II–3 Compliance and the Bone Fusion Rate of Low-Intensity Pulsed Ultrasound (LIPUS) Treatment
View through CrossRef
Objective:
The aim of this study is to clarify the necessity of the compliance (usage) rate to succeed bone union by LIPUS treatment device in the long bone fracture and the clavicular fracture.
Materials and Methods:
The subjects are 229 cases who were treated with the LIPUS device whose utilization rate was known after the surgical operation for fresh fractures, from 2008 to 2017. We have defined the LIPUS compliance rate (Number of days during which LIPUS was actually used/Number of days during which LIPUS was rented × 100). We have evaluated that LIPUS compliance rate. In a retrospective cohort study, outcomes were obtained that bone union was succeeded by LIPUS treatment alone. We investigated the compliance rate of LIPUS therapy equipment, conducted χ2 test of univariate analysis, and conducted Receiver Operator Characteristic (ROC) analysis on how much should be used.
Results:
Two hundred twenty-one Bones (96.5%) were bone union with LIPUS alone but 8 bones (3.5%) needed additional medical surgery. The average of compliance rate for LIPUS therapy was 75.9%. In the case of low compliance of LIPUS usage, there were failures of LIPUS treatment which is significant (P < 0.01). The cutoff value of the compliance rate in the ROC analysis was 57.1% (sensitivity 79.2%, specificity 75%, AUC [Area Under the Curve] = 0.779), for significantly difference in bone fusion ratio.
Conclusion:
The compliance (usage) rate of LIPUS treatment must be at least 60% to expect effective bone union.
Ovid Technologies (Wolters Kluwer Health)
Title: II–3 Compliance and the Bone Fusion Rate of Low-Intensity Pulsed Ultrasound (LIPUS) Treatment
Description:
Objective:
The aim of this study is to clarify the necessity of the compliance (usage) rate to succeed bone union by LIPUS treatment device in the long bone fracture and the clavicular fracture.
Materials and Methods:
The subjects are 229 cases who were treated with the LIPUS device whose utilization rate was known after the surgical operation for fresh fractures, from 2008 to 2017.
We have defined the LIPUS compliance rate (Number of days during which LIPUS was actually used/Number of days during which LIPUS was rented × 100).
We have evaluated that LIPUS compliance rate.
In a retrospective cohort study, outcomes were obtained that bone union was succeeded by LIPUS treatment alone.
We investigated the compliance rate of LIPUS therapy equipment, conducted χ2 test of univariate analysis, and conducted Receiver Operator Characteristic (ROC) analysis on how much should be used.
Results:
Two hundred twenty-one Bones (96.
5%) were bone union with LIPUS alone but 8 bones (3.
5%) needed additional medical surgery.
The average of compliance rate for LIPUS therapy was 75.
9%.
In the case of low compliance of LIPUS usage, there were failures of LIPUS treatment which is significant (P < 0.
01).
The cutoff value of the compliance rate in the ROC analysis was 57.
1% (sensitivity 79.
2%, specificity 75%, AUC [Area Under the Curve] = 0.
779), for significantly difference in bone fusion ratio.
Conclusion:
The compliance (usage) rate of LIPUS treatment must be at least 60% to expect effective bone union.
Related Results
Effect of Low-Intensity Pulsed Ultrasound (LIPUS) on Osteotomy
Effect of Low-Intensity Pulsed Ultrasound (LIPUS) on Osteotomy
Objective:
LIPUS following osteotomy has been covered by national health insurance since April 2015. We investigated the effect of LIPUS on the time period for callus f...
Low-intensity pulsed ultrasound does not promote bone healing and functional recovery after open wedge high tibial osteotomy
Low-intensity pulsed ultrasound does not promote bone healing and functional recovery after open wedge high tibial osteotomy
Aims To evaluate whether low-intensity pulsed ultrasound (LIPUS) accelerates bone healing at osteotomy sites and promotes functional recovery after open-wedge high tibial osteotomy...
Outcomes of Treatment with Low-Intensity Pulsed Ultrasound After Opening Wedge High Tibial Osteotomy
Outcomes of Treatment with Low-Intensity Pulsed Ultrasound After Opening Wedge High Tibial Osteotomy
Objective:
We have retrospectively reviewed the outcomes of treatment with low-intensity pulsed ultrasound (LIPUS) in patients who underwent opening wedge high tibial o...
The Nuclear Fusion Award
The Nuclear Fusion Award
The Nuclear Fusion Award ceremony for 2009 and 2010 award winners was held during the 23rd IAEA Fusion Energy Conference in Daejeon. This time, both 2009 and 2010 award winners w...
Enhancement of Low Intensity Pulsed Ultrasound on Spinal Fusion augmented with Stem cell-Bioceramic Composite
Enhancement of Low Intensity Pulsed Ultrasound on Spinal Fusion augmented with Stem cell-Bioceramic Composite
Low intensity pulsed ultrasound (LIPUS) was reported to enhance autograft implanted spinal fusion in rabbit model but none reported its effect on bioengineered stem cell-bioceramic...
Poster 107: The Use of Coacervate Sustained Release System to Identify the Most Potent BMP for Bone Regeneration
Poster 107: The Use of Coacervate Sustained Release System to Identify the Most Potent BMP for Bone Regeneration
Objectives: Bone morphogenetic proteins (BMPs) belong to the transforming growth factor superfamily that were first discovered by Marshall Urist. There are 14 BMPs identified to da...
Marginal Bone Changes around Dental Implants after LIPUS Application: CBCT Study
Marginal Bone Changes around Dental Implants after LIPUS Application: CBCT Study
To assess the effect of LIPUS on marginal bone regeneration during insertion and following loading using CBCT scan imaging, a trial of RCT of 22 subjects needing dental implant was...
Low intensity pulsed ultrasound in orthodontic tooth movement
Low intensity pulsed ultrasound in orthodontic tooth movement
Orthodontic treatment that takes a long time carries a large risk of complications such as caries or root resorption. Various surgical and nonsurgical interventions have been perfo...