Transprosthetic Fenestration With Electrified Wires. Experimental Evaluation of Three Multifilament Endografts
View through CrossRef
Purpose: In situ fenestration of aortic endografts is an alternative endovascular technique for treatment of complex aortic aneurysms. While this technique has been carried out also to pass stent-grafts in individual cases, its feasibility and safety using different stent-grafts needs to be evaluated. Methods: In a saline bath at water temperature of 37°C, a 0.018” Astato 30 guidewire was advanced through 3 different stent-grafts (RelayPro, Zenith and Endurant II) by applying external current of 180 W via an electrosurgery pencil. Puncture efficacy and quality of the fenestration after ballooning with a 6 mm percutaneous transluminal angioplasty (PTA) catheter were assessed. Then, balloon-expandable covered stents were deployed in the fenestrations and evaluated for stenosis, using microscopy and radiography. Results: Crossing of the electrified guidewire was instantaneous in the Zenith (n:10) and RelayPro (n:10) groups but not in 3 of 10 punctures in the Endurant group (p<.05). The fenestration area created after PTA was significantly larger in the RelayPro (5.3 mm2 ± 1.8, interquartile range [IQR] 1.6) and Zenith group (6.7 mm2 ± 0.7, IQR 0.5) compared to Endurant (2.3 mm2 ± 0.4, IQR 0.5, p<.001). Fraying was observed in all groups while graft shredding was found in 8 cases after PTA of the Zenith and Endurant endografts and in 5 of the RelayPro group, but the difference was not significant. Vertical tearing was detected after RelayPro (2 out of 10) and Zenith (6 out of 10) fenestrations, no damage was found in the Endurant group (p<.01). Residual stenosis at the level of the fenestration after implantation of a 6 × 79 mm VBX stent had to be corrected in all Endurant cases with a high-pressure PTA catheter. No stenosis was found in the RelayPro and Zenith groups before and after flaring. Conclusions: The “electrified wire” technique is a feasible tool that can be used to perform in situ fenestration by perforation of the endograft fabric. Based on this experimental evaluation the “ideal graft” for this technique could not be identified. Long-term fatigue tests and comparison with other fenestration techniques are required. Clinical Impact In situ endograft fenestration can be a useful technique in emergent aortic repair. Recently, the electrified wire technique has been proposed as alternative option to laser, radiofrequency and needle-based techniques. In comparison to these methods, the use of electrified wires can be performed without modifications of routine equipment. Additionally, the material costs can be substantially reduced. However, the effectiveness of this approach for fenestration of different prosthetic grafts is unknown. Based on our experimental studies, the electrified wire technique is feasible but the Endurant endograft requires more attempts, and the placement of a bridging stent should be completed with high-pressure balloons.
Title: Transprosthetic Fenestration With Electrified Wires. Experimental Evaluation of Three Multifilament Endografts
Description:
Purpose: In situ fenestration of aortic endografts is an alternative endovascular technique for treatment of complex aortic aneurysms.
While this technique has been carried out also to pass stent-grafts in individual cases, its feasibility and safety using different stent-grafts needs to be evaluated.
Methods: In a saline bath at water temperature of 37°C, a 0.
018” Astato 30 guidewire was advanced through 3 different stent-grafts (RelayPro, Zenith and Endurant II) by applying external current of 180 W via an electrosurgery pencil.
Puncture efficacy and quality of the fenestration after ballooning with a 6 mm percutaneous transluminal angioplasty (PTA) catheter were assessed.
Then, balloon-expandable covered stents were deployed in the fenestrations and evaluated for stenosis, using microscopy and radiography.
Results: Crossing of the electrified guidewire was instantaneous in the Zenith (n:10) and RelayPro (n:10) groups but not in 3 of 10 punctures in the Endurant group (p<.
05).
The fenestration area created after PTA was significantly larger in the RelayPro (5.
3 mm2 ± 1.
8, interquartile range [IQR] 1.
6) and Zenith group (6.
7 mm2 ± 0.
7, IQR 0.
5) compared to Endurant (2.
3 mm2 ± 0.
4, IQR 0.
5, p<.
001).
Fraying was observed in all groups while graft shredding was found in 8 cases after PTA of the Zenith and Endurant endografts and in 5 of the RelayPro group, but the difference was not significant.
Vertical tearing was detected after RelayPro (2 out of 10) and Zenith (6 out of 10) fenestrations, no damage was found in the Endurant group (p<.
01).
Residual stenosis at the level of the fenestration after implantation of a 6 × 79 mm VBX stent had to be corrected in all Endurant cases with a high-pressure PTA catheter.
No stenosis was found in the RelayPro and Zenith groups before and after flaring.
Conclusions: The “electrified wire” technique is a feasible tool that can be used to perform in situ fenestration by perforation of the endograft fabric.
Based on this experimental evaluation the “ideal graft” for this technique could not be identified.
Long-term fatigue tests and comparison with other fenestration techniques are required.
Clinical Impact In situ endograft fenestration can be a useful technique in emergent aortic repair.
Recently, the electrified wire technique has been proposed as alternative option to laser, radiofrequency and needle-based techniques.
In comparison to these methods, the use of electrified wires can be performed without modifications of routine equipment.
Additionally, the material costs can be substantially reduced.
However, the effectiveness of this approach for fenestration of different prosthetic grafts is unknown.
Based on our experimental studies, the electrified wire technique is feasible but the Endurant endograft requires more attempts, and the placement of a bridging stent should be completed with high-pressure balloons.
Related Results
Comparison of monofilament and multifilament bottom trammel nets regarding catch efficiency and chondrichthyan bycatch in Çanakkale, Türkiye
Comparison of monofilament and multifilament bottom trammel nets regarding catch efficiency and chondrichthyan bycatch in Çanakkale, Türkiye
Abstract
This study aimed to evaluate the catch efficiency and chondrichthyan bycatch of monofilament and multifilament bottom trammel nets in th...
Long-Term Experimental Evaluation of In Situ Laser-Fenestrated Versus Custom-Made Fenestrated Endografts
Long-Term Experimental Evaluation of In Situ Laser-Fenestrated Versus Custom-Made Fenestrated Endografts
Objective:
This study evaluates the biomechanical performance and long-term durability of in situ laser fenestration (ISLF) compared to custom-made fenestrated ...
A HIGH–PRESSURE MYSTERY: UNVEILING THE BODY‘S HIDDEN SHORTCUT
A HIGH–PRESSURE MYSTERY: UNVEILING THE BODY‘S HIDDEN SHORTCUT
Abstract
An arteriovenous fistula (AVF) is an abnormal connection between an artery and a vein, bypassing the capillary network. Acquired AVFs are most commonly caus...
Effectiveness of Fenestration and Discectomy among Patients Presenting with Lumbar Disc Herniation
Effectiveness of Fenestration and Discectomy among Patients Presenting with Lumbar Disc Herniation
Background:Research-StackExchange47, BackgroundLumbar disc herniation (LDH) is a frequent cause of back pain and neurological dysfunction that often requires surgical treatment. Th...
MALDI-TOF Analysis of K-wire Biofilms: A Pilot Study
MALDI-TOF Analysis of K-wire Biofilms: A Pilot Study
Category: Basic Sciences/Biologics Introduction/Purpose: In orthopedic foot and ankle surgery, temporary fixation utilizing percutaneous Kirschner wires (K- wires) is required for ...
Selectivity Evaluation for Two Experimental Gill-Net Configurations Used to Sample Lake Erie Walleyes
Selectivity Evaluation for Two Experimental Gill-Net Configurations Used to Sample Lake Erie Walleyes
Abstract
We used length frequencies of captured walleyes Sander vitreus to indirectly estimate and compare selectivity between two experimental gill-net configuratio...
REVIT SIMULATION APPLICATION IN EVALUATING FENESTRATION STRATEGIES TOWARDS ENERGY CONSERVATION IN MALAYSIAN INTERMEDIATE TERRACED HOUSES
REVIT SIMULATION APPLICATION IN EVALUATING FENESTRATION STRATEGIES TOWARDS ENERGY CONSERVATION IN MALAYSIAN INTERMEDIATE TERRACED HOUSES
With Malaysia’s population rising and rapid urbanisation, terraced houses were built to respond to new social and economic needs. The modern terraced house appeals to the demand f...
Application of the Bolton Relay Device for Thoracic Endografting In or Near the Aortic Arch
Application of the Bolton Relay Device for Thoracic Endografting In or Near the Aortic Arch
AbstractEndovascular correction of aortic arch pathology remains a challenge, with a variety of techniques proposed over the years to minimize complications and enhance the probabi...