential in creating myocardial electroanatomical mapping in arrhythmias.
The OCTARAY is a multi-spline mapping catheter with 48 closely spaced multielectrodes that enables high-resolution electroanatomical mapping, while the OPTRELL is a multi-electrode catheter with 36 electrodes arranged on eight radiating splines.
However, only a few studies have compared their performance.
In this study, we aimed to compare the OCTARAY and OPTRELL catheters in two areas: left atrial (LA) mapping during atrial fibrillation (AF) ablation; and premature ventricular contraction (PVC) mapping.
Methods
Twenty patients (Ten patients for LA mapping and ten for PVC mapping) were enrolled.
LA voltage mapping was performed twice, alternating between catheters post-AF ablation.
Parameters compared included mapping time, mapping points, catheter-induced premature atrial contraction (PACs), tissue proximity indication, low voltage area, and fluoroscopy time.
For PVC mapping, comparisons included mapping time, catheter-induced PVCs, earliest activation time measured from the onset of PVC QRS, earliest activation point, and fluoroscopy time.
Results
Compared with THE OCTARAY, mean voltage using the OPTRELL was higher (0.
192 mV[0.
072, 0.
48] vs.
0.
126 mV[0.
042, 0.
378];
P
= .
001) and the percentage of tissue proximity indication positive was also higher (14.
97% vs.
11.
45%;
P
< .
001).
However, there were no significant differences in low voltage area between the two groups (39.
5 m
2
[16.
5, 66.
8] vs.
40[23.
6, 61];
P
= .
861), and in other LA parameters.
In PVC mapping, catheter-induced PVCs using OPTRELL were significantly fewer than the OCTARAY (100 [32, 337] vs.
247 [110, 745],
P
= .
039), with fewer catheter induced PVCs per minute (15 [6, 23] vs.
35 [20, 71],
P
= .
039).
However, no significant differences were observed in other PVCs mapping parameters.
Conclusion
The OPTRELL catheter demonstrated higher voltage recordings in LA mapping and fewer catheter-induced PVCs compared with the OCTARAY catheter.
However, no significant difference was observed in other mapping parameters.
Related Results
Characteristics of the action potentials and the TRPC3-like channels in the cardiomyocytes from rabbit pulmonary vein sleeves
Characteristics of the action potentials and the TRPC3-like channels in the cardiomyocytes from rabbit pulmonary vein sleeves
Objective
Atrial fibrillation is the most prevalent clinical arrhythmia, and most of the paroxysmal atrial fibrillation was found to originate from the main pulmo...
Large Bore Catheters with Surface Treatments versus Untreated Catheters for Blood Access
Large Bore Catheters with Surface Treatments versus Untreated Catheters for Blood Access
Infection, thrombosis, and stenosis are among the most frequent complications associated with blood contacting catheters. Because these problems are usually related to surface prop...
Surface Treated Catheters with Ion Beam Based Process for Blood Access
Surface Treated Catheters with Ion Beam Based Process for Blood Access
Abstract: Infection, thrombosis, and stenosis are among the most frequent complications associated with blood contacting catheters. Because these problems are usually related to su...
C64. New Onset PVC Bigeminy Post Covid-19 Infection: Another Covid-Cardiac Related Problem
C64. New Onset PVC Bigeminy Post Covid-19 Infection: Another Covid-Cardiac Related Problem
Abstract
Background
Covid-19 is an infectious disease caused by a newly discovered coronavirus that caused acute severe respirat...
C64. New Onset PVC Bigeminy Post Covid-19 Infection: Another Covid-Cardiac Related Problem
C64. New Onset PVC Bigeminy Post Covid-19 Infection: Another Covid-Cardiac Related Problem
Abstract
Background
Covid-19 is an infectious disease caused by a newly discovered coronavirus that caused acute severe respirat...
Properties of Composite Films from Sugarcane Bagasse, Coconut Coir, and Water Hyacinth Fibers/Poly(Vinyl Chloride)
Properties of Composite Films from Sugarcane Bagasse, Coconut Coir, and Water Hyacinth Fibers/Poly(Vinyl Chloride)
Composite films from sugarcane bagasse fibers (SBF), coconut coir fibers (CCF) and water hyacinth fibers (WHF)/poly (vinyl chloride) (PVC) were successfully fabricated by solvent-c...
Utilization of Metal Oxides Nanoparticles in Modulating Polyvinyl Chloride Films to Resist Ultraviolet Light
Utilization of Metal Oxides Nanoparticles in Modulating Polyvinyl Chloride Films to Resist Ultraviolet Light
Modified poly(vinyl chloride) (PVC) films with organic groups (amino group from ethylene di-amine (en) and a suitable aromatic aldehyde from benzaldehyde (BEN)) were synthesized by...
Large Bore Catheters with Surface Treatments versus Untreated Catheters for Vascular Access in Hemodialysis
Large Bore Catheters with Surface Treatments versus Untreated Catheters for Vascular Access in Hemodialysis
Abstract: Infection, thrombosis, and stenosis are among the most frequent complications associated with blood‐contacting catheters. Complications resulting from infection remain a...