Count rate performance of brain-dedicated PET scanners: a Monte Carlo simulation study
View through CrossRef
Abstract
In recent years, there has been a renewed interest in brain-dedicated PET imaging systems, particularly in the context of combined PET/MR imaging. We are currently designing a brain-dedicated PET insert suitable for an ultra-high field brain-dedicated MR scanner, the Siemens Magnetom 7T MR scanner. In this paper, an investigation on the count rate performance of several possible detectors through a series of Monte Carlo simulations is reported. Brain-dedicated PET scanners with a lutetium oxyorthosilicate scintillator and a detector area of 0.04 (1 crystal per detector) to 101.37 (2500 crystals per detector) cm
2
, detector thickness of 10 to 20 mm and a fixed crystal pitch of ~2 mm were simulated. The count rate performance of each scanner was evaluated as a function of detector deadtime type and constant, coincidence timing window and lower level discriminator. Also, the effects of activity outside the field-of-view (FOV) on the count rate performance of each scanner were studied. For each detector geometry and performance metric, the scanner singles rate, scanner sensitivity and noise equivalent count rate as a function of activity in the FOV were measured. It was seen that scanners with detectors comprised a few crystal elements showed reduced scanner sensitivity due to a high number of inter-detector scattering. The count rate performance of scanners with large detectors, on the other hand, was mainly determined by the deadtime properties of the detectors. A model for the count rate performance of the scanner with each studied detector is presented in this work.
Title: Count rate performance of brain-dedicated PET scanners: a Monte Carlo simulation study
Description:
Abstract
In recent years, there has been a renewed interest in brain-dedicated PET imaging systems, particularly in the context of combined PET/MR imaging.
We are currently designing a brain-dedicated PET insert suitable for an ultra-high field brain-dedicated MR scanner, the Siemens Magnetom 7T MR scanner.
In this paper, an investigation on the count rate performance of several possible detectors through a series of Monte Carlo simulations is reported.
Brain-dedicated PET scanners with a lutetium oxyorthosilicate scintillator and a detector area of 0.
04 (1 crystal per detector) to 101.
37 (2500 crystals per detector) cm
2
, detector thickness of 10 to 20 mm and a fixed crystal pitch of ~2 mm were simulated.
The count rate performance of each scanner was evaluated as a function of detector deadtime type and constant, coincidence timing window and lower level discriminator.
Also, the effects of activity outside the field-of-view (FOV) on the count rate performance of each scanner were studied.
For each detector geometry and performance metric, the scanner singles rate, scanner sensitivity and noise equivalent count rate as a function of activity in the FOV were measured.
It was seen that scanners with detectors comprised a few crystal elements showed reduced scanner sensitivity due to a high number of inter-detector scattering.
The count rate performance of scanners with large detectors, on the other hand, was mainly determined by the deadtime properties of the detectors.
A model for the count rate performance of the scanner with each studied detector is presented in this work.
Related Results
Brain Organoids, the Path Forward?
Brain Organoids, the Path Forward?
Photo by Maxim Berg on Unsplash
INTRODUCTION
The brain is one of the most foundational parts of being human, and we are still learning about what makes humans unique. Advancements ...
SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
A enfermagem é o pilar da assistência pois está na linha de frente do cuidado holístico, todavia esta é estigmatizada e desvalorizada, assim como não possui reconhecimento consider...
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
ObjectiveTo compare the diagnostic differences between total-body PET/CT (positron emission tomography/computed tomography) and PET/MR (positron emission tomography/magnetic resona...
[RETRACTED] Gro-X Brain Reviews - Is Gro-X Brain A Scam? v1
[RETRACTED] Gro-X Brain Reviews - Is Gro-X Brain A Scam? v1
[RETRACTED]➢Item Name - Gro-X Brain➢ Creation - Natural Organic Compound➢ Incidental Effects - NA➢ Accessibility - Online➢ Rating - ⭐⭐⭐⭐⭐➢ Click Here To Visit - Official Website - ...
Inter-scanner Aβ-amyloid PET harmonization using barrel phantom spatial resolution matching
Inter-scanner Aβ-amyloid PET harmonization using barrel phantom spatial resolution matching
Abstract
The standardized uptake value ratio (SUVR) is used to measure Aβ uptake in PET images of the brain. Variations in PET scanner technologies and image recons...
Monte Carlo methods: barrier option pricing with stable Greeks and multilevel Monte Carlo learning
Monte Carlo methods: barrier option pricing with stable Greeks and multilevel Monte Carlo learning
For discretely observed barrier options, there exists no closed solution under the Black-Scholes model. Thus, it is often helpful to use Monte Carlo simulations, which are easily a...
Abstract PO5-07-11: Use of 64Cu-DOTA trastuzumab -PET to predict response to trastuzumab-deruxtecan (TDXd) in patients with metastatic disease to the brain: Study in Progress
Abstract PO5-07-11: Use of 64Cu-DOTA trastuzumab -PET to predict response to trastuzumab-deruxtecan (TDXd) in patients with metastatic disease to the brain: Study in Progress
Abstract
Background: We have utilized 64Cu-DOTA trastuzumab-PET imaging in patients with advanced breast cancer. In our experience, uptake on 64Cu-DOTA trastuzumab-P...
Comparison of 18F-FDG-PET and 18F-FDG-PET/CT for the diagnostic performance in thyroid nodules with indeterminate cytology
Comparison of 18F-FDG-PET and 18F-FDG-PET/CT for the diagnostic performance in thyroid nodules with indeterminate cytology
Abstract
Backgrounds:
We performed a meta-analysis to compare 18F-FDG-PET and 18F-FDG-PET/CT for the diagnostic performance in thyroid nodules wi...