Javascript must be enabled to continue!
The difference of removal force values on various preload application on 5º taper implant transfer abutment to determine the optimal preload
View through CrossRef
Background: Dental implants successfully become long-term solution in Indonesia with the two-piece variety being prevalent. The 5º tapered internal connections are favored for their stability compared to external connections. Preload is essential for stability relying on friction between the implant-abutment connection. Purpose: Evaluate the differences in removal force value resulting from various preload applications on 5º taper implant transfer abutment to determine the optimal preload. Methods: Implant abutments with 5º taper implant transfer abutments were evaluated in the study. The control group of this study derived from the manufacturer’s preload recommendation of 35 Ncm. Preloads from 10 Ncm, 15 Ncm, 20 Ncm, 25 Ncm, 30 Ncm, 40 Ncm, 45 Ncm, 50 Ncm were used as independent variables in this experiment (n=3). The abutment screw was removed before the tensile test using a universal testing machine (UTM). Statistical data analysis was conducted using the One-way ANOVA and Tukey HSD (p<0,05). Results: The 35 Ncm preload has the highest result with a mean value of (723 N ± 53.8). The 10 Ncm preload had the lowest results with a mean value of (267 N ± 71.4). There was no remarkable difference between the 30 Ncm and the 35 Ncm preload groups. Conclusion: There is a difference in removal force values at various preloads given. The removal force value at 35 Ncm is the optimal preload recommended by the manufacturer in line with a preload of 30 Ncm which is not significantly different from 35 Ncm.
Title: The difference of removal force values on various preload application on 5º taper implant transfer abutment to determine the optimal preload
Description:
Background: Dental implants successfully become long-term solution in Indonesia with the two-piece variety being prevalent.
The 5º tapered internal connections are favored for their stability compared to external connections.
Preload is essential for stability relying on friction between the implant-abutment connection.
Purpose: Evaluate the differences in removal force value resulting from various preload applications on 5º taper implant transfer abutment to determine the optimal preload.
Methods: Implant abutments with 5º taper implant transfer abutments were evaluated in the study.
The control group of this study derived from the manufacturer’s preload recommendation of 35 Ncm.
Preloads from 10 Ncm, 15 Ncm, 20 Ncm, 25 Ncm, 30 Ncm, 40 Ncm, 45 Ncm, 50 Ncm were used as independent variables in this experiment (n=3).
The abutment screw was removed before the tensile test using a universal testing machine (UTM).
Statistical data analysis was conducted using the One-way ANOVA and Tukey HSD (p<0,05).
Results: The 35 Ncm preload has the highest result with a mean value of (723 N ± 53.
8).
The 10 Ncm preload had the lowest results with a mean value of (267 N ± 71.
4).
There was no remarkable difference between the 30 Ncm and the 35 Ncm preload groups.
Conclusion: There is a difference in removal force values at various preloads given.
The removal force value at 35 Ncm is the optimal preload recommended by the manufacturer in line with a preload of 30 Ncm which is not significantly different from 35 Ncm.
Related Results
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct
Introduction
Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
REUSE OF THE INTRAOSSEOUS PART OF THE DENTAL IMPLANT (THE CLINICAL CASE)
REUSE OF THE INTRAOSSEOUS PART OF THE DENTAL IMPLANT (THE CLINICAL CASE)
Introduction. Replacement of a small included defect in the frontal area with a fixed structure based on the implant is the optimal solution. It is not always required to remove th...
<span style="mso-ansi-language: EN;">Retrieval of Fractured Abutment Screw of Dental Implant. Case Report<span style="font-size: 10.0pt; mso-ansi-font-size: 18.0pt; font-family: 'Times New Roman',serif; mso-ascii-font-family: 'Palatino Linotype';
<span style="mso-ansi-language: EN;">Retrieval of Fractured Abutment Screw of Dental Implant. Case Report<span style="font-size: 10.0pt; mso-ansi-font-size: 18.0pt; font-family: 'Times New Roman',serif; mso-ascii-font-family: 'Palatino Linotype';
Background: The advancement of dental implants marks a significant milestone in modern dentistry, offering an effective solution for tooth replacement while improving both function...
Factors Influencing Marginal Bone Loss around Dental Implants: A Narrative Review
Factors Influencing Marginal Bone Loss around Dental Implants: A Narrative Review
Implant supported dental prostheses are increasingly used in dental practice. The aim of this narrative review is to present the influence of transmucosal surface of prosthetic abu...
Testing the mechanical performance of a single taper-retained implant–abutment connection under static and dynamic loads: an in vitro study
Testing the mechanical performance of a single taper-retained implant–abutment connection under static and dynamic loads: an in vitro study
The mechanical strength and stability of single taper-retained implant–abutment connections in dental implantations have not been determined. We evaluated the mechanical performanc...
Crown-to-implant ratio of short locking-taper implants placed in posterior areas of the maxilla: a retrospective study 2 years after loading
Crown-to-implant ratio of short locking-taper implants placed in posterior areas of the maxilla: a retrospective study 2 years after loading
Introduction: The purpose was to study any relationship between crown-to-implant ratio and peri-implant bone loss of short, plateau-design, locking taper implants in posterior maxi...
A study on magnetic field assisted laser
percussion drilling and its effect
on surface integrity
A study on magnetic field assisted laser
percussion drilling and its effect
on surface integrity
Purpose: of this paper is to reduce the taper angle and surface roughness of the laser drilled
hole on Aluminium alloy with the assistance of magnetic field. At lower laser powers...
Stress and Strain Analysis of the Bone-Implant Interface: A Comparison of Fiber-Reinforced Composite and Titanium Implants Utilizing 3-Dimensional Finite Element Study
Stress and Strain Analysis of the Bone-Implant Interface: A Comparison of Fiber-Reinforced Composite and Titanium Implants Utilizing 3-Dimensional Finite Element Study
This study analyzed stress and strain mediated by 2 different implant materials, titanium (Ti) and experimental fiber-reinforced composite (FRC), on the implant and on the bone tis...

