Javascript must be enabled to continue!
e0486 Transcatheter closure of no rim large atrial septal defect with amplatzer occluders: technical considerations, safety and feasibility
View through CrossRef
Backgrounds
Transcatheter closuring no rim or large atrial septal defect (ASD) with Amplatzer occluding device (AOD) was technically challenged. The present study was to address technical issues, and to test the safety and feasibility for transcatheter closuring large and no rim ASDs with AOD.
Methods
Patients, with large ASDs of 325 mm and with no rims at least in one defect border detected echocardiographically, were included in the study. 49 patients eligible underwent transcatheter closure of ASDs and divided into group A (n=26, large ASDs with intact rims) and group B (n=23, large ASDs with no rims). Three occluding methods i.e. the conventional releasing, the waist pre-releasing, and the dumbbell-shaped releasing were sequentially attempted for all patients if necessary. For the waist pre-releasing method, the waist of AOD was released immediately following the expansion of the distal umbrella and withdrawn to wedge the defect in order to enhance the self-centralisation of the occluder; for the dumbbell-shaped releasing method, the distal umbrella was released within the upper left pulmonary vein to constrain the expansion of the umbrella, and the delivering catheter was further withdrawn slowly until the proximal umbrella was expanded in the right atrium, and finally, the original shape of the distal umbrella could quickly recover by slightly shaking or pulling the catheter, meanwhile the AOD could well fixed the defect.
Results
The average diameter of ASDs in group A and B were 27.2±11.7 mm and 28.5±11.9 mm, respectively (p>0.05), and the average diameter of finally used AODs was 34.5±10.2mm and 38.7±11.9 mm, respectively (p<0.01). The technical successful rates respectively for group A and B were 61.5% and 26.1% (p<0.05) by using the conventional releasing method, increased to 70.1% and 39.1% (p<0.05) by trying the waist pre-releasing method, and further increased to 100% and 100% by attempting the dumbbell-shaped releasing method. Neither major complications no occluder dislodging occurred peri-procedurally in the two groups.
Conclusions
Trans-catheter closure of no rim large-to-huge ASD with the AOD may be safe and feasible; closuring no rim large-to-huge ASD needs bigger AODs and more use of the dumbbell-shaped releasing method.
Title: e0486 Transcatheter closure of no rim large atrial septal defect with amplatzer occluders: technical considerations, safety and feasibility
Description:
Backgrounds
Transcatheter closuring no rim or large atrial septal defect (ASD) with Amplatzer occluding device (AOD) was technically challenged.
The present study was to address technical issues, and to test the safety and feasibility for transcatheter closuring large and no rim ASDs with AOD.
Methods
Patients, with large ASDs of 325 mm and with no rims at least in one defect border detected echocardiographically, were included in the study.
49 patients eligible underwent transcatheter closure of ASDs and divided into group A (n=26, large ASDs with intact rims) and group B (n=23, large ASDs with no rims).
Three occluding methods i.
e.
the conventional releasing, the waist pre-releasing, and the dumbbell-shaped releasing were sequentially attempted for all patients if necessary.
For the waist pre-releasing method, the waist of AOD was released immediately following the expansion of the distal umbrella and withdrawn to wedge the defect in order to enhance the self-centralisation of the occluder; for the dumbbell-shaped releasing method, the distal umbrella was released within the upper left pulmonary vein to constrain the expansion of the umbrella, and the delivering catheter was further withdrawn slowly until the proximal umbrella was expanded in the right atrium, and finally, the original shape of the distal umbrella could quickly recover by slightly shaking or pulling the catheter, meanwhile the AOD could well fixed the defect.
Results
The average diameter of ASDs in group A and B were 27.
2±11.
7 mm and 28.
5±11.
9 mm, respectively (p>0.
05), and the average diameter of finally used AODs was 34.
5±10.
2mm and 38.
7±11.
9 mm, respectively (p<0.
01).
The technical successful rates respectively for group A and B were 61.
5% and 26.
1% (p<0.
05) by using the conventional releasing method, increased to 70.
1% and 39.
1% (p<0.
05) by trying the waist pre-releasing method, and further increased to 100% and 100% by attempting the dumbbell-shaped releasing method.
Neither major complications no occluder dislodging occurred peri-procedurally in the two groups.
Conclusions
Trans-catheter closure of no rim large-to-huge ASD with the AOD may be safe and feasible; closuring no rim large-to-huge ASD needs bigger AODs and more use of the dumbbell-shaped releasing method.
Related Results
Abstract 18349: Long-term Results Comparing Recurrent Neurological Event Rates of Three Different Devices for Percutaneous PFO-Closure
Abstract 18349: Long-term Results Comparing Recurrent Neurological Event Rates of Three Different Devices for Percutaneous PFO-Closure
Background:
Percutaneous patent foramen ovale (PFO)-closure for secondary stroke prevention is discussed controversially, and long-term data comparing different closure...
ASSA13-15-6 Transcatheter Closure of Patent Ductus Arteriosus with Severe Pulmonary Hypertension Using Domestic Occluder in Adults
ASSA13-15-6 Transcatheter Closure of Patent Ductus Arteriosus with Severe Pulmonary Hypertension Using Domestic Occluder in Adults
Background
To evaluate the clinical efficacy of transcatheter closure of patent ductus arteriosus (PDA) with severe pulmonary hypertention using domestic occluder...
P920Understanding arrhythmia mechanisms in patients with atrial septal defects
P920Understanding arrhythmia mechanisms in patients with atrial septal defects
Abstract
Background
Atrial arrhythmias represent a major cause of morbidity and hospitalization in patients with atrial septal d...
Long-term results of transcatheter closure of large secundum atrial septal defect: comparison of intraoperative device closure
Long-term results of transcatheter closure of large secundum atrial septal defect: comparison of intraoperative device closure
Background
Both Transcatheter device closure (TCDC) or intraoperative device closure (IODC) have emerged as new minimally invasive alternatives to the classic sur...
Effects of Transcatheter Atrial Septal Defect Closure in Elderly Patients with Long-Standing Persistent Atrial Fibrillation
Effects of Transcatheter Atrial Septal Defect Closure in Elderly Patients with Long-Standing Persistent Atrial Fibrillation
Abstract
Purpose
Although the safety and efficacy of transcatheter atrial septal defect (ASD) closure has been reported in elderly patients, postprocedural outcomes in eld...
INTERVENTIONAL THERAPY OF MULTIPLE ATRIAL SEPTAL DEFECTS
INTERVENTIONAL THERAPY OF MULTIPLE ATRIAL SEPTAL DEFECTS
Objectives
To evaluate the safety and efficiency of transactheter closure of multiple atrial septal defects.
Met...
The clinical evaluation of transcatheter closure of perimembranous ventricular septal defects with the Amplatzer duct occluder II
The clinical evaluation of transcatheter closure of perimembranous ventricular septal defects with the Amplatzer duct occluder II
Objectives
Investigate the feasibility of perimembranous ventricular septal defects closure with the Amplatzer duct occluder II.
...
e0487 One year outcome of transcatheter closure of very large atrial septal defect with amplatzer occluders
e0487 One year outcome of transcatheter closure of very large atrial septal defect with amplatzer occluders
Objective
This study was to investigate the outcome of transcatheter closuring large-to-huge ASD with the AOD within one year.
...