Click-on fluorescence detectors: using robotic surgical instruments to characterize molecular tissue aspects
View through CrossRef
AbstractFluorescence imaging is increasingly being implemented in surgery. One of the drawbacks of its application is the need to switch back-and-forth between fluorescence- and white-light-imaging settings and not being able to dissect safely under fluorescence guidance. The aim of this study was to engineer ‘click-on’ fluorescence detectors that transform standard robotic instruments into molecular sensing devices that enable the surgeon to detect near-infrared (NIR) fluorescence in a white-light setting. This NIR-fluorescence detector setup was engineered to be press-fitted onto standard forceps instruments of the da Vinci robot. Following system characterization in a phantom setting (i.e., spectral properties, sensitivity and tissue signal attenuation), the performance with regard to different clinical indocyanine green (ICG) indications (e.g., angiography and lymphatic mapping) was determined via robotic surgery in pigs. To evaluate in-human applicability, the setup was also used for ICG-containing lymph node specimens from robotic prostate cancer surgery. The resulting Click-On device allowed for NIR ICG signal identification down to a concentration of 4.77 × 10–6 mg/ml. The fully assembled system could be introduced through the trocar and grasping, and movement abilities of the instrument were preserved. During surgery, the system allowed for the identification of blood vessels and assessment of vascularization (i.e., bowel, bladder and kidney), as well as localization of pelvic lymph nodes. During human specimen evaluation, it was able to distinguish sentinel from non-sentinel lymph nodes. With this introduction of a NIR-fluorescence Click-On sensing detector, a next step is made towards using surgical instruments in the characterization of molecular tissue aspects.
Springer Science and Business Media LLC
Title: Click-on fluorescence detectors: using robotic surgical instruments to characterize molecular tissue aspects
Description:
AbstractFluorescence imaging is increasingly being implemented in surgery.
One of the drawbacks of its application is the need to switch back-and-forth between fluorescence- and white-light-imaging settings and not being able to dissect safely under fluorescence guidance.
The aim of this study was to engineer ‘click-on’ fluorescence detectors that transform standard robotic instruments into molecular sensing devices that enable the surgeon to detect near-infrared (NIR) fluorescence in a white-light setting.
This NIR-fluorescence detector setup was engineered to be press-fitted onto standard forceps instruments of the da Vinci robot.
Following system characterization in a phantom setting (i.
e.
, spectral properties, sensitivity and tissue signal attenuation), the performance with regard to different clinical indocyanine green (ICG) indications (e.
g.
, angiography and lymphatic mapping) was determined via robotic surgery in pigs.
To evaluate in-human applicability, the setup was also used for ICG-containing lymph node specimens from robotic prostate cancer surgery.
The resulting Click-On device allowed for NIR ICG signal identification down to a concentration of 4.
77 × 10–6 mg/ml.
The fully assembled system could be introduced through the trocar and grasping, and movement abilities of the instrument were preserved.
During surgery, the system allowed for the identification of blood vessels and assessment of vascularization (i.
e.
, bowel, bladder and kidney), as well as localization of pelvic lymph nodes.
During human specimen evaluation, it was able to distinguish sentinel from non-sentinel lymph nodes.
With this introduction of a NIR-fluorescence Click-On sensing detector, a next step is made towards using surgical instruments in the characterization of molecular tissue aspects.
Related Results
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct
Introduction
Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
Robotic Surgery
Robotic Surgery
Robots have revolutionized industrial production, from automobiles to pharmaceutical manufacturing, and offer an exciting, novel approach to surgical diseases. Robots employed in s...
Robotic Surgery
Robotic Surgery
Robots have revolutionized industrial production, from automobiles to pharmaceutical manufacturing, and offer an exciting, novel approach to surgical diseases. Robots employed in s...
Endoscopic and Robotic Assisted Transvaginal Hysterectomy: A Feasibility Study
Endoscopic and Robotic Assisted Transvaginal Hysterectomy: A Feasibility Study
Purpose: To perform a feasibility study of the novel approach “transvaginal natural orifice transluminal endoscopic surgery” (vNOTES) with Da-Vinci robotic assistance. The primary ...
PROMs in Total Hip Arthroplasty: A Comparison between Robotic Mako Assisted Versus Manual Technique
PROMs in Total Hip Arthroplasty: A Comparison between Robotic Mako Assisted Versus Manual Technique
Study Purpose: The purpose of this observational study is to evaluate the degree of satisfaction and improvement in quality of life through PROMs (Patient Related Outcome Measures)...
Global Surgery Research: An Overview and the Role of Medical Students and Surgical Trainees in Advancing Global Surgery Research in LMICs
Global Surgery Research: An Overview and the Role of Medical Students and Surgical Trainees in Advancing Global Surgery Research in LMICs
Global surgery research is a critical area of study aimed at enhancing access to safe and effective surgical care for patients in low- and middle-income countries (LMICs). It is es...
Effects of the Multilayer Structure on the Responsivity of Pyroelectric Thin Film Detectors
Effects of the Multilayer Structure on the Responsivity of Pyroelectric Thin Film Detectors
AbstractPyroelectric thin film detectors have advantages of wavelength independent sensitivity, room temperature operation and direct incorporation with integrated circuit amplifie...