A combined laboratory and field evaluation of the Cholestech LDX and CardioChek PA point-of-care testing lipid and glucose analysers

View through CrossRef
Background Combined lipid and glucose point-of-care testing (POCT) devices could facilitate widespread population screening for cardiovascular disease (CVD) and diabetes as part of the NHS Vascular Risk Assessment and Management Program (NHS Health Checks). An evaluation of the Cholestech LDX and CardioChek PA POCT analysers was performed in collaboration with the Wolverhampton City Primary Care Trust (PCT). Methods Intra-/inter-batch imprecision, between-analyser variation and the effect of haematocrit and ascorbic acid assay interference were investigated. Accuracy of the POCT capillary whole blood total cholesterol (TC), high-density-lipoprotein cholesterol (HDL-C) and glucose measurements was estimated by comparison with those from the laboratory analysis of paired venous samples. POCT usability and clinical governance were also assessed. Results The LDX exhibited lower intra- and inter-batch imprecision and external quality assessment (EQA) scheme between-analyser variation for the measurement of TC, HDL-C and glucose when compared to the CardioChek. Ascorbic acid negatively interfered in all three assays on both POCT analysers and results reported by the CardioChek were influenced by the specimen haematocrit. The LDX displayed closer agreement with the laboratory methods for the measurement of TC and HDL-C but both the LDX and the CardioChek displayed positive bias for the measurement of glucose. Conclusions POCT has clear advantages for delivering NHS Health Checks over the laboratory-based approach although device performance does differ. Users should also be aware of the potential clinical governance and interference issues associated with these devices.
Title: A combined laboratory and field evaluation of the Cholestech LDX and CardioChek PA point-of-care testing lipid and glucose analysers
Description:
Background Combined lipid and glucose point-of-care testing (POCT) devices could facilitate widespread population screening for cardiovascular disease (CVD) and diabetes as part of the NHS Vascular Risk Assessment and Management Program (NHS Health Checks).
An evaluation of the Cholestech LDX and CardioChek PA POCT analysers was performed in collaboration with the Wolverhampton City Primary Care Trust (PCT).
Methods Intra-/inter-batch imprecision, between-analyser variation and the effect of haematocrit and ascorbic acid assay interference were investigated.
Accuracy of the POCT capillary whole blood total cholesterol (TC), high-density-lipoprotein cholesterol (HDL-C) and glucose measurements was estimated by comparison with those from the laboratory analysis of paired venous samples.
POCT usability and clinical governance were also assessed.
Results The LDX exhibited lower intra- and inter-batch imprecision and external quality assessment (EQA) scheme between-analyser variation for the measurement of TC, HDL-C and glucose when compared to the CardioChek.
Ascorbic acid negatively interfered in all three assays on both POCT analysers and results reported by the CardioChek were influenced by the specimen haematocrit.
The LDX displayed closer agreement with the laboratory methods for the measurement of TC and HDL-C but both the LDX and the CardioChek displayed positive bias for the measurement of glucose.
Conclusions POCT has clear advantages for delivering NHS Health Checks over the laboratory-based approach although device performance does differ.
Users should also be aware of the potential clinical governance and interference issues associated with these devices.

Related Results

CUT-OFF POINT FOR FASTING GLUCOSE IN DIAGNOSING PREDIABETES
CUT-OFF POINT FOR FASTING GLUCOSE IN DIAGNOSING PREDIABETES
Objective. This study aimed to evaluate the feasibility of using fasting glucose as a primary diagnostic criterion for prediabetes, and to determine the optimal cut-off point for d...
Bioinformatics Analysis of Gefitinib or Rapamycin on Inhibiting the Survival of Hela in the Low Glucose and High Lactic Acid Environment
Bioinformatics Analysis of Gefitinib or Rapamycin on Inhibiting the Survival of Hela in the Low Glucose and High Lactic Acid Environment
Objective: To explore on the antitumor effect of gefitinib and rapamycin and possible mechanism in normal glucose and high lactic acid microenvironment. Methods: Hela cells are cul...
Abstract 1133: Glucose deprivation-induced intracellular reactive oxygen species activates the PI3K-AKT axis
Abstract 1133: Glucose deprivation-induced intracellular reactive oxygen species activates the PI3K-AKT axis
Abstract Glucose is the most efficient energy source and various cancer cells depend on glycolysis for their energy production. On the other hand, tumor microenviron...
Glucose intolerance is associated with resting heart rate among individuals without diabetes
Glucose intolerance is associated with resting heart rate among individuals without diabetes
Elevated resting heart rate is associated with cardiovascular diseases and all-cause mortality. Unmanaged diabetes is associated with high blood pressure and high resting heart rat...
Glutamate dehydrogenase 1 mediated glutaminolysis sustains HCC cells proliferation and survival under glucose deprivation
Glutamate dehydrogenase 1 mediated glutaminolysis sustains HCC cells proliferation and survival under glucose deprivation
Abstract Background: It is generally believed that tumor cells could sustain its proliferation and survival under different nutrient status according to a so-called metabol...
New and simple Ohmic definition of insulin resistance in lean and obese subjects
New and simple Ohmic definition of insulin resistance in lean and obese subjects
objective:: Insulin enhances the influx of glucose into cells. However, the relationship between glucose and insulin is complex and insulin sensitivity varies widely with age, ethn...
Effects of herbal tea (Platostoma palustre) on blood glucose regulation in vivo
Effects of herbal tea (Platostoma palustre) on blood glucose regulation in vivo
Platostoma palustre jelly is a traditional food. Platostoma palustre has been used as folk medicine and is effective against heat-shock, hypertension and diabetes. Therefore, the a...
Synthesis of a Ni–TiO2 Nanocomposite as an Enzyme–Less, Amperometric Sensor for Glucose Sensing & Monitoring
Synthesis of a Ni–TiO2 Nanocomposite as an Enzyme–Less, Amperometric Sensor for Glucose Sensing & Monitoring
In 1994, the Center for Disease Control declared that diabetes had reached epidemic proportions. Since then, however, little has been done to suppress the yearly increasing statist...

Back to Top