The efficacy of apneic oxygenation during intubation using a prototype of an oxygenation laryngoscope - a technical simulation
View through CrossRef
Abstract
Background
Recently, a non-commercial oxygenation laryngoscope was able to maintain apneic oxygenation during simulated intubation efforts. Since that prototype was 3 mm wider than a standard Macintosh laryngoscope blade, the intubation performance of this device may differ from standard blades. A new prototype of an oxygenation laryngoscope was developed, consisting of a standard-size Macintosh blade and a fixed oxygen supply line to the side. Actually, it is unclear at which point of this blade the oxygen supply line should end to facilitate the best possible oxygen supply for apneic oxygenation.
Methods
In this simulation study using a standardized human airway manikin, the efficacy of apneic oxygenation by oxygen insufflation using standard and modified Macintosh blades was compared: a standard Macintosh blade without oxygen supply line as control, one with an additional oxygen supply line ending proximal near the handle, one with the line ending at the middle of the blade, and one with the line ending near the tip. A preoxygenated test lung was connected to an oximeter with a flow rate of 200ml/min, simulating oxygen consumption of a male adult, and to the trachea of an anatomically correctly shaped airway manikin. Apneic oxygenation was performed and oxygen content was measured over a 20-minutes observation period. Experiments were repeated five times for each laryngoscope blade.
Results
Oxygen percentage in the test lung dropped from 100 ± 0% at the start of the experiment to 53 ± 1.5% in the room air control group (p < 0.001 compared to all other groups), and to 74 ± 2.5% in the proximal oxygen line group, whereas oxygen percentage remained at 100% in both the medium and distal oxygen line groups (p = 1 between these groups; p < 0.001 between all other groups).
Conclusions
In this simulation study with a preoxygenated airway manikin, the use of a modified Macintosh laryngoscope blade with oxygen line attached at the tip or at the middle were able to maintain apneic oxygenation without measurable drop of oxygen content over 20 min. Proximal placement of the oxygen supply line still showed an advantage against room air, however it did not completely prevent room air from entering the airway.
Trial registration
Not applicable.
Springer Science and Business Media LLC
Title: The efficacy of apneic oxygenation during intubation using a prototype of an oxygenation laryngoscope - a technical simulation
Description:
Abstract
Background
Recently, a non-commercial oxygenation laryngoscope was able to maintain apneic oxygenation during simulated intubation efforts.
Since that prototype was 3 mm wider than a standard Macintosh laryngoscope blade, the intubation performance of this device may differ from standard blades.
A new prototype of an oxygenation laryngoscope was developed, consisting of a standard-size Macintosh blade and a fixed oxygen supply line to the side.
Actually, it is unclear at which point of this blade the oxygen supply line should end to facilitate the best possible oxygen supply for apneic oxygenation.
Methods
In this simulation study using a standardized human airway manikin, the efficacy of apneic oxygenation by oxygen insufflation using standard and modified Macintosh blades was compared: a standard Macintosh blade without oxygen supply line as control, one with an additional oxygen supply line ending proximal near the handle, one with the line ending at the middle of the blade, and one with the line ending near the tip.
A preoxygenated test lung was connected to an oximeter with a flow rate of 200ml/min, simulating oxygen consumption of a male adult, and to the trachea of an anatomically correctly shaped airway manikin.
Apneic oxygenation was performed and oxygen content was measured over a 20-minutes observation period.
Experiments were repeated five times for each laryngoscope blade.
Results
Oxygen percentage in the test lung dropped from 100 ± 0% at the start of the experiment to 53 ± 1.
5% in the room air control group (p < 0.
001 compared to all other groups), and to 74 ± 2.
5% in the proximal oxygen line group, whereas oxygen percentage remained at 100% in both the medium and distal oxygen line groups (p = 1 between these groups; p < 0.
001 between all other groups).
Conclusions
In this simulation study with a preoxygenated airway manikin, the use of a modified Macintosh laryngoscope blade with oxygen line attached at the tip or at the middle were able to maintain apneic oxygenation without measurable drop of oxygen content over 20 min.
Proximal placement of the oxygen supply line still showed an advantage against room air, however it did not completely prevent room air from entering the airway.
Trial registration
Not applicable.
Related Results
Comparison of king vision video laryngoscope versus mccoy laryngoscope for endotracheal intubation in patients with immobilized cervical spine
Comparison of king vision video laryngoscope versus mccoy laryngoscope for endotracheal intubation in patients with immobilized cervical spine
Background: Objective of this study was to compare King vision video laryngoscope versus McCoy laryngoscope for endotracheal intubation in patients with simulated immobilized cervi...
COMPARATIVE EVALUATION OF AIRTRAQ OPTICAL LARYNGOSCOPE AND HUGE MED VIDEO LARYNGOSCOPE FOR LARYNGEAL VISUALISATION AND ENDOTRACHEAL INTUBATION IN ADULTS WITH ANTICIPATED DIFFICULT INTUBATION
COMPARATIVE EVALUATION OF AIRTRAQ OPTICAL LARYNGOSCOPE AND HUGE MED VIDEO LARYNGOSCOPE FOR LARYNGEAL VISUALISATION AND ENDOTRACHEAL INTUBATION IN ADULTS WITH ANTICIPATED DIFFICULT INTUBATION
Background and Aims: Complications arising from difficult or failed tracheal intubation still remains a leading cause of anaesthetic morbidity and mortality.This has lead to develo...
Comparison of Airtraq® Laryngoscope, Bonfils Endoscope and Fiberoptic Bronchoscope for Awake Tracheal Intubation: A Randomized, Controlled Trial
Comparison of Airtraq® Laryngoscope, Bonfils Endoscope and Fiberoptic Bronchoscope for Awake Tracheal Intubation: A Randomized, Controlled Trial
Over the last decades several indirect laryngoscopes have been developed to provide a significant better glottic view and improved the success rate in difficult intubations. Some c...
Impact clinique des techniques non-invasives d'oxygénation au cours de l'insuffisance respiratoire aiguë
Impact clinique des techniques non-invasives d'oxygénation au cours de l'insuffisance respiratoire aiguë
L’oxygénothérapie conventionnelle, l’oxygénothérapie nasale à haut débit (OHD, et la ventilation non-invasive (VNI), sont les trois techniques d’oxygénation disponibles en réanimat...
Comparison Between C-MAC Blade D Video Laryngoscope and Macintosh Laryngoscope for Insertion of Double Lumen Tube in Patients Undergoing Elective Thoracic Surgeries
Comparison Between C-MAC Blade D Video Laryngoscope and Macintosh Laryngoscope for Insertion of Double Lumen Tube in Patients Undergoing Elective Thoracic Surgeries
The insertion of a double lumen tube (DLT) for lung isolation during thoracic surgeries is technically challenging and may be associated with airway trauma and significant hemodyna...
DirEct versus VIdeo LaryngosCopE (DEVICE): protocol and statistical analysis plan for a randomised clinical trial in critically ill adults undergoing emergency tracheal intubation
DirEct versus VIdeo LaryngosCopE (DEVICE): protocol and statistical analysis plan for a randomised clinical trial in critically ill adults undergoing emergency tracheal intubation
IntroductionAmong critically ill patients undergoing orotracheal intubation in the emergency department (ED) or intensive care unit (ICU), failure to visualise the vocal cords and ...
Proficiencies of military medical officers in intubating difficult airways
Proficiencies of military medical officers in intubating difficult airways
Abstract
Background
This study sheds light on the proficiency of military medical officers who had received between 2 and...
Proficiencies of Military Medical Officers in Intubating Difficult Airways
Proficiencies of Military Medical Officers in Intubating Difficult Airways
Abstract
Background This study sheds light on the proficiency of military medical officers who had received between 2-3 years of post-graduate training, in the handling of ...