Characterization of a novel VenusX orthogonal dual-layer multileaf collimator

View through CrossRef
Abstract Background We investigated and characterized the performance of a novel orthogonal dual-layer multileaf collimator (αMLC) mounted on the LinaTech VenusX medical linear accelerator. Methods We evaluated leaf positioning accuracy and reproducibility using electronic portal imaging device (EPID) images of the picket fence test. The average, interleaf, and leaf-end transmissions of the upper, lower, and both αMLC layers were measured using an ionization chamber or EPID. The square and rhombus fields were used to evaluate the leaf penumbra of αMLC. To investigate the advantages of an orthogonal dual-layer MLC in field shaping, a right triangular pattern field, and a circular pattern field were formed using both layers and a single layer of the αMLC. These two evaluation pattern fields imaged were acquired by EPID and compared. Results The deviations in the positioning accuracy of the upper and lower MLC were 0.76 mm and 0.62 mm, respectively. Reproducibility was 0.29 mm and 0.26 mm correspondingly. The measured average transmissions were 1.86%, 1.82% and 0.03% for upper-, lower- and dual-layer MLC, respectively. The maximum interleaf transmission of the lower-layer MLC was 2.22% and 0.21% for the dual-layer. The maximum leaf-end transmissions were 23.90% and 0.44% correspondingly. Penumbra of the square field is 6.2 mm in X direction and 8.0 mm in Y direction. Average penumbras of the rhombus fields with side lengths of 5 cm and 10 cm were 3.6 mm and 4.9 mm, respectively. For the right triangular and circular fields, the fields shaped by the dual-layer of the αMLC were much closer to the planned field than the single-layer MLC. The dose undulation amplitude of the 50% isodose lines and the leaf stepping angle change of both layers was smaller than the single-layer MLC. Conclusions The αMLC benefits from its orthogonal dual-layer design. Leaf transmission, dose undulations at the field edge, and MLC field dependence of the leaf stepping angle of the dual-layer αMLC were remarkably reduced. αMLC can potentially improve the quality of IMRT and VMAT plans.
Title: Characterization of a novel VenusX orthogonal dual-layer multileaf collimator
Description:
Abstract Background We investigated and characterized the performance of a novel orthogonal dual-layer multileaf collimator (αMLC) mounted on the LinaTech VenusX medical linear accelerator.
Methods We evaluated leaf positioning accuracy and reproducibility using electronic portal imaging device (EPID) images of the picket fence test.
The average, interleaf, and leaf-end transmissions of the upper, lower, and both αMLC layers were measured using an ionization chamber or EPID.
The square and rhombus fields were used to evaluate the leaf penumbra of αMLC.
To investigate the advantages of an orthogonal dual-layer MLC in field shaping, a right triangular pattern field, and a circular pattern field were formed using both layers and a single layer of the αMLC.
These two evaluation pattern fields imaged were acquired by EPID and compared.
Results The deviations in the positioning accuracy of the upper and lower MLC were 0.
76 mm and 0.
62 mm, respectively.
Reproducibility was 0.
29 mm and 0.
26 mm correspondingly.
The measured average transmissions were 1.
86%, 1.
82% and 0.
03% for upper-, lower- and dual-layer MLC, respectively.
The maximum interleaf transmission of the lower-layer MLC was 2.
22% and 0.
21% for the dual-layer.
The maximum leaf-end transmissions were 23.
90% and 0.
44% correspondingly.
Penumbra of the square field is 6.
2 mm in X direction and 8.
0 mm in Y direction.
Average penumbras of the rhombus fields with side lengths of 5 cm and 10 cm were 3.
6 mm and 4.
9 mm, respectively.
For the right triangular and circular fields, the fields shaped by the dual-layer of the αMLC were much closer to the planned field than the single-layer MLC.
The dose undulation amplitude of the 50% isodose lines and the leaf stepping angle change of both layers was smaller than the single-layer MLC.
Conclusions The αMLC benefits from its orthogonal dual-layer design.
Leaf transmission, dose undulations at the field edge, and MLC field dependence of the leaf stepping angle of the dual-layer αMLC were remarkably reduced.
αMLC can potentially improve the quality of IMRT and VMAT plans.

Related Results

Study on the energy capture efficiency of flapping airfoil power generator using semiactive dual‐layer airfoils
Study on the energy capture efficiency of flapping airfoil power generator using semiactive dual‐layer airfoils
AbstractTo address the problem that the energy capture efficiencies of existing flapping airfoil power generators are lower than those of traditional turbine generators, a dual‐lay...
Design and analysis of dual-curing systems
Design and analysis of dual-curing systems
Dual-curing processing is a method to prepare thermoset materials through two polymerization reactions carried out simultaneously or sequentially. In these processes, a firm under...
Synchronizability and eigenvalues of two-layer star networks
Synchronizability and eigenvalues of two-layer star networks
From the study of multilayer networks, scientists have found that the properties of the multilayer networks show great difference from those of the traditional complex networks. In...
Penetrative convection in Nocturnal ABL: Numerical Simulations
Penetrative convection in Nocturnal ABL: Numerical Simulations
<p>After the sunset, under calm and clear sky conditions, aerosol laden surface air-layer, cools rapidly due to radiative cooling<sup>[1, 2, & 3...
A holistic aerosol model for Uranus and Neptune, including Dark Spots
A holistic aerosol model for Uranus and Neptune, including Dark Spots
<p>Previous studies of the reflectance spectra of Uranus and Neptune concentrated on individual, narrow wavelength regions, inferring solutions for the vertical struc...
SU‐E‐T‐458: Impact of Parameters for Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer
SU‐E‐T‐458: Impact of Parameters for Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer
Purpose:Different parameters Result in different plans. As the developing of volumetric modulated arc therapy for rectal cancer, the purpose of this study was to work out appropria...

Back to Top