How Do Isothermal Calorimeters Age? Repeatability and Reproducibility Across Calorimeter Generations
View through CrossRef
AbstractReliable experimental data is essential for reproducible science. As our lab has access to five instruments of the isothermal (heat conduction) type of different age and product generations, we became interested in the quality of experimental data obtained from the different instruments. Each instrument consists of 8 individual calorimeters. An intralaboratory study of the first 72 hours of cement hydration was conducted with all five instruments. The objective was to determine the differences between the available instruments with respect to repeatability (precision of the 8 calorimeters within one instrument) and reproducibility (precision between the different instruments). The experimental program comprises 120 calorimetry runs in total (5 instruments each with 8 calorimeters, triple measurement performed). Results show that the oldest instrument version with its 8 calorimeters has a stronger higher standard deviation of the heat released after 72 hours compared to the latest state‐of‐the‐art instrument model (246.7 ± 2.7 vs. 232.3 ± 0.5 Jg–1). This is attributed to the technical design differences of the instruments and continuous use over time. In terms of reproducibility, an increasing standard deviation for the heat released from 1 day to 3 days (136.2 ± 2.6 and 236.4 ± 5.9 J g–1) is observed. From our point of view, this test study indicates that isothermal heat conduction calorimetry is a robust and precise measurement method to determine the heat of cement hydration for 72 hours.
Title: How Do Isothermal Calorimeters Age? Repeatability and Reproducibility Across Calorimeter Generations
Description:
AbstractReliable experimental data is essential for reproducible science.
As our lab has access to five instruments of the isothermal (heat conduction) type of different age and product generations, we became interested in the quality of experimental data obtained from the different instruments.
Each instrument consists of 8 individual calorimeters.
An intralaboratory study of the first 72 hours of cement hydration was conducted with all five instruments.
The objective was to determine the differences between the available instruments with respect to repeatability (precision of the 8 calorimeters within one instrument) and reproducibility (precision between the different instruments).
The experimental program comprises 120 calorimetry runs in total (5 instruments each with 8 calorimeters, triple measurement performed).
Results show that the oldest instrument version with its 8 calorimeters has a stronger higher standard deviation of the heat released after 72 hours compared to the latest state‐of‐the‐art instrument model (246.
7 ± 2.
7 vs.
232.
3 ± 0.
5 Jg–1).
This is attributed to the technical design differences of the instruments and continuous use over time.
In terms of reproducibility, an increasing standard deviation for the heat released from 1 day to 3 days (136.
2 ± 2.
6 and 236.
4 ± 5.
9 J g–1) is observed.
From our point of view, this test study indicates that isothermal heat conduction calorimetry is a robust and precise measurement method to determine the heat of cement hydration for 72 hours.
Related Results
Design for Accuracy and Repeatability for Robots Using Taguchi Methods
Design for Accuracy and Repeatability for Robots Using Taguchi Methods
In this paper, we apply Taguchi Methods in the framework of concurrent engineering design for robot-based manufacturing sys tems to obtain higher accuracy and repeatability for rob...
Designing Calorimeter Made from Various Wastes
Designing Calorimeter Made from Various Wastes
<p>A research has been conducted which aims to design a waste-based calorimeter as an alternative to measuring heat for chemistry practicum in schools. Research products deve...
Repeatability and reproducibility of Keratograph 5M corneal topography
Repeatability and reproducibility of Keratograph 5M corneal topography
Background
Corneal topography is an important diagnostic tool and highly repeatable and reproducible topographic devices are essential in eye care practice. Placido ...
New cycle, same old mistakes? Overlapping vs. discrete generations in long-term recurrent selection
New cycle, same old mistakes? Overlapping vs. discrete generations in long-term recurrent selection
Abstract
Background
Recurrent selection is a foundational breeding method for quantitative trait improvement. It typically feat...
Uncertainty evaluation of measurement of orientation repeatability for industrial robots
Uncertainty evaluation of measurement of orientation repeatability for industrial robots
PurposeThis paper aims to propose a reasonable method to evaluate uncertainty of measurement of industrial robots’ orientation repeatability and solve the non-linear problem existi...
KINETICS OF ISOTHERMAL AND NON-ISOTHERMAL PYROLYSIS OF OIL SHALE; pp. 415–424
KINETICS OF ISOTHERMAL AND NON-ISOTHERMAL PYROLYSIS OF OIL SHALE; pp. 415–424
The kinetics of decomposition of Huadian oil shale was studied by non-isoÂthermal and isothermal methods. The non-isothermal weight loss was recorded at different heating rates (2...
Improving Reproducibility in AI Research: Four Mechanisms Adopted by JAIR
Improving Reproducibility in AI Research: Four Mechanisms Adopted by JAIR
Background: Lately, the reproducibility of scientific results has become an increasing worry in the scientific community. Several studies show that artificial intelligence research...
THE ATLAS ELECTROMAGNETIC CALORIMETERS: FEATURES AND PERFORMANCE
THE ATLAS ELECTROMAGNETIC CALORIMETERS: FEATURES AND PERFORMANCE
In this paper it is shown how a sampling electromagnetic calorimeter based on the liquid argon technique satisfies the very demanding requirements of an experiment at the LHC. Sect...