The Accuracy of Pulse Oxygen Saturation, Heart Rate, Blood Pressure, and Respiratory Rate Raised by a Contactless Telehealth Portal: Validation Study (Preprint)

View through CrossRef
BACKGROUND The traditional measurement of heart rate (HR), oxygen saturation (SpO<sub>2</sub>), blood pressure (BP), and respiratory rate (RR) via physical examination can be challenging, and the recent pandemic has accelerated trends toward telehealth and remote monitoring. Instead of going to the physician to check these vital signs, measuring them at home would be more convenient. Vital sign monitors, also known as physiological parameter monitors, are electronic devices that measure and display biological information about patients under constant monitoring. OBJECTIVE The purpose of this study was to validate the accuracy of the pulse SpO<sub>2</sub>, HR, BP, and RR raised by Docsun Telehealth Portal by comparing it with approved medical devices. METHODS This is a noninvasive, self-check, system-based study conducted to validate the detection of vital signs (SpO<sub>2</sub>, HR, BP, and RR) raised by Docsun Telehealth Portal. The input for software processing involves facial screening without any accessories on the face, scanning directly through the software application portal. The participant’s facial features are detected and screened for the extraction of necessary readings. RESULTS For the validation of HR, SpO<sub>2</sub>, BP, and RR measurements, the main outcomes were the mean of the absolute difference between the respective investigational devices and the reference values as well as the absolute percentage difference between the respective investigational devices and the reference values. If the HR was within ±10% of the reference standard or 5 beats per minute, it was considered acceptable for clinical purposes. The average absolute difference between the Docsun Telehealth Portal and the reference values was 1.41 (SD 1.14) beats per minute. The mean absolute percentage difference was 1.69% (SD 1.37). Therefore, the Docsun Telehealth Portal met the predefined accuracy cutoff for HR measurements. If the RR was within ±10% of the reference standard or 3 breaths per minute, it was considered acceptable for clinical purposes. The average absolute difference between the Docsun Telehealth Portal and the reference values was 0.86 breaths per minute. The mean absolute percentage difference was 4.72%. Therefore, the Docsun Telehealth Portal met the predefined accuracy cutoff for RR measurements. SpO<sub>2</sub> levels were considered acceptable if the average absolute difference between the Docsun Telehealth Portal and the reference values was ±3%. The mean absolute percentage difference was 0.59%. Therefore, the Docsun Telehealth Portal met the predefined accuracy cutoff for SpO<sub>2</sub> measurements. The Docsun Telehealth Portal predicted systolic BP with an accuracy of 94.81% and diastolic BP with an accuracy of 95.71%. CONCLUSIONS The results of the study show that the accuracy of the HR, BP, SpO<sub>2</sub>, and RR values raised by the Docsun Telehealth Portal, compared against the clinically approved medical devices, proved to be accurate by meeting predefined accuracy guidelines.
Title: The Accuracy of Pulse Oxygen Saturation, Heart Rate, Blood Pressure, and Respiratory Rate Raised by a Contactless Telehealth Portal: Validation Study (Preprint)
Description:
BACKGROUND The traditional measurement of heart rate (HR), oxygen saturation (SpO<sub>2</sub>), blood pressure (BP), and respiratory rate (RR) via physical examination can be challenging, and the recent pandemic has accelerated trends toward telehealth and remote monitoring.
Instead of going to the physician to check these vital signs, measuring them at home would be more convenient.
Vital sign monitors, also known as physiological parameter monitors, are electronic devices that measure and display biological information about patients under constant monitoring.
OBJECTIVE The purpose of this study was to validate the accuracy of the pulse SpO<sub>2</sub>, HR, BP, and RR raised by Docsun Telehealth Portal by comparing it with approved medical devices.
METHODS This is a noninvasive, self-check, system-based study conducted to validate the detection of vital signs (SpO<sub>2</sub>, HR, BP, and RR) raised by Docsun Telehealth Portal.
The input for software processing involves facial screening without any accessories on the face, scanning directly through the software application portal.
The participant’s facial features are detected and screened for the extraction of necessary readings.
RESULTS For the validation of HR, SpO<sub>2</sub>, BP, and RR measurements, the main outcomes were the mean of the absolute difference between the respective investigational devices and the reference values as well as the absolute percentage difference between the respective investigational devices and the reference values.
If the HR was within ±10% of the reference standard or 5 beats per minute, it was considered acceptable for clinical purposes.
The average absolute difference between the Docsun Telehealth Portal and the reference values was 1.
41 (SD 1.
14) beats per minute.
The mean absolute percentage difference was 1.
69% (SD 1.
37).
Therefore, the Docsun Telehealth Portal met the predefined accuracy cutoff for HR measurements.
If the RR was within ±10% of the reference standard or 3 breaths per minute, it was considered acceptable for clinical purposes.
The average absolute difference between the Docsun Telehealth Portal and the reference values was 0.
86 breaths per minute.
The mean absolute percentage difference was 4.
72%.
Therefore, the Docsun Telehealth Portal met the predefined accuracy cutoff for RR measurements.
SpO<sub>2</sub> levels were considered acceptable if the average absolute difference between the Docsun Telehealth Portal and the reference values was ±3%.
The mean absolute percentage difference was 0.
59%.
Therefore, the Docsun Telehealth Portal met the predefined accuracy cutoff for SpO<sub>2</sub> measurements.
The Docsun Telehealth Portal predicted systolic BP with an accuracy of 94.
81% and diastolic BP with an accuracy of 95.
71%.
CONCLUSIONS The results of the study show that the accuracy of the HR, BP, SpO<sub>2</sub>, and RR values raised by the Docsun Telehealth Portal, compared against the clinically approved medical devices, proved to be accurate by meeting predefined accuracy guidelines.

Related Results

[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Improving Rural Health Through Telehealth-Guided Provider-to-Provider Communication
Improving Rural Health Through Telehealth-Guided Provider-to-Provider Communication
Objectives. To assess the use, effectiveness, and implementation of telehealth-supported provider-to-provider communication and collaboration for the provision of healthcare servic...
Continued Use of Telehealth In Oncology Patients (Preprint)
Continued Use of Telehealth In Oncology Patients (Preprint)
BACKGROUND Prior to the Covid-19 pandemic, telehealth represented a small portion of our oncology practice. Despite offering telehealth availability, fewer ...
High Concentration Oxygen and Hypercapnia in Respiratory Disease
High Concentration Oxygen and Hypercapnia in Respiratory Disease
<p>Oxygen-induced elevations in arterial carbon dioxide tension have been demonstrated in patients with chronic obstructive pulmonary disease (COPD), asthma, pneumonia, obesi...
0864 Severe Central Sleep Apnea
0864 Severe Central Sleep Apnea
Abstract Introduction Central sleep apnea (CSA) is a rare form of sleep disordered breathing with repeated apneic episodes with ...
Knowledge comparison amongst telehealth service utilized and never-utilized adults in Bangladesh: a cross-sectional study
Knowledge comparison amongst telehealth service utilized and never-utilized adults in Bangladesh: a cross-sectional study
Background: Telehealth service is an approach to health care delivery that uses various telecommunication technologies, where the knowledge of the patient plays an impo...

Back to Top