Narrow beam breast CT: proof‐of‐concept

View through CrossRef
AbstractBackgroundIn breast CT, scattered photons form a large portion of the acquired signal, adversely impacting image quality throughout the frequency response of the imaging system. Prior studies provided evidence for a new image acquisition design, dubbed Narrow Beam Breast CT (NB‐bCT), in preventing scatter acquisition.PurposeHere, we report the design, implementation, and initial characterization of the first NB‐bCT prototype.MethodsThe imaging system's apparatus is composed of two primary assemblies: a dynamic Fluence Modulator (collimator) and a photon‐counting line detector. The design of the assemblies enables them to operate in lockstep during image acquisition, converting sourced x‐rays into a moving narrow beam. During a projection, this narrow beam sweeps the entire fan angle coverage of the imaging system. The assemblies are each comprised of a metal housing, a sensory system, and a robotic system. A controller unit handles their relative movements. To study the impact of fluence modulation on the signal received in the detector, three physical breast phantoms, representative of small, average, and large size breasts, were developed and imaged, and acquired projections analyzed. The scatter acquisition in each projection as a function of breast phantom size was investigated. The imaging system's spatial resolution at the center and periphery of the field of view was measured.ResultsMinimal acquisition of scattered rays occurs during image acquisition with NB‐bCT; results in minimal scatter to primary ratios in small, average, and large breast phantoms imaged were 0.05, 0.07, and 0.9, respectively. System spatial resolution of 5.2 lp/mm at 10% max MTF and 2.9 lp/mm at 50% max MTF at the center of the field of view was achieved, with minimal loss with the shift toward the corner (5.0 lp/mm at 10% max MTF and 2.5 lp/mm at 50% max MTF).ConclusionThe disclosed development, implementation, and characterization of a physical NB‐bCT prototype system demonstrates a new method of CT‐based image acquisition that yields high spatial resolution while minimizing scatter‐components in acquired projections. This methodology holds promise for high‐resolution CT‐imaging applications in which reduction of scatter contamination is desirable.
Title: Narrow beam breast CT: proof‐of‐concept
Description:
AbstractBackgroundIn breast CT, scattered photons form a large portion of the acquired signal, adversely impacting image quality throughout the frequency response of the imaging system.
Prior studies provided evidence for a new image acquisition design, dubbed Narrow Beam Breast CT (NB‐bCT), in preventing scatter acquisition.
PurposeHere, we report the design, implementation, and initial characterization of the first NB‐bCT prototype.
MethodsThe imaging system's apparatus is composed of two primary assemblies: a dynamic Fluence Modulator (collimator) and a photon‐counting line detector.
The design of the assemblies enables them to operate in lockstep during image acquisition, converting sourced x‐rays into a moving narrow beam.
During a projection, this narrow beam sweeps the entire fan angle coverage of the imaging system.
The assemblies are each comprised of a metal housing, a sensory system, and a robotic system.
A controller unit handles their relative movements.
To study the impact of fluence modulation on the signal received in the detector, three physical breast phantoms, representative of small, average, and large size breasts, were developed and imaged, and acquired projections analyzed.
The scatter acquisition in each projection as a function of breast phantom size was investigated.
The imaging system's spatial resolution at the center and periphery of the field of view was measured.
ResultsMinimal acquisition of scattered rays occurs during image acquisition with NB‐bCT; results in minimal scatter to primary ratios in small, average, and large breast phantoms imaged were 0.
05, 0.
07, and 0.
9, respectively.
System spatial resolution of 5.
2 lp/mm at 10% max MTF and 2.
9 lp/mm at 50% max MTF at the center of the field of view was achieved, with minimal loss with the shift toward the corner (5.
0 lp/mm at 10% max MTF and 2.
5 lp/mm at 50% max MTF).
ConclusionThe disclosed development, implementation, and characterization of a physical NB‐bCT prototype system demonstrates a new method of CT‐based image acquisition that yields high spatial resolution while minimizing scatter‐components in acquired projections.
This methodology holds promise for high‐resolution CT‐imaging applications in which reduction of scatter contamination is desirable.

Related Results

Desmoid-Type Fibromatosis of The Breast: A Case Series
Desmoid-Type Fibromatosis of The Breast: A Case Series
Abstract IntroductionDesmoid-type fibromatosis (DTF), also called aggressive fibromatosis, is a rare, benign, locally aggressive condition. Mammary DTF originates from fibroblasts ...
Breast Carcinoma within Fibroadenoma: A Systematic Review
Breast Carcinoma within Fibroadenoma: A Systematic Review
Abstract Introduction Fibroadenoma is the most common benign breast lesion; however, it carries a potential risk of malignant transformation. This systematic review provides an ove...
Spanish Breast Cancer Research Group (GEICAM)
Spanish Breast Cancer Research Group (GEICAM)
This section provides current contact details and a summary of recent or ongoing clinical trials being coordinated by Spanish Breast Cancer Research Group (GEICAM). Clinical trials...
Abstract OI-1: OI-1 Decoding breast cancer predisposition genes
Abstract OI-1: OI-1 Decoding breast cancer predisposition genes
Abstract Women with one or more first-degree female relatives with a history of breast cancer have a two-fold increased risk of developing breast cancer. This risk i...
International Breast Cancer Study Group (IBCSG)
International Breast Cancer Study Group (IBCSG)
This section provides current contact details and a summary of recent or ongoing clinical trials being coordinated by International Breast Cancer Study Group (IBCSG). Clinical tria...
Abstract IA11: Life course epidemiology and breast cancer: translating risk into prevention
Abstract IA11: Life course epidemiology and breast cancer: translating risk into prevention
Abstract Molecular histology may be conceptualized as the microscopic and molecular characteristics of normal tissues that are required for physiologic function. Ove...

Back to Top