A New Iterative Metal Artifact Reduction Algorithm for Both Energy-Integrating and Photon-Counting CT Systems
View through CrossRef
Objectives
The aim of this study was to introduce and evaluate a new metal artifact reduction framework (iMARv2) that addresses the drawbacks (residual artifacts after correction and user preferences for image quality) associated with the current clinically applied iMAR.
Materials and Methods
A new iMARv2 has been introduced, combining the current iMAR with new modular components to remove residual metal artifacts after image correction. The postcorrection image impression is adjustable with user-selectable strength settings. Phantom scans from an energy-integrating and a photon-counting detector CT were used to assess image quality, including a Gammex phantom and anthropomorphic phantoms. In addition, 36 clinical cases (with metallic implants such as dental fillings, hip replacements, and spinal screws) were reconstructed and evaluated in a blinded and randomized reader study.
Results
The Gammex phantom showed lower HU errors compared with the uncorrected image at almost all iMAR and iMARv2 settings evaluated, with only minor differences between iMAR and the different iMARv2 settings. In addition, the anthropomorphic phantoms showed a trend toward lower errors with higher iMARv2 strength settings. On average, the iMARv2 strength 3 performed best of all the clinical reconstructions evaluated, with a significant increase in diagnostic confidence and decrease in artifacts. All hip and dental cases showed a significant increase in diagnostic confidence and decrease in artifact strength, and the improvements from iMARv2 in the dental cases were significant compared with iMAR. There were no significant improvements in the spine.
Conclusions
This work has introduced and evaluated a new method for metal artifact reduction and demonstrated its utility in routine clinical datasets. The greatest improvements were seen in dental fillings, where iMARv2 significantly improved image quality compared with conventional iMAR.
Ovid Technologies (Wolters Kluwer Health)
Title: A New Iterative Metal Artifact Reduction Algorithm for Both Energy-Integrating and Photon-Counting CT Systems
Description:
Objectives
The aim of this study was to introduce and evaluate a new metal artifact reduction framework (iMARv2) that addresses the drawbacks (residual artifacts after correction and user preferences for image quality) associated with the current clinically applied iMAR.
Materials and Methods
A new iMARv2 has been introduced, combining the current iMAR with new modular components to remove residual metal artifacts after image correction.
The postcorrection image impression is adjustable with user-selectable strength settings.
Phantom scans from an energy-integrating and a photon-counting detector CT were used to assess image quality, including a Gammex phantom and anthropomorphic phantoms.
In addition, 36 clinical cases (with metallic implants such as dental fillings, hip replacements, and spinal screws) were reconstructed and evaluated in a blinded and randomized reader study.
Results
The Gammex phantom showed lower HU errors compared with the uncorrected image at almost all iMAR and iMARv2 settings evaluated, with only minor differences between iMAR and the different iMARv2 settings.
In addition, the anthropomorphic phantoms showed a trend toward lower errors with higher iMARv2 strength settings.
On average, the iMARv2 strength 3 performed best of all the clinical reconstructions evaluated, with a significant increase in diagnostic confidence and decrease in artifacts.
All hip and dental cases showed a significant increase in diagnostic confidence and decrease in artifact strength, and the improvements from iMARv2 in the dental cases were significant compared with iMAR.
There were no significant improvements in the spine.
Conclusions
This work has introduced and evaluated a new method for metal artifact reduction and demonstrated its utility in routine clinical datasets.
The greatest improvements were seen in dental fillings, where iMARv2 significantly improved image quality compared with conventional iMAR.
Related Results
The photon blockade effect of a complete Buck-Sukumar model
The photon blockade effect of a complete Buck-Sukumar model
The Buck-Sukumar (BS) model, with a nonlinear coupling between the atom and the light field, is well defined only when its coupling strength is lower than a critical coupling. Its ...
Dosimetric impact of metal artifact reduction for spinal implants in stereotactic body radiotherapy
Dosimetric impact of metal artifact reduction for spinal implants in stereotactic body radiotherapy
Abstract
Background
Metal artifacts due to spinal implants can affect the accuracy of dose calculation for radiotherapy. However, the dosimetric impact of metal artifact r...
Shadow and photon ring of black hole in asymptotically safe gravity
Shadow and photon ring of black hole in asymptotically safe gravity
In this paper, we focus on discussing the influence of thin disk accretion and asymptotically safe (AS) gravity correction parameter on the shadow and photon ring of black holes. F...
Increasing familiarity with the heartbeat counting task does not affect performance
Increasing familiarity with the heartbeat counting task does not affect performance
Background: Interoception is typically defined as the processing and perception of internal signals. A common evaluation of interoceptive abilities is via the perception of heartbe...
CT Metal Artifact Reduction based on Virtual Generated Artifacts Using Modified pix2pix
CT Metal Artifact Reduction based on Virtual Generated Artifacts Using Modified pix2pix
Abstract
Background: Metal artifacts introduce challenges in image-guided diagnosis or accurate dose calculations. This study aims to reduce metal artifacts from the spinal...
Multi‐Photon Microscopy
Multi‐Photon Microscopy
AbstractIn this series of papers on light microscopy imaging, we have covered the fundamentals of microscopy, super‐resolution microscopy, and lightsheet microscopy. This last revi...
[RETRACTED] Prima Weight Loss Dragons Den UK v1
[RETRACTED] Prima Weight Loss Dragons Den UK v1
[RETRACTED]Prima Weight Loss Dragons Den UK :-Obesity is a not kidding medical issue brought about by devouring an excessive amount of fat, eating terrible food sources, and practi...
[RETRACTED] Prima Weight Loss Dragons Den UK v1
[RETRACTED] Prima Weight Loss Dragons Den UK v1
[RETRACTED]Prima Weight Loss Dragons Den UK :-Obesity is a not kidding medical issue brought about by devouring an excessive amount of fat, eating terrible food sources, and practi...