The efficacy of apneic oxygenation during intubation using a prototype of an oxygenation laryngoscope - a technical simulation

View through CrossRef
Abstract Background Recently, a non-commercial oxygenation laryngoscope was able to maintain apneic oxygenation during simulated intubation efforts. Since that prototype was 3 mm wider than a standard Macintosh laryngoscope blade, the intubation performance of this device may differ from standard blades. A new prototype of an oxygenation laryngoscope was developed, consisting of a standard-size Macintosh blade and a fixed oxygen supply line to the side. Actually, it is unclear at which point of this blade the oxygen supply line should end to facilitate the best possible oxygen supply for apneic oxygenation. Methods In this simulation study using a standardized human airway manikin, the efficacy of apneic oxygenation by oxygen insufflation using standard and modified Macintosh blades was compared: a standard Macintosh blade without oxygen supply line as control, one with an additional oxygen supply line ending proximal near the handle, one with the line ending at the middle of the blade, and one with the line ending near the tip. A preoxygenated test lung was connected to an oximeter with a flow rate of 200ml/min, simulating oxygen consumption of a male adult, and to the trachea of an anatomically correctly shaped airway manikin. Apneic oxygenation was performed and oxygen content was measured over a 20-minutes observation period. Experiments were repeated five times for each laryngoscope blade. Results Oxygen percentage in the test lung dropped from 100 ± 0% at the start of the experiment to 53 ± 1.5% in the room air control group (p < 0.001 compared to all other groups), and to 74 ± 2.5% in the proximal oxygen line group, whereas oxygen percentage remained at 100% in both the medium and distal oxygen line groups (p = 1 between these groups; p < 0.001 between all other groups). Conclusions In this simulation study with a preoxygenated airway manikin, the use of a modified Macintosh laryngoscope blade with oxygen line attached at the tip or at the middle were able to maintain apneic oxygenation without measurable drop of oxygen content over 20 min. Proximal placement of the oxygen supply line still showed an advantage against room air, however it did not completely prevent room air from entering the airway. Trial registration Not applicable.
Title: The efficacy of apneic oxygenation during intubation using a prototype of an oxygenation laryngoscope - a technical simulation
Description:
Abstract Background Recently, a non-commercial oxygenation laryngoscope was able to maintain apneic oxygenation during simulated intubation efforts.
Since that prototype was 3 mm wider than a standard Macintosh laryngoscope blade, the intubation performance of this device may differ from standard blades.
A new prototype of an oxygenation laryngoscope was developed, consisting of a standard-size Macintosh blade and a fixed oxygen supply line to the side.
Actually, it is unclear at which point of this blade the oxygen supply line should end to facilitate the best possible oxygen supply for apneic oxygenation.
Methods In this simulation study using a standardized human airway manikin, the efficacy of apneic oxygenation by oxygen insufflation using standard and modified Macintosh blades was compared: a standard Macintosh blade without oxygen supply line as control, one with an additional oxygen supply line ending proximal near the handle, one with the line ending at the middle of the blade, and one with the line ending near the tip.
A preoxygenated test lung was connected to an oximeter with a flow rate of 200ml/min, simulating oxygen consumption of a male adult, and to the trachea of an anatomically correctly shaped airway manikin.
Apneic oxygenation was performed and oxygen content was measured over a 20-minutes observation period.
Experiments were repeated five times for each laryngoscope blade.
Results Oxygen percentage in the test lung dropped from 100 ± 0% at the start of the experiment to 53 ± 1.
5% in the room air control group (p < 0.
001 compared to all other groups), and to 74 ± 2.
5% in the proximal oxygen line group, whereas oxygen percentage remained at 100% in both the medium and distal oxygen line groups (p = 1 between these groups; p < 0.
001 between all other groups).
Conclusions In this simulation study with a preoxygenated airway manikin, the use of a modified Macintosh laryngoscope blade with oxygen line attached at the tip or at the middle were able to maintain apneic oxygenation without measurable drop of oxygen content over 20 min.
Proximal placement of the oxygen supply line still showed an advantage against room air, however it did not completely prevent room air from entering the airway.
Trial registration Not applicable.

Related Results

Approach Combining the Airway Scope and the Bougie for Minimizing Movement of the Cervical Spine during Endotracheal Intubation
Approach Combining the Airway Scope and the Bougie for Minimizing Movement of the Cervical Spine during Endotracheal Intubation
Background The Airway Scope (AWS, AWS-S100; Hoya-Pentax, Tokyo, Japan), a recently introduced video laryngoscope, has been reported to reduce movement of the cervical s...
Clinical evaluation of Ambu® AuraGain™ as a conduit for intubation in paediatric patients: a descriptive study
Clinical evaluation of Ambu® AuraGain™ as a conduit for intubation in paediatric patients: a descriptive study
Introduction: Many paediatric difficult airway guidelines have recommended supraglottic airway devices (SGAs) as an indispensable tool in the algorithm for managing failed intubati...
Rocuronium versus Cisatracurium for Rapid Sequence Induction of Anesthesia in Morbidly Obese Patients
Rocuronium versus Cisatracurium for Rapid Sequence Induction of Anesthesia in Morbidly Obese Patients
Introduction and Aim: Obesity is a challenging problem for anesthetists. Difficult airway with risk of aspiration and regurgitation increases with obesity. Therefore, the need for ...
A spray-as-you-go airway topical anesthesia attenuates cardiovascular responses for double-lumen tube tracheal intubation
A spray-as-you-go airway topical anesthesia attenuates cardiovascular responses for double-lumen tube tracheal intubation
Abstract Background: The spray-as-you-go airway topical anesthesia and nerve block technique are commonly used in awake tracheal intubation. However, their effects have not...
Learning endotracheal intubation with the Video Endotracheal Tube Guide
Learning endotracheal intubation with the Video Endotracheal Tube Guide
Introduction Video laryngoscopes facilitate the visualization of the glottis but do not guarantee endotracheal intubation due to difficulties in guiding the endotracheal tube to t...

Back to Top