Novel Perceval Sizing Technique: Single Center Experience
View through CrossRef
Abstract
Objective
Aim of this study was to compare old (manufacturer recommended) and new (institution based) sizing techniques for sutureless valve.
Materials
A 226 consecutively operated patients underwent aortic valve replacement with Perceval sutureless valve (Corcym) and had CT scan with contrast enhancement were included to this study. The final decision of appropriate size is based on intra-operative obturator sizing. Briefly, we have measured on the CT scans the annular ring surface and perimeter in order to estimate the prosthesis size. New sizing technique uses only white obturator of Perceval and it should passe trough the annulus with slight friction, which practically under-sizes with respect to manufacturer’s recommendations.
Results
The operative mortality was 1 (0.4424779%). There were no prosthesis migration neither annular rupture in any group. The mean follow up was lower in new group (3.3 ± 2.0 vs 9.7 ± 1.6, p < 0.01). At the discharge the patients who have used the new sizing technique had less gradient on the prosthetic valve (13.4 ± 5.0 vs 15.2 ± 5.5, p = 0.02). The new sizing was less prone to degeneration at the follow-up which would require intervention (13.4 ± 5.0 vs 15.2 ± 5.5, p = 0.02). Oversizing of 22.6% had significant role on valve gradient increase and structural degeneration (
p
< 0.05).
Conclusions
New sizing technique is safe and reproducible. It seem to deliver better immediate and long term benefits for Perceval sutureless valve, less postoperative gradient, less probability or structural degeneration.
Title: Novel Perceval Sizing Technique: Single Center Experience
Description:
Abstract
Objective
Aim of this study was to compare old (manufacturer recommended) and new (institution based) sizing techniques for sutureless valve.
Materials
A 226 consecutively operated patients underwent aortic valve replacement with Perceval sutureless valve (Corcym) and had CT scan with contrast enhancement were included to this study.
The final decision of appropriate size is based on intra-operative obturator sizing.
Briefly, we have measured on the CT scans the annular ring surface and perimeter in order to estimate the prosthesis size.
New sizing technique uses only white obturator of Perceval and it should passe trough the annulus with slight friction, which practically under-sizes with respect to manufacturer’s recommendations.
Results
The operative mortality was 1 (0.
4424779%).
There were no prosthesis migration neither annular rupture in any group.
The mean follow up was lower in new group (3.
3 ± 2.
0 vs 9.
7 ± 1.
6, p < 0.
01).
At the discharge the patients who have used the new sizing technique had less gradient on the prosthetic valve (13.
4 ± 5.
0 vs 15.
2 ± 5.
5, p = 0.
02).
The new sizing was less prone to degeneration at the follow-up which would require intervention (13.
4 ± 5.
0 vs 15.
2 ± 5.
5, p = 0.
02).
Oversizing of 22.
6% had significant role on valve gradient increase and structural degeneration (
p
< 0.
05).
Conclusions
New sizing technique is safe and reproducible.
It seem to deliver better immediate and long term benefits for Perceval sutureless valve, less postoperative gradient, less probability or structural degeneration.
Related Results
Small Aortic Annulus: Aortic Allografts vs Perceval S
Small Aortic Annulus: Aortic Allografts vs Perceval S
Цель. Оценить и сравнить гемодинамические показатели и изменение качества жизни после имплантации протезов Perceval S и аортальных аллографтов, а также частоту и спектр осложнений ...
Specificity of European Theatre Directing of the Flemish Wave: Metaphorical Images Luk Perceval
Specificity of European Theatre Directing of the Flemish Wave: Metaphorical Images Luk Perceval
The purpose of the article is to reveal the peculiarities of the creative activity of one of the leading representatives of theatre directing of the "Flemish wave" Luk Perceval. Re...
Perceval S and Coronary Artery Bypass Grafting, Contradiction or Full Harmony?
Perceval S and Coronary Artery Bypass Grafting, Contradiction or Full Harmony?
Background / Study Objective: Coronary artery disease is very common in patients who are referred to aortic valve replacement. Concomitant coronary artery bypass grafting (CABG) pr...
Abstract 16654: Sizing Strategy is a Major Determinant for Postoperative Pressure Gradients in Commonly Implanted Stented Tissue Valves
Abstract 16654: Sizing Strategy is a Major Determinant for Postoperative Pressure Gradients in Commonly Implanted Stented Tissue Valves
Background
Heated debates revolve around hemodynamic performance of stented tissue valves. We and others reported differences in valve size nomenclature confusing direc...
Pre-sizing online tool for photovoltaic water pumping system: development, testing, validation, and evaluation
Pre-sizing online tool for photovoltaic water pumping system: development, testing, validation, and evaluation
AbstractRenewable energy has become increasingly popular worldwide, and solar energy is one of the most sustainable and widely used forms of renewable energy. However, the lack of ...
Is Sub-Commissural Annuloplasty a Safe Adjunct to Sutureless Perceval-S Aortic Valve Implantation?
Is Sub-Commissural Annuloplasty a Safe Adjunct to Sutureless Perceval-S Aortic Valve Implantation?
Background: Sutureless bioprosthesis aortic valves simplify surgery for aortic valve replacement (AVR) but some unexpected anatomical features of the recipients aortic annulus migh...
Factors Influencing on Retention of Alkylketene Dimer
Factors Influencing on Retention of Alkylketene Dimer
Mechanisms of retention of AKD are studied in terms of effects of cationic polymer additions, and roles of fines and carboxyl groups of pulp fibers. Handsheets were prepared from n...
PR-335-203810-R01 Review of Xray Computed Tomography Performance
PR-335-203810-R01 Review of Xray Computed Tomography Performance
X-Ray Computed Tomography (XRCT) has been utilized for decades in medical and industrial imaging applications. The technology uses penetrating X-ray radiation to image the internal...