A Recapping System for Automatic, Semiautomatic, and Manual Use
View through CrossRef
Abstract
Context.
—Closed sample tubes are needed for stored as well as in-process samples because evaporation significantly changes analyte concentrations. In modern laboratories, it is essential that the procedures for ordering tests on stored serum samples be simple and time-efficient. Therefore, recapping and reopening of tubes should be readily possible in laboratories to save resources and time.
Objective.
—To evaluate the recapping system MagCap, in comparison to established snap caps and unsealed tubes.
Design.
—Specific balls, with an iron core, fit tightly by their specific weight into openings of primary and secondary tubes as well as microtiter plates. Recapping of the tubes can be performed individually by a single dispenser or rackwise by a multidispenser. Balls can be removed individually by a pencillike magnet or rackwise. Balls can be reused after cleaning in a glassware washer or used as disposables.
Results.
—Balls seal the sample tubes as effectively as snap caps and decrease evaporation significantly in comparison to unsealed tubes. Evaporation effects of up to 13% were detected in unsealed tubes (0.5 mL) within 24 hours, whereas resealing of the tubes diminished the concentration effect to less than 1%. Open samples (0.5 mL) showed a concentration effect in the refrigerator within 7 days of up to 25%. The evaporation effect of the sealed tubes was less than 3%. Multiple reopenening and recapping procedures are easily possible using magnetic forces.
Conclusions.
—The MagCap recapping system seals primary and secondary containers securely and cost-effectively with the ball cap and thus facilitates a necessary advance in sample quality.
Archives of Pathology and Laboratory Medicine
Title: A Recapping System for Automatic, Semiautomatic, and Manual Use
Description:
Abstract
Context.
—Closed sample tubes are needed for stored as well as in-process samples because evaporation significantly changes analyte concentrations.
In modern laboratories, it is essential that the procedures for ordering tests on stored serum samples be simple and time-efficient.
Therefore, recapping and reopening of tubes should be readily possible in laboratories to save resources and time.
Objective.
—To evaluate the recapping system MagCap, in comparison to established snap caps and unsealed tubes.
Design.
—Specific balls, with an iron core, fit tightly by their specific weight into openings of primary and secondary tubes as well as microtiter plates.
Recapping of the tubes can be performed individually by a single dispenser or rackwise by a multidispenser.
Balls can be removed individually by a pencillike magnet or rackwise.
Balls can be reused after cleaning in a glassware washer or used as disposables.
Results.
—Balls seal the sample tubes as effectively as snap caps and decrease evaporation significantly in comparison to unsealed tubes.
Evaporation effects of up to 13% were detected in unsealed tubes (0.
5 mL) within 24 hours, whereas resealing of the tubes diminished the concentration effect to less than 1%.
Open samples (0.
5 mL) showed a concentration effect in the refrigerator within 7 days of up to 25%.
The evaporation effect of the sealed tubes was less than 3%.
Multiple reopenening and recapping procedures are easily possible using magnetic forces.
Conclusions.
—The MagCap recapping system seals primary and secondary containers securely and cost-effectively with the ball cap and thus facilitates a necessary advance in sample quality.
Related Results
Correlation between manual and automatic measurements of the atrial signal in VDD leadless pacemakers (Micra AV)
Correlation between manual and automatic measurements of the atrial signal in VDD leadless pacemakers (Micra AV)
Abstract
Introduction
The Micra AV leadless pacemaker (model MC1AVR1, Medtronic, MN, US) provides atrio-ventricular (AV) synchro...
Closed-Loop Oxygen Control
Closed-Loop Oxygen Control
<p>Guidelines recommend that oxygen should be titrated to achieve a target oxygen saturation (SpO2 ) range in acutely unwell patients, a concept colloquially known as “swimmi...
Semiautomatic inverse filtering of vowels with differing degrees of nasality
Semiautomatic inverse filtering of vowels with differing degrees of nasality
Most approaches to inverse filtering are either manual or automatic. Manual methods, while providing high-quality fine-grained analysis, require user expertise and manpower to anal...
Converting Malfunctioned Fully Automatic (Electro-Mechanical Driven) to Semiautomatic Washing Machine
Converting Malfunctioned Fully Automatic (Electro-Mechanical Driven) to Semiautomatic Washing Machine
This study converted a damaged fully automatic washing machine into a semi -automatic washing machine. Conversion of the machine was done due to unavailability of its electronic sp...
Solving the problems of underwater mecha-nized and automatic arc welding of various metal structures with a wet method
Solving the problems of underwater mecha-nized and automatic arc welding of various metal structures with a wet method
A complex of problems related to semi-automatic and automatic underwater wet welding, concerning both equipment and ways of implementation of arc process, has been considered. An a...
Interobserver Agreement in Automatic Segmentation Annotation of Prostate Magnetic Resonance Imaging
Interobserver Agreement in Automatic Segmentation Annotation of Prostate Magnetic Resonance Imaging
We aimed to compare the performance and interobserver agreement of radiologists manually segmenting images or those assisted by automatic segmentation. We further aimed to reduce i...
Measurement of the Particulate Matter Emission from Internal Combustion Engines Using an Automatic Method
Measurement of the Particulate Matter Emission from Internal Combustion Engines Using an Automatic Method
<div class="section abstract"><div class="htmlview paragraph">The aim of this study was to evaluate the accuracy and precision of measurements of an automatic weighing ...
Evaluation of automatic pericardial segmentation methods in computed tomography images
Evaluation of automatic pericardial segmentation methods in computed tomography images
Abstract
Funding Acknowledgements
Type of funding sources: Public grant(s) – National budget only. Main funding source(s): Funda...