The effects of scatter in x‐ray computed tomography

View through CrossRef
The effects of detection of scattered radiation in x‐ray transmission CT are studied both theoretically and experimentally. It is shown that scatter induces nonlinear errors in the measurement of attenuation values which can lead to cupping, streaks, and CT number inaccuracies. It is shown that scatter effects predominate over beam spectrum hardening effects for large body parts, and that the artifact propensity is a direct function of the scatter‐to‐primary ratio. The presence of scatter induced streaks were demonstrated experimentally on both a third‐ and a fourth‐generation CT scanner using an appropriate water‐equivalent phantom. A simple model with constant scatter background leads to a correction algorithm which was tested on several phantoms and one human pelvis. The algorithm worked well on the smaller phantoms but was less successful for the larger objects. Nevertheless, it still gave substantial improvement in the pelvic scan. We demonstrated that, at least in the pelvis, scatter is a more significant source of error than beam hardening and that improved scatter correction algorithms are needed. The consequences for the quantitative interpretation of CT numbers for clinical diagnoses are discussed.
Title: The effects of scatter in x‐ray computed tomography
Description:
The effects of detection of scattered radiation in x‐ray transmission CT are studied both theoretically and experimentally.
It is shown that scatter induces nonlinear errors in the measurement of attenuation values which can lead to cupping, streaks, and CT number inaccuracies.
It is shown that scatter effects predominate over beam spectrum hardening effects for large body parts, and that the artifact propensity is a direct function of the scatter‐to‐primary ratio.
The presence of scatter induced streaks were demonstrated experimentally on both a third‐ and a fourth‐generation CT scanner using an appropriate water‐equivalent phantom.
A simple model with constant scatter background leads to a correction algorithm which was tested on several phantoms and one human pelvis.
The algorithm worked well on the smaller phantoms but was less successful for the larger objects.
Nevertheless, it still gave substantial improvement in the pelvic scan.
We demonstrated that, at least in the pelvis, scatter is a more significant source of error than beam hardening and that improved scatter correction algorithms are needed.
The consequences for the quantitative interpretation of CT numbers for clinical diagnoses are discussed.

Related Results

Recalculation of scatter fractions for homogeneous and heterogeneous geometries using Geant4 Monte Carlo simulations
Recalculation of scatter fractions for homogeneous and heterogeneous geometries using Geant4 Monte Carlo simulations
Abstract The primary aim of this study was to determine the scatter fraction from patient, particularly in the presence of patient heterogeneities, using Monte Carlo...
Using dual-energy CT to discriminate sediment facies in a varved sequence 
Using dual-energy CT to discriminate sediment facies in a varved sequence 
The attenuation of X-ray is influenced by the density (electron density, ρe) and the elemental composition (effective atomic number, Zeff) of the object being imaged. The ...
Determination of the Transition Temperature Scatter Using the Charpy Data Scatter
Determination of the Transition Temperature Scatter Using the Charpy Data Scatter
For a pressure vessels integrity assessment it is necessary to know the transition temperature of metal. It is defined from the hyperbolic tangent curve, which best approximates th...
Health Risk From Radioactive Iodine (RAI) Therapy And Medical Imaging- A Short Review
Health Risk From Radioactive Iodine (RAI) Therapy And Medical Imaging- A Short Review
Different role of radioactive iodine (RAI) has a different dosage. RAI is the effective treatment for cancer but has a short or long side effects such as xerostomia and loss of tas...
A Review: Tomography Systems in Medical and Industrial Processes
A Review: Tomography Systems in Medical and Industrial Processes
Conventional methods previously guided reading meter to monitor the process and steps must be taken to interpret the readings and what need to adjustment be implemented. With the u...

Back to Top