Leadless Pacemaker tine fracture and dislocation: two case reports
View through CrossRef
Abstract
The rate of leadless pacemaker implantation is increasing worldwide. To date, there have been few reports of leadless pacemaker dislocation and extraction. This article describes two cases of leadless pacemaker tine fracture and dislocation. One case was an elderly male, and the pacemaker was found to have fallen off to the right pulmonary artery one month after surgery, and a new leadless pacemaker was implanted; the other was an elderly female, and the leadless pacemaker fell off to the right ventricle 9 months after surgery. The leadless pacemaker was removed, and a dual-chamber permanent pacemaker was implanted. After leadless pacemaker implantation, regular programming should be performed to detect abnormalities early and perform clinical intervention.
Springer Science and Business Media LLC
Title: Leadless Pacemaker tine fracture and dislocation: two case reports
Description:
Abstract
The rate of leadless pacemaker implantation is increasing worldwide.
To date, there have been few reports of leadless pacemaker dislocation and extraction.
This article describes two cases of leadless pacemaker tine fracture and dislocation.
One case was an elderly male, and the pacemaker was found to have fallen off to the right pulmonary artery one month after surgery, and a new leadless pacemaker was implanted; the other was an elderly female, and the leadless pacemaker fell off to the right ventricle 9 months after surgery.
The leadless pacemaker was removed, and a dual-chamber permanent pacemaker was implanted.
After leadless pacemaker implantation, regular programming should be performed to detect abnormalities early and perform clinical intervention.
Related Results
Aveir Leadless Pacemaker implantation in pediatric population: a case series
Aveir Leadless Pacemaker implantation in pediatric population: a case series
Abstract
Background
While the Medtronic Micra pacemaker provided a small device for leadless pacemaker implantation, the Aveir d...
P185 FOLLOW–UP OF PATIENTS UNDERGOING POST–EXTRACTION LEADLESS PACEMAKER IMPLANTATION
P185 FOLLOW–UP OF PATIENTS UNDERGOING POST–EXTRACTION LEADLESS PACEMAKER IMPLANTATION
Abstract
Introduction
The implantation of traditional stimulation devices involves numerous complications, in the short and long...
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct
Introduction
Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
A Comparative Study of the Old Tuberculin Tine Test and the PPD-Tine Test
A Comparative Study of the Old Tuberculin Tine Test and the PPD-Tine Test
Forty-one children suspected of having tuberculosis were simultaneously given an intermediate strength Mantoux test, an old tuberculin tine test (OT), and a purified protein deriva...
Fracture Modelling Using Seismic Based Fracture Intensity Volume, a Case Study in Middle East
Fracture Modelling Using Seismic Based Fracture Intensity Volume, a Case Study in Middle East
Abstract
In this paper, a case study in a fractured carbonate reservoir is presented to demonstrate the approach of fracture modeling using fracture intensity vol...
P186 LEADLESS PACEMAKER VERSUS TRANSVENOUS PACEMAKER: THE BEST ALTERNATIVE FOR THE POST–EXTRACTION PATIENT
P186 LEADLESS PACEMAKER VERSUS TRANSVENOUS PACEMAKER: THE BEST ALTERNATIVE FOR THE POST–EXTRACTION PATIENT
Abstract
Introduction
As complications associated with pacemakers or defibrillators implant increase, the number of patients und...
Device–device interference triggered by an abandoned pacemaker: a case report
Device–device interference triggered by an abandoned pacemaker: a case report
Abstract
Background
Cardiac implantable electronic devices (CIEDs) are prone to electromagnetic interference. Common sources inc...
P425Single procedure pace and ablate. evaluation of efficacy and safety comparing three different vascular routes
P425Single procedure pace and ablate. evaluation of efficacy and safety comparing three different vascular routes
Abstract
Funding Acknowledgements
No financial support for this study
Introduc...