SU‐E‐P‐39: Characterization of Dose Impact On Different Beam Fields Size Attenuation for Elekta IGRT Couch
View through CrossRef
Purpose:Quantification of the dosimetric impact of the Elekta IGRT treatment couch in different beam field sizes. Established the relationship of relative dose versus beam angle at different beam field sizes.Methods:Measurements of couch attenuation were performed at gantry angles from 180° to 120°, using a 0.125cc semiflex ionization chamber, isocentrically placed in the center of a homogeneous cylindric sliced RW3 phantom for 6 photon beams. Measurements were performed at six different field sizes (3×3, 5×5, 7×7,10×10, 12×12 and 15×15 cm2). The phantom were positioned at the center of the couche,100 MU were delivered at every gantry angle. The dose difference to the ion chamber was determined.Results:For oblique fields with 6 MV photons at the same gantry angle the attenuation coefficient value from the lagest to the smallest the order is field size 7 cm2,5 cm2,10 cm2,12 cm2,15 cm2 and 3 cm2. The biggest couch attenuation by up to 4.15% at the gantry angle of 140°for the field size of 7 cm2, while for the field size of 3 cm2 the couch attenuated value only 3.5%. The other field size couch attenuation values are between the couch attenuated value of field size of 7 cm2 and 3 cm2Conclusion:The recommended treatment couch attenuation measured beam field size is 10×10 or 12×12 cm2. When measured using the beam field size 3×3 cm2 the tested value will be lower, while measuerd using the beam field size 7×7 cm2 the tested value will be higer than the recommended beam field size. This should be noted when modeling the treatment couch in the treatment planning system.
Title: SU‐E‐P‐39: Characterization of Dose Impact On Different Beam Fields Size Attenuation for Elekta IGRT Couch
Description:
Purpose:Quantification of the dosimetric impact of the Elekta IGRT treatment couch in different beam field sizes.
Established the relationship of relative dose versus beam angle at different beam field sizes.
Methods:Measurements of couch attenuation were performed at gantry angles from 180° to 120°, using a 0.
125cc semiflex ionization chamber, isocentrically placed in the center of a homogeneous cylindric sliced RW3 phantom for 6 photon beams.
Measurements were performed at six different field sizes (3×3, 5×5, 7×7,10×10, 12×12 and 15×15 cm2).
The phantom were positioned at the center of the couche,100 MU were delivered at every gantry angle.
The dose difference to the ion chamber was determined.
Results:For oblique fields with 6 MV photons at the same gantry angle the attenuation coefficient value from the lagest to the smallest the order is field size 7 cm2,5 cm2,10 cm2,12 cm2,15 cm2 and 3 cm2.
The biggest couch attenuation by up to 4.
15% at the gantry angle of 140°for the field size of 7 cm2, while for the field size of 3 cm2 the couch attenuated value only 3.
5%.
The other field size couch attenuation values are between the couch attenuated value of field size of 7 cm2 and 3 cm2Conclusion:The recommended treatment couch attenuation measured beam field size is 10×10 or 12×12 cm2.
When measured using the beam field size 3×3 cm2 the tested value will be lower, while measuerd using the beam field size 7×7 cm2 the tested value will be higer than the recommended beam field size.
This should be noted when modeling the treatment couch in the treatment planning system.
Related Results
On Flores Island, do "ape-men" still exist? https://www.sapiens.org/biology/flores-island-ape-men/
On Flores Island, do "ape-men" still exist? https://www.sapiens.org/biology/flores-island-ape-men/
<span style="font-size:11pt"><span style="background:#f9f9f4"><span style="line-height:normal"><span style="font-family:Calibri,sans-serif"><b><spa...
Hubungan Perilaku Pola Makan dengan Kejadian Anak Obesitas
Hubungan Perilaku Pola Makan dengan Kejadian Anak Obesitas
<p><em><span style="font-size: 11.0pt; font-family: 'Times New Roman',serif; mso-fareast-font-family: 'Times New Roman'; mso-ansi-language: EN-US; mso-fareast-langua...
Automation of the Monte Carlo simulation of medical linear accelerators
Automation of the Monte Carlo simulation of medical linear accelerators
The main result of this thesis is a software system, called PRIMO, which simulates clinical linear accelerators and the subsequent dose distributions using the Monte Carlo method. ...
Even Star Decomposition of Complete Bipartite Graphs
Even Star Decomposition of Complete Bipartite Graphs
<p><span lang="EN-US"><span style="font-family: 宋体; font-size: medium;">A decomposition (</span><span><span style="font-family: 宋体; font-size: medi...
Increased life expectancy of heart failure patients in a rural center by a multidisciplinary program
Increased life expectancy of heart failure patients in a rural center by a multidisciplinary program
Abstract
Funding Acknowledgements
Type of funding sources: None.
INTRODUCTION Patients with heart failure (HF)...
TU‐B‐201‐03: Accounting for MV Imaging Dose and the Future of MV Imaging
TU‐B‐201‐03: Accounting for MV Imaging Dose and the Future of MV Imaging
Radiotherapy is in the era of image guided radiation therapy (IGRT) in which imaging procedures are routinely performed for patient positioning, and target localization. The additi...
Assessment of Peripheral Photon Beam Skin Dose using MTS-7 TL Crystal for 15 MV Elekta Medical Linac
Assessment of Peripheral Photon Beam Skin Dose using MTS-7 TL Crystal for 15 MV Elekta Medical Linac
High energy linear accelerators are being used for treatment of cancer patients. In radiotherapy, the peripheral doses also attack the normal and healthy cells. According to the su...
Establishing time series of flux and grain size of suspended sand in rivers using an acoustic method
Establishing time series of flux and grain size of suspended sand in rivers using an acoustic method
<p>Measuring the concentration and grain size of suspended sand in rivers continuously remains a scientific challenge due to its pronounced spatio-temporal variabilit...