Comparison of LA and PVC mapping using OCTARAY and OPTRELL catheters
View through CrossRef
AbstractBackgroundMultielectrode mapping catheters, such as the OCTARAY and OPTRELL, are essential in creating myocardial electroanatomical mapping in arrhythmias. The OCTARAY is a multi-spline mapping catheter with 48 closely spaced multielectrodes that enables high-resolution electroanatomical mapping, while the OPTRELL is a multi-electrode catheter with 36 electrodes arranged on eight radiating splines. However, only a few studies have compared their performance. In this study, we aimed to compare the OCTARAY and OPTRELL catheters in two areas: left atrial (LA) mapping during atrial fibrillation (AF) ablation; and premature ventricular contraction (PVC) mapping.MethodsTwenty patients (Ten patients for LA mapping and ten for PVC mapping) were enrolled. LA voltage mapping was performed twice, alternating between catheters post-AF ablation. Parameters compared included mapping time, mapping points, catheter-induced premature atrial contraction (PACs), tissue proximity indication, low voltage area, and fluoroscopy time. For PVC mapping, comparisons included mapping time, catheter-induced PVCs, earliest activation time measured from the onset of PVC QRS, earliest activation point, and fluoroscopy time.ResultsCompared with THE OCTARAY, mean voltage using the OPTRELL was higher (0.192 mV[0.072, 0.48] vs. 0.126 mV[0.042, 0.378];P= .001) and the percentage of tissue proximity indication positive was also higher (14.97% vs. 11.45%;P< .001). However, there were no significant differences in low voltage area between the two groups (39.5 m2[16.5, 66.8] vs. 40[23.6, 61];P= .861), and in other LA parameters. In PVC mapping, catheter-induced PVCs using OPTRELL were significantly fewer than the OCTARAY (100 [32, 337] vs. 247 [110, 745],P= .039), with fewer catheter induced PVCs per minute (15 [6, 23] vs. 35 [20, 71],P= .039). However, no significant differences were observed in other PVCs mapping parameters.ConclusionThe OPTRELL catheter demonstrated higher voltage recordings in LA mapping and fewer catheter-induced PVCs compared with the OCTARAY catheter. However, no significant difference was observed in other mapping parameters.
Title: Comparison of LA and PVC mapping using OCTARAY and OPTRELL catheters
Description:
AbstractBackgroundMultielectrode mapping catheters, such as the OCTARAY and OPTRELL, are essential in creating myocardial electroanatomical mapping in arrhythmias.
The OCTARAY is a multi-spline mapping catheter with 48 closely spaced multielectrodes that enables high-resolution electroanatomical mapping, while the OPTRELL is a multi-electrode catheter with 36 electrodes arranged on eight radiating splines.
However, only a few studies have compared their performance.
In this study, we aimed to compare the OCTARAY and OPTRELL catheters in two areas: left atrial (LA) mapping during atrial fibrillation (AF) ablation; and premature ventricular contraction (PVC) mapping.
MethodsTwenty patients (Ten patients for LA mapping and ten for PVC mapping) were enrolled.
LA voltage mapping was performed twice, alternating between catheters post-AF ablation.
Parameters compared included mapping time, mapping points, catheter-induced premature atrial contraction (PACs), tissue proximity indication, low voltage area, and fluoroscopy time.
For PVC mapping, comparisons included mapping time, catheter-induced PVCs, earliest activation time measured from the onset of PVC QRS, earliest activation point, and fluoroscopy time.
ResultsCompared with THE OCTARAY, mean voltage using the OPTRELL was higher (0.
192 mV[0.
072, 0.
48] vs.
0.
126 mV[0.
042, 0.
378];P= .
001) and the percentage of tissue proximity indication positive was also higher (14.
97% vs.
11.
45%;P< .
001).
However, there were no significant differences in low voltage area between the two groups (39.
5 m2[16.
5, 66.
8] vs.
40[23.
6, 61];P= .
861), and in other LA parameters.
In PVC mapping, catheter-induced PVCs using OPTRELL were significantly fewer than the OCTARAY (100 [32, 337] vs.
247 [110, 745],P= .
039), with fewer catheter induced PVCs per minute (15 [6, 23] vs.
35 [20, 71],P= .
039).
However, no significant differences were observed in other PVCs mapping parameters.
ConclusionThe OPTRELL catheter demonstrated higher voltage recordings in LA mapping and fewer catheter-induced PVCs compared with the OCTARAY catheter.
However, no significant difference was observed in other mapping parameters.
Related Results
Anticorrosive and Morphological Properties of Polyaniline/Polyvinyl Chloride Blend on AA7075-T6
Anticorrosive and Morphological Properties of Polyaniline/Polyvinyl Chloride Blend on AA7075-T6
Introduction
Aluminium alloys are lightweight materials with good mechanical properties widely used in automotive and aerospace industries. Polyanili...
Characteristics of the action potentials and the TRPC3-like channels in the cardiomyocytes from rabbit pulmonary vein sleeves
Characteristics of the action potentials and the TRPC3-like channels in the cardiomyocytes from rabbit pulmonary vein sleeves
Objective
Atrial fibrillation is the most prevalent clinical arrhythmia, and most of the paroxysmal atrial fibrillation was found to originate from the main pulmo...
C64. New Onset PVC Bigeminy Post Covid-19 Infection: Another Covid-Cardiac Related Problem
C64. New Onset PVC Bigeminy Post Covid-19 Infection: Another Covid-Cardiac Related Problem
Abstract
Background
Covid-19 is an infectious disease caused by a newly discovered coronavirus that caused acute severe respirat...
C64. New Onset PVC Bigeminy Post Covid-19 Infection: Another Covid-Cardiac Related Problem
C64. New Onset PVC Bigeminy Post Covid-19 Infection: Another Covid-Cardiac Related Problem
Abstract
Background
Covid-19 is an infectious disease caused by a newly discovered coronavirus that caused acute severe respirat...
Acute Dialysis Catheters
Acute Dialysis Catheters
Acute dialysis catheters are non‐cuffed, non‐tunnelled catheters used for immediate vascular access. They are primarily used for acute renal failure in bed‐bound patients and for s...
Prospective Study of Incidence and Predictors of Peripheral IV Catheter–Induced Complications
Prospective Study of Incidence and Predictors of Peripheral IV Catheter–Induced Complications
BackgroundPeripheral intravenous catheters (PIVs) are invasive catheters that may endure risks of clinical complications affecting health care outcomes and patient satisfaction. Pa...
810. Resistance to the Minocycline-Rifampin-Chlorhexidine (MRCH) combination does not emerge in biofilms of Tetracycline and Rifampin resistant bacteria grown on MRCH catheters depleted below antimicrobially effective concentrations
810. Resistance to the Minocycline-Rifampin-Chlorhexidine (MRCH) combination does not emerge in biofilms of Tetracycline and Rifampin resistant bacteria grown on MRCH catheters depleted below antimicrobially effective concentrations
Abstract
Background
Central Line Associated Bloodstream Infections (CLABSIs) remain a significant medical problem for critically...
Structural, electrical and optical properties of PVC/ZnTe nanocomposite thin films
Structural, electrical and optical properties of PVC/ZnTe nanocomposite thin films
Abstract
ZnTe nanoparticles (NPs) were synthesized using hydrothermal method and then were dispersed in poly(vinyl-chloride) (PVC) to prepare hybrid PVC/ZnTe nanocomposite ...