Dosimetric Characteristics of Radiophotoluminescent Glass Dosimeters for Proton Beams
View through CrossRef
Purpose:
The purpose of the study was to investigate the dosimetric characteristics of radiophotoluminescent glass dosimeters (RGDs) for pencil beam scanning proton therapy. The RGD’s end-to-end testing of intensity-modulated proton therapy (IMPT) plans was also evaluated.
Materials and Methods:
The dosimetric characteristics of the GD-302M type glass dosimeter were studied in terms of uniformity, short-term and long-term reproducibility, stability of the magazine position readout, dose linearity in the range from 0.2 to 20 Gy, energy response in 70–220 MeV, and fading effect. The reference conditions of the spot scanning beam from the Varian ProBeam Compact system were operation at 160 MeV, a 2 cm water-equivalent depth in a solid water phantom, a 10 cm × 10 cm field size at the isocenter, and 2 Gy dose delivery. End-to-end testing of IMPT plans for the head, abdomen, and pelvis was verified using the Alderson Rando phantom. The overall uncertainty analysis was confirmed in this study.
Results:
The relative response of RGDs for the uniformity test was within 0.95–1.05. The percentages of the coefficients of variation for short-term and long-term reproducibility were 1.16% and 1.50%, respectively. The dose ACE glass dosimetry reader FGD-1000 showed a stable magazine position readout. The dose was found to be linear with R
2 = 0.9988. The energy response relative to 160 MeV was approximately within 4.0%. The fading effect was within 2.4%. For the end-to-end test, the difference between the treatment plan and RGD measurement was within 1.0%. The overall uncertainty of the RGD measurement for the proton beam was 4.6%, which covered all energy ranges in this study.
Conclusion:
The experimental study indicates that the RGDs have the potential to be used in the dosimetry of therapeutic proton beams, including end-to-end dosimetry.
Title: Dosimetric Characteristics of Radiophotoluminescent Glass Dosimeters for Proton Beams
Description:
Purpose:
The purpose of the study was to investigate the dosimetric characteristics of radiophotoluminescent glass dosimeters (RGDs) for pencil beam scanning proton therapy.
The RGD’s end-to-end testing of intensity-modulated proton therapy (IMPT) plans was also evaluated.
Materials and Methods:
The dosimetric characteristics of the GD-302M type glass dosimeter were studied in terms of uniformity, short-term and long-term reproducibility, stability of the magazine position readout, dose linearity in the range from 0.
2 to 20 Gy, energy response in 70–220 MeV, and fading effect.
The reference conditions of the spot scanning beam from the Varian ProBeam Compact system were operation at 160 MeV, a 2 cm water-equivalent depth in a solid water phantom, a 10 cm × 10 cm field size at the isocenter, and 2 Gy dose delivery.
End-to-end testing of IMPT plans for the head, abdomen, and pelvis was verified using the Alderson Rando phantom.
The overall uncertainty analysis was confirmed in this study.
Results:
The relative response of RGDs for the uniformity test was within 0.
95–1.
05.
The percentages of the coefficients of variation for short-term and long-term reproducibility were 1.
16% and 1.
50%, respectively.
The dose ACE glass dosimetry reader FGD-1000 showed a stable magazine position readout.
The dose was found to be linear with R
2 = 0.
9988.
The energy response relative to 160 MeV was approximately within 4.
0%.
The fading effect was within 2.
4%.
For the end-to-end test, the difference between the treatment plan and RGD measurement was within 1.
0%.
The overall uncertainty of the RGD measurement for the proton beam was 4.
6%, which covered all energy ranges in this study.
Conclusion:
The experimental study indicates that the RGDs have the potential to be used in the dosimetry of therapeutic proton beams, including end-to-end dosimetry.
Related Results
Dosimetric characteristics of radiophotoluminescent glass dosimeters for proton beams
Dosimetric characteristics of radiophotoluminescent glass dosimeters for proton beams
Abstract
Background: The study of radiophotoluminescent glass dosimeters (RGDs) in the clinical usage of proton beams is limited. The aim of this study was to investigate t...
Odd version Mathieu-Gaussian beam based on Green function
Odd version Mathieu-Gaussian beam based on Green function
Like the theoretical pattern of non-diffracting Bessel beams, ideal non-diffracting Mathieu beams also carry infinite energy, but cannot be generated as a physically realizable ent...
Construction of Bi-Pearcey beams and their mathematical mechanism
Construction of Bi-Pearcey beams and their mathematical mechanism
We present a theoretical expression in the form of the Pearcey function by deducing the Fresnel diffraction distribution of an elliptic line. Then, we numerically simulate and expe...
Dynamic Characteristics Analysis of Three-Layer Steel–Concrete Composite Beams
Dynamic Characteristics Analysis of Three-Layer Steel–Concrete Composite Beams
The dynamic behavior of three-layer composite beams, consisting of concrete slabs and steel beams, is influenced by the structural configuration of each layer as well as the shear ...
Current Developments in Radiotherapy Quality Control Methods
Current Developments in Radiotherapy Quality Control Methods
Quality control is of critical importance to ensure the accurate and reliable implementation of treatment in radiotherapy. Radiotherapy is a primary method of treating diseases suc...
Experimental study and computational analysis of structural performance of reinforced geopolymer concrete beams
Experimental study and computational analysis of structural performance of reinforced geopolymer concrete beams
PurposeIn this study, the aim is to explore the effects of geopolymer concrete (GPC) strength and reinforcement ratio on the flexural performance of reinforced GPC beams. Furthermo...
Investigation of alanine dosimeter for output measurement in radiotherapy: Results from multicenter in Thailand
Investigation of alanine dosimeter for output measurement in radiotherapy: Results from multicenter in Thailand
Background: Alanine dosimeters are generally used in high-dose industrial applications (kGy). Later, research into employing alanine as a dosimeter in radiotherapy (1-20 Gy) has in...
Unraveling the mechanism of proton translocation in the extracellular half-channel of bacteriorhodopsin
Unraveling the mechanism of proton translocation in the extracellular half-channel of bacteriorhodopsin
AbstractBacteriorhodopsin, a light activated protein that creates a proton gradient in halobacteria, has long served as a simple model of proton pumps. Within bacteriorhodopsin, se...