Nonlinearly scaled prior image‐controlled frequency split for high‐frequency metal artifact reduction in computed tomography
View through CrossRef
AbstractPurposeThis paper introduces a new approach for the dedicated reduction of high‐frequency metal artifacts, which applies a nonlinear scaling (NLS) transfer function on the high‐frequency projection domain to reduce artifacts, while preserving edge information and anatomic detail by incorporating prior image information.MethodsAn NLS function is applied to suppress high‐frequency streak artifacts, but to restrict the correction to metal projections only, scaling is performed in the sinogram domain. Anatomic information should be preserved and is excluded from scaling by incorporating a prior image from tissue classification. The corrected high‐frequency sinogram is reconstructed and combined with the low‐frequency component of a normalized metal artifact reduction (NMAR) image. Scans of different anthropomorphic phantoms were acquired (unilateral hip, bilateral hip, dental implants, and embolization coil). Multiple regions of interest (ROIs) were drawn around the metal implants and hounsfield unit (HU) deviations were analyzed. Clinical data sets including single image slices of dental fillings, a bilateral hip implant, spinal fixation screws, and an aneurysm coil were reconstructed and assessed.ResultsThe prior image‐controlled NLS can remove streak artifacts while preserving anatomic detail within the bone and soft tissue. The qualitative analysis of clinical cases showed a tremendous enhancement within dental fillings and neuro coils, and a significant enhancement within spinal screws or hip implants. The phantom scan measurements support this observation. In all phantom setups, the NLS‐corrected result showed lowest HU derivation and the best visualization of the data.ConclusionsThe prior image‐controlled NLS provides a method to reduce high‐frequency streaks in metal‐corrupted computed tomography (CT) data.
Title: Nonlinearly scaled prior image‐controlled frequency split for high‐frequency metal artifact reduction in computed tomography
Description:
AbstractPurposeThis paper introduces a new approach for the dedicated reduction of high‐frequency metal artifacts, which applies a nonlinear scaling (NLS) transfer function on the high‐frequency projection domain to reduce artifacts, while preserving edge information and anatomic detail by incorporating prior image information.
MethodsAn NLS function is applied to suppress high‐frequency streak artifacts, but to restrict the correction to metal projections only, scaling is performed in the sinogram domain.
Anatomic information should be preserved and is excluded from scaling by incorporating a prior image from tissue classification.
The corrected high‐frequency sinogram is reconstructed and combined with the low‐frequency component of a normalized metal artifact reduction (NMAR) image.
Scans of different anthropomorphic phantoms were acquired (unilateral hip, bilateral hip, dental implants, and embolization coil).
Multiple regions of interest (ROIs) were drawn around the metal implants and hounsfield unit (HU) deviations were analyzed.
Clinical data sets including single image slices of dental fillings, a bilateral hip implant, spinal fixation screws, and an aneurysm coil were reconstructed and assessed.
ResultsThe prior image‐controlled NLS can remove streak artifacts while preserving anatomic detail within the bone and soft tissue.
The qualitative analysis of clinical cases showed a tremendous enhancement within dental fillings and neuro coils, and a significant enhancement within spinal screws or hip implants.
The phantom scan measurements support this observation.
In all phantom setups, the NLS‐corrected result showed lowest HU derivation and the best visualization of the data.
ConclusionsThe prior image‐controlled NLS provides a method to reduce high‐frequency streaks in metal‐corrupted computed tomography (CT) data.
Related Results
Quantitative analysis of metal artifact reduction in total hip arthroplasty using virtual monochromatic imaging and orthopedic metal artifact reduction, a phantom study
Quantitative analysis of metal artifact reduction in total hip arthroplasty using virtual monochromatic imaging and orthopedic metal artifact reduction, a phantom study
Abstract
Objective
To quantify metal artifact reduction using 130 keV virtual monochromatic imaging (VMI) with and without orthopedic metal artifact...
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Objective: To determine the frequency of common chromosomal aberrations in local population idiopathic determine the frequency of common chromosomal aberrations in local population...
Dosimetric impact of metal artifact reduction for spinal implants in stereotactic body radiotherapy
Dosimetric impact of metal artifact reduction for spinal implants in stereotactic body radiotherapy
Abstract
Background
Metal artifacts due to spinal implants can affect the accuracy of dose calculation for radiotherapy. However, the dosimetric impact of metal artifact r...
CT Metal Artifact Reduction based on Virtual Generated Artifacts Using Modified pix2pix
CT Metal Artifact Reduction based on Virtual Generated Artifacts Using Modified pix2pix
Abstract
Background: Metal artifacts introduce challenges in image-guided diagnosis or accurate dose calculations. This study aims to reduce metal artifacts from the spinal...
Small Cell Lung Cancer and Tarlatamab: A Meta-Analysis of Clinical Trials
Small Cell Lung Cancer and Tarlatamab: A Meta-Analysis of Clinical Trials
Abstract
Introduction
Tarlatamab is a Delta-like ligand 3 (DLL3) -directed bispecific T-cell engager recently approved for use in patients with advanced small cell lung cancer (SCL...
A nonlinear scaling‐based normalized metal artifact reduction to reduce low‐frequency artifacts in energy‐integrating and photon‐counting CT
A nonlinear scaling‐based normalized metal artifact reduction to reduce low‐frequency artifacts in energy‐integrating and photon‐counting CT
AbstractBackgroundMetal within the scan plane can cause severe artifacts when reconstructing X‐ray computed tomography (CT) scans. Both in clinical use and recent research, normali...
Characterization of surgically transposed ovaries in integrated PET/CT scan in patients with cervical cancer
Characterization of surgically transposed ovaries in integrated PET/CT scan in patients with cervical cancer
AbstractBackground. The purpose of this study was to determine the ovarian findings on integrated positron emission tomography/computed tomography scans during follow‐up in cervica...
Metal artifact reduction through MVCBCT and kVCT in radiotherapy
Metal artifact reduction through MVCBCT and kVCT in radiotherapy
AbstractThis study proposes a new method for removal of metal artifacts from megavoltage cone beam computed tomography (MVCBCT) and kilovoltage CT (kVCT) images. Both images were c...