Evaluation of Performance and Longevity of Ti-Cu Dry Electrodes: Degradation Analysis Using Anodic Stripping Voltammetry
View through CrossRef
This study aimed to investigate the degradation of dry biopotential electrodes using the anodic stripping voltammetry (ASV) technique. The electrodes were based on Ti-Cu thin films deposited on different polymeric substrates (polyurethane, polylactic acid, and cellulose) by Direct Current (DC) magnetron sputtering. TiCu0.34 thin films (chemical composition of 25.4 at.% Cu and 74.6 at.% Ti) were prepared by sputtering a composite Ti target. For comparison purposes, a Cu-pure thin film was prepared under the same conditions and used as a reference. Both films exhibited dense microstructures with differences in surface topography and crystalline structure. The degradation process involved immersing TiCu0.34 and Cu-pure thin films in artificial sweat (prepared following the ISO standard 3160-2) for different durations (1 h, 4 h, 24 h, 168 h, and 240 h). ASV was the technique selected to quantify the amount of Cu(II) released by the electrodes immersed in the sweat solution. The optimal analysis conditions were set for 120 s and −1.0 V for time deposition and potential deposition, respectively, with a quantification limit of 0.050 ppm and a detection limit of 0.016 ppm. The results showed that TiCu0.34 electrodes on polyurethane substrates were significantly more reliable over time compared to Cu-pure electrodes. After 240 h of immersion, the TiCu0.34 electrodes released a maximum of 0.06 ppm Cu, while Cu-pure electrodes released 16 ppm. The results showed the significant impact of the substrate on the electrode’s longevity, with cellulose bases performing poorly. TiCu0.34 thin films on cellulose released 1.15 µg/cm2 of copper after 240 h, compared to 1.12 mg/cm2 from Cu-pure films deposited on the same substrate. Optical microscopy revealed that electrodes based on polylactic acid substrates were more prone to corrosion over time, whereas TiCu thin-film metallic glass-like structures on PU substrates showed extended lifespan. This study underscored the importance of assessing the degradation of dry biopotential electrodes for e-health applications, contributing to developing more durable and reliable sensing devices. While the study simulated real-world conditions using artificial sweat, it did not involve in vivo measurements.
Title: Evaluation of Performance and Longevity of Ti-Cu Dry Electrodes: Degradation Analysis Using Anodic Stripping Voltammetry
Description:
This study aimed to investigate the degradation of dry biopotential electrodes using the anodic stripping voltammetry (ASV) technique.
The electrodes were based on Ti-Cu thin films deposited on different polymeric substrates (polyurethane, polylactic acid, and cellulose) by Direct Current (DC) magnetron sputtering.
TiCu0.
34 thin films (chemical composition of 25.
4 at.
% Cu and 74.
6 at.
% Ti) were prepared by sputtering a composite Ti target.
For comparison purposes, a Cu-pure thin film was prepared under the same conditions and used as a reference.
Both films exhibited dense microstructures with differences in surface topography and crystalline structure.
The degradation process involved immersing TiCu0.
34 and Cu-pure thin films in artificial sweat (prepared following the ISO standard 3160-2) for different durations (1 h, 4 h, 24 h, 168 h, and 240 h).
ASV was the technique selected to quantify the amount of Cu(II) released by the electrodes immersed in the sweat solution.
The optimal analysis conditions were set for 120 s and −1.
0 V for time deposition and potential deposition, respectively, with a quantification limit of 0.
050 ppm and a detection limit of 0.
016 ppm.
The results showed that TiCu0.
34 electrodes on polyurethane substrates were significantly more reliable over time compared to Cu-pure electrodes.
After 240 h of immersion, the TiCu0.
34 electrodes released a maximum of 0.
06 ppm Cu, while Cu-pure electrodes released 16 ppm.
The results showed the significant impact of the substrate on the electrode’s longevity, with cellulose bases performing poorly.
TiCu0.
34 thin films on cellulose released 1.
15 µg/cm2 of copper after 240 h, compared to 1.
12 mg/cm2 from Cu-pure films deposited on the same substrate.
Optical microscopy revealed that electrodes based on polylactic acid substrates were more prone to corrosion over time, whereas TiCu thin-film metallic glass-like structures on PU substrates showed extended lifespan.
This study underscored the importance of assessing the degradation of dry biopotential electrodes for e-health applications, contributing to developing more durable and reliable sensing devices.
While the study simulated real-world conditions using artificial sweat, it did not involve in vivo measurements.
Related Results
Preparation of Silver-Based Mercury Film Electrodes for Anodic Stripping Voltammetry
Preparation of Silver-Based Mercury Film Electrodes for Anodic Stripping Voltammetry
The aim of this study was to develop an electrochemical method for reproducible preparation and evaluation of silver-based mercury film electrodes (SBMFE) for applications such as ...
Structured 3D Printed Dry ECG Electrodes Using Copper Based Filament
Structured 3D Printed Dry ECG Electrodes Using Copper Based Filament
Commercial wet Silver and Silver Chloride electrodes are used to monitor electrocardiogram (ECG) signals in numerous bioimpedance applications. These electrodes are frequently sing...
Effects of binder and solvent on cathode manufacturing for Li-ion batteries
Effects of binder and solvent on cathode manufacturing for Li-ion batteries
Lithium-ion batteries (LIBs) have played a significant role in consumer electronics and electric vehicles. The major focus of research in Li-ion batteries has been to maximize perf...
To evaluate the outcome of varicose vein surgery with and without venous stripping
To evaluate the outcome of varicose vein surgery with and without venous stripping
Aim: The aim of this study to evaluate the outcome of varicose vein surgery with and without venous stripping.Material and methods: This research covered 100 varicose vein operatio...
Amalgam Electrodes for Electroanalysis
Amalgam Electrodes for Electroanalysis
AbstractLiquid mercury is a unique material for the indicator electrode in voltammetry. One reason for this is the high overvoltage for hydrogen formation, thus extending the actua...
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Abstract
The rapid growth of open access publishing (OAP) has significantly improved the accessibility and dissemination of scientific knowledge. However, this expansion has also c...
Remote Stripping Analysis of Lead and Copper by a Mercury‐Coated Platinum Microelectrode
Remote Stripping Analysis of Lead and Copper by a Mercury‐Coated Platinum Microelectrode
AbstractThe performance of a remote stripping sensor based on mercury microelectrodes (MM‐RS) for the in situ detection of trace metals in aquatic systems, was investigated. The su...
Cauchy, Cauchy–Santos–Sartori–Faria, Logit, and Probit Functions for Estimating Seed Longevity in Soybean
Cauchy, Cauchy–Santos–Sartori–Faria, Logit, and Probit Functions for Estimating Seed Longevity in Soybean
Seed longevity is characterized as the time for which seed remains viable during storage. Seed longevity can be estimated by a Probit model that determines the period in which 50% ...