AbstractBackgroundData on the factors that trigger repetitive nonreentrant ventriculoatrial synchrony (RNRVAS) are limited. We hypothesize that loss of atrial capture may trigger RNRVAS. We aimed to use an atrial threshold test to observe the development of RNRVAS upon loss of atrial capture in patients with implantable cardiac electronic devices (CIED).MethodsPatients with DDD mode CIEDs [177 patients, 67.5 ± 14.8 (70) years; 70 women] were included. Atrial threshold test was done in DDD mode at a rate at least 10 beats above the basal heart rate, with an AV delay of 300 ms (range 250–350). A multivariable logistic regression model was used to assess the independent predictors of RNRVAS.ResultsRNRVAS was observed in 69 of the 177 patients (39.0%) during atrial threshold test. In patients with VA conduction, incidence of RNRVAS increased to 76.7%. In univariate analysis, younger age (p = .038) and the presence of VA conduction (p < .001) were associated with an increased risk of RNRVAS, whereas complete AV block or any AV node conduction defect (p < .001) and the ventricular pacing ratio (p = .001) were inversely related to the risk of RNRVAS occurrence after loss of atrial capture. In multivariate analysis complete AV block (p = .009) and ventricular pacing ratio (p = .029) appeared as independent factors inversely related to the risk of RNRVAS development.ConclusionIn this study, we demonstrated that loss of atrial capture results in RNRVAS in one‐third of patients with a CIED in DDD mode, and in three‐fourths of those with VA conduction under certain predisposing CIED settings.
Title: Potential triggering of repetitive nonreentrant ventriculoatrial synchrony by loss of atrial capture
Description:
AbstractBackgroundData on the factors that trigger repetitive nonreentrant ventriculoatrial synchrony (RNRVAS) are limited.
We hypothesize that loss of atrial capture may trigger RNRVAS.
We aimed to use an atrial threshold test to observe the development of RNRVAS upon loss of atrial capture in patients with implantable cardiac electronic devices (CIED).
MethodsPatients with DDD mode CIEDs [177 patients, 67.
5 ± 14.
8 (70) years; 70 women] were included.
Atrial threshold test was done in DDD mode at a rate at least 10 beats above the basal heart rate, with an AV delay of 300 ms (range 250–350).
A multivariable logistic regression model was used to assess the independent predictors of RNRVAS.
ResultsRNRVAS was observed in 69 of the 177 patients (39.
0%) during atrial threshold test.
In patients with VA conduction, incidence of RNRVAS increased to 76.
7%.
In univariate analysis, younger age (p = .
038) and the presence of VA conduction (p < .
001) were associated with an increased risk of RNRVAS, whereas complete AV block or any AV node conduction defect (p < .
001) and the ventricular pacing ratio (p = .
001) were inversely related to the risk of RNRVAS occurrence after loss of atrial capture.
In multivariate analysis complete AV block (p = .
009) and ventricular pacing ratio (p = .
029) appeared as independent factors inversely related to the risk of RNRVAS development.
ConclusionIn this study, we demonstrated that loss of atrial capture results in RNRVAS in one‐third of patients with a CIED in DDD mode, and in three‐fourths of those with VA conduction under certain predisposing CIED settings.

Related Results

Potential triggering of Repetitive Nonreentrant Ventriculoatrial Synchrony (RNRVAS) by loss of atrial capture.
Potential triggering of Repetitive Nonreentrant Ventriculoatrial Synchrony (RNRVAS) by loss of atrial capture.
Background Data on the factors that trigger repetitive non-reentrant ventriculoatrial synchrony (RNRVAS) are limited. We hypothesize that loss of atrial capture may trigger RNRVAS....
Pseudo repetitive non‐reentrant ventriculoatrial synchrony: Current challenges
Pseudo repetitive non‐reentrant ventriculoatrial synchrony: Current challenges
AbstractBackgroundWe recently described a novel pacemaker‐mediated arrhythmia in Abbott cardiac implantable electronic devices (CIED), called pseudo‐RNRVAS, that mimics repetitive ...
Left atrial function assessment by speckle tracking echocardiography in low-risk atrial fibrillation patients
Left atrial function assessment by speckle tracking echocardiography in low-risk atrial fibrillation patients
Objective: To compare the left atrial function utilising speckle tracking echocardiography in patients with low-risk and high-risk non-valvular atrial fibrillation.Methods: The des...
P920Understanding arrhythmia mechanisms in patients with atrial septal defects
P920Understanding arrhythmia mechanisms in patients with atrial septal defects
Abstract Background Atrial arrhythmias represent a major cause of morbidity and hospitalization in patients with atrial septal d...
RELATIONSHIP BETWEEN ATRIAL FIBRILLATION CARDIOVERSION AND F
RELATIONSHIP BETWEEN ATRIAL FIBRILLATION CARDIOVERSION AND F
Objectives To investigate the relationship between atrial fibrillation cardioversion and f wave in electrocardiogram, providing an ordinary and noninvasive method...
Inhibitory Motifs Quench Synchrony Induced by Excitatory Motifs in Biological Neuronal Networks
Inhibitory Motifs Quench Synchrony Induced by Excitatory Motifs in Biological Neuronal Networks
A bstract The connectivity in biological neuronal networks is known to deviate significantly from the random network (Erdős–Rényi...
Left atrial strain parameters are able to predict presence of atrial fibrillation
Left atrial strain parameters are able to predict presence of atrial fibrillation
Abstract Background Three-dimensional speckle-tracking echocardiography allows evaluation of left atrial function, left atrial f...

Back to Top