Inter-study reproducibility of cardiovascular magnetic resonance-derived hemodynamic force assessments

View through CrossRef
AbstractCardiovascular magnetic resonance (CMR)-derived hemodynamic force (HDF) analyses have been introduced recently enabling more in-depth cardiac function evaluation. Inter-study reproducibility is important for a widespread clinical use but has not been quantified for this novel CMR post-processing tool yet. Serial CMR imaging was performed in 11 healthy participants in a median interval of 63 days (range 49–87). HDF assessment included left ventricular (LV) longitudinal, systolic peak and impulse, systolic/diastolic transition, diastolic deceleration as well as atrial thrust acceleration forces. Inter-study reproducibility and study sample sizes required to demonstrate 10%, 15% or 20% relative changes of HDF measurements were calculated. In addition, intra- and inter-observer analyses were performed. Intra- and inter-observer reproducibility was excellent for all HDF parameters according to intraclass correlation coefficient (ICC) values (> 0.80 for all). Inter-study reproducibility of all HDF parameters was excellent (ICC ≥ 0.80 for all) with systolic parameters showing lower coeffients of variation (CoV) than diastolic measurements (CoV 15.2% for systolic impulse vs. CoV 30.9% for atrial thrust). Calculated sample sizes to detect relative changes ranged from n = 12 for the detection of a 20% relative change in systolic impulse to n = 200 for the detection of 10% relative change in atrial thrust. Overall inter-study reproducibility of CMR-derived HDF assessments was sufficient with systolic HDF measurements showing lower inter-study variation than diastolic HDF analyses.
Title: Inter-study reproducibility of cardiovascular magnetic resonance-derived hemodynamic force assessments
Description:
AbstractCardiovascular magnetic resonance (CMR)-derived hemodynamic force (HDF) analyses have been introduced recently enabling more in-depth cardiac function evaluation.
Inter-study reproducibility is important for a widespread clinical use but has not been quantified for this novel CMR post-processing tool yet.
Serial CMR imaging was performed in 11 healthy participants in a median interval of 63 days (range 49–87).
HDF assessment included left ventricular (LV) longitudinal, systolic peak and impulse, systolic/diastolic transition, diastolic deceleration as well as atrial thrust acceleration forces.
Inter-study reproducibility and study sample sizes required to demonstrate 10%, 15% or 20% relative changes of HDF measurements were calculated.
In addition, intra- and inter-observer analyses were performed.
Intra- and inter-observer reproducibility was excellent for all HDF parameters according to intraclass correlation coefficient (ICC) values (> 0.
80 for all).
Inter-study reproducibility of all HDF parameters was excellent (ICC ≥ 0.
80 for all) with systolic parameters showing lower coeffients of variation (CoV) than diastolic measurements (CoV 15.
2% for systolic impulse vs.
CoV 30.
9% for atrial thrust).
Calculated sample sizes to detect relative changes ranged from n = 12 for the detection of a 20% relative change in systolic impulse to n = 200 for the detection of 10% relative change in atrial thrust.
Overall inter-study reproducibility of CMR-derived HDF assessments was sufficient with systolic HDF measurements showing lower inter-study variation than diastolic HDF analyses.

Related Results

Magnetic cloak made of NdFeB permanent magnetic material
Magnetic cloak made of NdFeB permanent magnetic material
In the past few years, the concept of an electromagnetic invisibility cloak has received much attention. Based on the pioneering theoretical work, invisibility cloaks have been gre...
Is the nuclear attraction just magnetic moment interaction between nucleons?
Is the nuclear attraction just magnetic moment interaction between nucleons?
Abstract By reexamining the magnetic moment interaction between nucleons, it is argued in this article that the formula for calculating magnetic induction generated by magn...
Association between dog and cat ownership with cardiovascular disease: A systematic review and meta-analysis
Association between dog and cat ownership with cardiovascular disease: A systematic review and meta-analysis
Background: Numerous studies have described the correlation of pet ownership with cardiovascular diseases, with dog and cat ownership emerging as the predominant forms of pet compa...
Advanced Hemodynamic Monitoring in Polytraumatized Patients: A Comprehensive Review
Advanced Hemodynamic Monitoring in Polytraumatized Patients: A Comprehensive Review
INTRODUCTION Advanced hemodynamic monitoring plays a crucial role in optimizing the perioperative management of polytraumatized patients, allowing for real-time assessment of cardi...
Method of liquid consumption measuring in nuclear magnetic flowmeters-relaxometers
Method of liquid consumption measuring in nuclear magnetic flowmeters-relaxometers
The need to expand the functionality of systems for monitoring the parameters of the flow of liquid media is substantiated. The advantages of using meters based on the phenomenon o...
Testing inter-scanner reproducibility of CrCEST at 3T
Testing inter-scanner reproducibility of CrCEST at 3T
Motivation: CrCEST MRI can map muscle-specific oxidative phosphorylation (OXPHOS), yet its reproducibility across different scanners has not been tested. Goal(s): To evaluate the i...

Back to Top