Comparability of four clinical laboratory measurement methods for GGT and commutability of candidate reference materials
View through CrossRef
AbstractBackgroundThis study was conducted to evaluate the progress in the standardization of the gamma‐glutamyl transferase (GGT) to achieve metrological traceability of routine in vitro diagnosis (IVD) medical devices.MethodsWe collected 25 single fresh frozen serum samples for GGT analysis. Candidate reference materials (RMs), calibrators, internal quality controls (IQC), and external quality assessment (EQA) materials from the National Center for Clinical Laboratory (NCCL), Beijing Center for Clinical Laboratory (BCCL), and College of American Pathologists (CAP) were randomly added to these serum samples. A total of 42 samples were examined using IFCC reference method and four different IVD medical devices to perform the comparability and commutability study.ResultsThe four IVD medical devices achieved trueness assessment within the measurement range. Linear analysis showed the agreement of Siemens ADVIA 2400, Hitachi 7600‐020/BioSino, Beckman AU 5800, and Roche Cobas 501 with the reference method. These assay pairs were comparable at the medical decision levels. The GGT in‐house candidate RMs, and Beckmann and Roche calibrators were all within the limits of the 95% prediction intervals, the commutability of BioSino calibrators was indeterminate, and some internal and external quality controls were not commutable for comparisons of certain IVD medical devices vs the reference method.ConclusionsBy comparing with the reference method, we found that performance of GGT conventional measurement systems to be traceable to the higher order references was improved. The commutable materials for calibration and trueness controls of routine methods were significant to promote the standardization of GGT analysis.
Title: Comparability of four clinical laboratory measurement methods for GGT and commutability of candidate reference materials
Description:
AbstractBackgroundThis study was conducted to evaluate the progress in the standardization of the gamma‐glutamyl transferase (GGT) to achieve metrological traceability of routine in vitro diagnosis (IVD) medical devices.
MethodsWe collected 25 single fresh frozen serum samples for GGT analysis.
Candidate reference materials (RMs), calibrators, internal quality controls (IQC), and external quality assessment (EQA) materials from the National Center for Clinical Laboratory (NCCL), Beijing Center for Clinical Laboratory (BCCL), and College of American Pathologists (CAP) were randomly added to these serum samples.
A total of 42 samples were examined using IFCC reference method and four different IVD medical devices to perform the comparability and commutability study.
ResultsThe four IVD medical devices achieved trueness assessment within the measurement range.
Linear analysis showed the agreement of Siemens ADVIA 2400, Hitachi 7600‐020/BioSino, Beckman AU 5800, and Roche Cobas 501 with the reference method.
These assay pairs were comparable at the medical decision levels.
The GGT in‐house candidate RMs, and Beckmann and Roche calibrators were all within the limits of the 95% prediction intervals, the commutability of BioSino calibrators was indeterminate, and some internal and external quality controls were not commutable for comparisons of certain IVD medical devices vs the reference method.
ConclusionsBy comparing with the reference method, we found that performance of GGT conventional measurement systems to be traceable to the higher order references was improved.
The commutable materials for calibration and trueness controls of routine methods were significant to promote the standardization of GGT analysis.
Related Results
E19 Gamma-glutamyl transpeptidase level is correlated with disease activity in oligoarticular and polyarticular Juvenile idiopathic arthritis
E19 Gamma-glutamyl transpeptidase level is correlated with disease activity in oligoarticular and polyarticular Juvenile idiopathic arthritis
Abstract
Introduction
Gamma-glutamyl transpeptidase (GGT) level was found to be correlated with disease activity indexes in rheu...
e0365 Clinical study on relationship between serum gamma-glutamyltransferase and coronary heart disease in women
e0365 Clinical study on relationship between serum gamma-glutamyltransferase and coronary heart disease in women
Objective
To study the relationship between serum gamma-glutamyltransferase (GGT) and coronary heart disease (CHD) in women.
...
Serum Gamma Glutamyl Transferase: Understanding its Contribution as a Potential Predictor of the Occurrence of Type 2 Diabetes
Serum Gamma Glutamyl Transferase: Understanding its Contribution as a Potential Predictor of the Occurrence of Type 2 Diabetes
Introduction:
The liver and kidneys are the primary locations of the glutathione metabolism enzyme gamma-glutamyl transferase (GGT). The two main factors contributing to an
increas...
Gambaran Kadar Gamma Glutamyl Transferase Pada Pasien Tuberkulosis Paru Yang Menjalani Terapi Oat Di Puskesmas Wonorejo
Gambaran Kadar Gamma Glutamyl Transferase Pada Pasien Tuberkulosis Paru Yang Menjalani Terapi Oat Di Puskesmas Wonorejo
Terapi OAT dalam jangka panjang dapat menyebabkan Drug Induced Liver Injury karena hepatotoksisitas sehingga diperlukan pemeriksaan Gamma Glutamyl Transferase untuk melihat fungsi ...
Accounting Comparability and Debt Cost
Accounting Comparability and Debt Cost
<p align="justify">This paper takes China's A-share listed companies in 2010-2017 as the research object to explore the impact of accounting comparability on debt costs. Firs...
Transaminases and gamma glutamyl transpeptidase for detecting nonalcoholic steatohepatitis and fibrosis in nonalcoholic fatty liver disease
Transaminases and gamma glutamyl transpeptidase for detecting nonalcoholic steatohepatitis and fibrosis in nonalcoholic fatty liver disease
Background: Nonalcoholic steatohepatitis (NASH) and advanced fibrosis are the spectrum of nonalcoholic fatty liver disease (NAFLD) that may progress to cirrhosis.Objective: We aime...
THE EFFECT OF PERORAL POLYVINYL CHLORIDE MICROPLASTIC ON ALKALINE PHOSPHATASE AND GAMMA-GLUTAMYL TRANSFERASE LEVELS IN RATTUS NORVEGICUS WISTAR STRAIN
THE EFFECT OF PERORAL POLYVINYL CHLORIDE MICROPLASTIC ON ALKALINE PHOSPHATASE AND GAMMA-GLUTAMYL TRANSFERASE LEVELS IN RATTUS NORVEGICUS WISTAR STRAIN
Introduction: Human exposure to microplastics continues to rise. Microplastics are harmful and are suspected of contributing to various health problems in humans, including reduced...
Serum γ‐Glutamyltransferase: Independent Predictor of Risk of Diabetes, Hypertension, Metabolic Syndrome, and Coronary Disease
Serum γ‐Glutamyltransferase: Independent Predictor of Risk of Diabetes, Hypertension, Metabolic Syndrome, and Coronary Disease
Serum γ‐glutamyltransferase (GGT) is associated with oxidative stress and hepatic steatosis. The extent to which its value in determining incident cardiometabolic risk (coronary he...