Evaluation of First-Draw Whole Blood, Point-of-Care Cardiac Markers in the Context of the Universal Definition of Myocardial Infarction: A Comparison of a Multimarker Panel to Troponin Alone and to Testing in the Central Laboratory
View through CrossRef
Abstract
Context
.—Previous studies evaluating point-of-care testing (POCT) for cardiac biomarkers did not use current recommendations for troponin cutoff values or recognize the recent universal definition of acute myocardial infarction. Traditionally, achieving optimal sensitivity for the detection of myocardial injury on initial presentation required combining cardiac troponin and/or creatine kinase isoenzyme MB with an early marker, usually myoglobin. In recent years, the performance of central laboratory combining cardiac troponin assays has improved significantly, potentially obviating the need for a multimarker panel to achieve optimum sensitivity.
Objective
.—To compare 2 commonly used POCT strategies to a fourth generation, central laboratory cardiac troponin T assay on first-draw specimens from patients being evaluated for acute myocardial infarction in the emergency department. The 2 strategies included a traditional POCT multimarker panel and a newer POCT method using cardiac troponin I alone.
Design
.—Blood specimens from 204 patients presenting to the emergency department with signs and/or symptoms of myocardial ischemia were measured on the 2 POCT systems and by a central laboratory method. The diagnosis for each patient was determined by retrospective chart review.
Results
.—The cardiac troponin T assasy alone was more sensitive for acute myocardial infarction than the multimarker POCT panel with equal or better specificity. When compared with a POCT troponin I, the cardiac troponin T was also more sensitive, but this difference was not significant. The POCT troponin I alone also had the same sensitivity as the multimarker panel.
Conclusions
.—Testing for combining cardiac troponin alone using newer, commercially available, central laboratory or POCT assays performed with equal or greater sensitivity to acute myocardial infarction as the older, traditional, multimarker panel. In the near future, high-sensitivity, central laboratory troponins will be available for routine clinical use. As a result, the quality gap between central laboratories and older POCT methods will continue to widen, unless the performance of the POCT methods is improved.
Archives of Pathology and Laboratory Medicine
Title: Evaluation of First-Draw Whole Blood, Point-of-Care Cardiac Markers in the Context of the Universal Definition of Myocardial Infarction: A Comparison of a Multimarker Panel to Troponin Alone and to Testing in the Central Laboratory
Description:
Abstract
Context
.
—Previous studies evaluating point-of-care testing (POCT) for cardiac biomarkers did not use current recommendations for troponin cutoff values or recognize the recent universal definition of acute myocardial infarction.
Traditionally, achieving optimal sensitivity for the detection of myocardial injury on initial presentation required combining cardiac troponin and/or creatine kinase isoenzyme MB with an early marker, usually myoglobin.
In recent years, the performance of central laboratory combining cardiac troponin assays has improved significantly, potentially obviating the need for a multimarker panel to achieve optimum sensitivity.
Objective
.
—To compare 2 commonly used POCT strategies to a fourth generation, central laboratory cardiac troponin T assay on first-draw specimens from patients being evaluated for acute myocardial infarction in the emergency department.
The 2 strategies included a traditional POCT multimarker panel and a newer POCT method using cardiac troponin I alone.
Design
.
—Blood specimens from 204 patients presenting to the emergency department with signs and/or symptoms of myocardial ischemia were measured on the 2 POCT systems and by a central laboratory method.
The diagnosis for each patient was determined by retrospective chart review.
Results
.
—The cardiac troponin T assasy alone was more sensitive for acute myocardial infarction than the multimarker POCT panel with equal or better specificity.
When compared with a POCT troponin I, the cardiac troponin T was also more sensitive, but this difference was not significant.
The POCT troponin I alone also had the same sensitivity as the multimarker panel.
Conclusions
.
—Testing for combining cardiac troponin alone using newer, commercially available, central laboratory or POCT assays performed with equal or greater sensitivity to acute myocardial infarction as the older, traditional, multimarker panel.
In the near future, high-sensitivity, central laboratory troponins will be available for routine clinical use.
As a result, the quality gap between central laboratories and older POCT methods will continue to widen, unless the performance of the POCT methods is improved.
Related Results
THE CLINICAL VALUE OF MYOCARDIAL ENZYMES AND TROPONIN I COMBINED DETECTION FOR EARLY DIAGNOSIS OF ACUTE MYOCARDIAL INFARCTION
THE CLINICAL VALUE OF MYOCARDIAL ENZYMES AND TROPONIN I COMBINED DETECTION FOR EARLY DIAGNOSIS OF ACUTE MYOCARDIAL INFARCTION
Objectives
To investigate the clinical value of myocardial enzymes and troponin I combined detection for early diagnosis of acute myocardial infarction.
...
e0447 Combination of conventional biomarkers for risk stratification in non ST elevation acute coronary syndrome
e0447 Combination of conventional biomarkers for risk stratification in non ST elevation acute coronary syndrome
Objective
To investigate the predicting value of electrocardiography (ECG) deviation, plasma levels of troponin I (cTnI) and Combination of conventional biomarker...
Additional PKA phosphorylation sites in human cardiac troponin I
Additional PKA phosphorylation sites in human cardiac troponin I
We used mass spectrometry to monitor cAMP‐dependent protein kinase catalysed phosphorylation of human cardiac troponin I in vitro. Phosphorylation of isolated troponin I by cAMP‐de...
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Misleading Elevation of Troponin T caused by Polymyositis
Misleading Elevation of Troponin T caused by Polymyositis
Background: Elevations of cardiac enzymes are commonly used to indicate myocardial ischemia, but they can be elevated due to other conditions. Different forms of Troponin (cTnT, sT...
Possible effects and changes of myocardial injury markers in percutaneous coronary intervention
Possible effects and changes of myocardial injury markers in percutaneous coronary intervention
Objective
Percutaneous Coronary Intervention (PCI) is a technique of cardiac intervention treatment in revascularisation of coronary vessel. The use of PCI in pat...
Structural dynamics of the intrinsically disordered linker region of cardiac troponin T
Structural dynamics of the intrinsically disordered linker region of cardiac troponin T
ABSTRACT
The cardiac troponin complex, composed of troponins I, T, and C, plays a central role in regulating the calcium-dependent interactions between myosin and t...
Study of Ischemia Modified Albumin as New Potential Diagnostic Biomarker In Acute Myocardial Infarction.
Study of Ischemia Modified Albumin as New Potential Diagnostic Biomarker In Acute Myocardial Infarction.
Background: Because of the varied presentation and associated high mortality the identification of patients with acute myocardial infarction is very critical for the patient manage...