Comparison of field olfactometers in a controlled chamber using hydrogen sulfide as the test odorant

View through CrossRef
A standard method for measuring and quantifying odour in the ambient air utilizes a portable odour detecting and measuring device known as a field olfactometer (US Public Health Service Project Grant A-58-541). The field olfactometer dynamically dilutes the ambient air with carbon-filtered air in distinct ratios known as “Dilutions-to-Threshold” dilution factors (D/Ts), i.e. 2, 4, 7, 15, etc. Thirteen US states and several cities in North America currently utilize field olfactometry as a key component of determining compliance to odour regulations and ordinances. A controlled environmental chamber was utilized, with hydrogen sulfide as the known test odorant. A hydrogen sulfide environment was created in this controlled chamber using an Advanced Calibration Designs, Inc. Cal2000 Hydrogen Sulfide Generator. The hydrogen sulfide concentration inside the chamber was monitored using an Arizona Instruments, Inc. Jerome Model 631 H2S Analyzer. When the environmental chamber reached a desired test concentration, test operators entered the chamber. The dilution-to-threshold odour concentration was measured using a Nasal Ranger Field Olfactometer (St Croix Sensory, Inc.) and a Barnebey Sutcliffe Corp. Scentometer. The actual hydrogen sulfide concentration was also measured at the location in the room where the operators were standing while using the two types of field olfactometers. This paper presents a correlation between dilution-to-threshold values (D/T) and hydrogen sulfide ambient concentration. For example, a D/T of 7 corresponds to ambient H2S concentrations of 5.7-15.6 μg/m3 (4-11 ppbv). During this study, no significant difference was found between results obtained using the Scentometer or the Nasal Ranger® (r = 0.82). Also, no significant difference was found between results of multiple Nasal Ranger® users (p = 0.309). The field olfactometers yielded hydrogen sulfide thresholds of 0.7-3.0 μg/m3 (0.5-2.0 ppbv). Laboratory olfactometry yielded comparable thresholds of 0.64-1.3 μg/m3 (0.45-0.9 ppbv). These thresholds are consistent with published values.
Title: Comparison of field olfactometers in a controlled chamber using hydrogen sulfide as the test odorant
Description:
A standard method for measuring and quantifying odour in the ambient air utilizes a portable odour detecting and measuring device known as a field olfactometer (US Public Health Service Project Grant A-58-541).
The field olfactometer dynamically dilutes the ambient air with carbon-filtered air in distinct ratios known as “Dilutions-to-Threshold” dilution factors (D/Ts), i.
e.
2, 4, 7, 15, etc.
Thirteen US states and several cities in North America currently utilize field olfactometry as a key component of determining compliance to odour regulations and ordinances.
A controlled environmental chamber was utilized, with hydrogen sulfide as the known test odorant.
A hydrogen sulfide environment was created in this controlled chamber using an Advanced Calibration Designs, Inc.
Cal2000 Hydrogen Sulfide Generator.
The hydrogen sulfide concentration inside the chamber was monitored using an Arizona Instruments, Inc.
Jerome Model 631 H2S Analyzer.
When the environmental chamber reached a desired test concentration, test operators entered the chamber.
The dilution-to-threshold odour concentration was measured using a Nasal Ranger Field Olfactometer (St Croix Sensory, Inc.
) and a Barnebey Sutcliffe Corp.
Scentometer.
The actual hydrogen sulfide concentration was also measured at the location in the room where the operators were standing while using the two types of field olfactometers.
This paper presents a correlation between dilution-to-threshold values (D/T) and hydrogen sulfide ambient concentration.
For example, a D/T of 7 corresponds to ambient H2S concentrations of 5.
7-15.
6 μg/m3 (4-11 ppbv).
During this study, no significant difference was found between results obtained using the Scentometer or the Nasal Ranger® (r = 0.
82).
Also, no significant difference was found between results of multiple Nasal Ranger® users (p = 0.
309).
The field olfactometers yielded hydrogen sulfide thresholds of 0.
7-3.
0 μg/m3 (0.
5-2.
0 ppbv).
Laboratory olfactometry yielded comparable thresholds of 0.
64-1.
3 μg/m3 (0.
45-0.
9 ppbv).
These thresholds are consistent with published values.

Related Results

Safe Boundaries of High-Temperature Fracturing Fluids
Safe Boundaries of High-Temperature Fracturing Fluids
Abstract In successful hydraulic fracturing operations, fracturing fluids should possess sufficient viscosity in order to transmit high pumping pressure downhole. Th...
Study of High-sulfur Natural Gas Field Water Treatment
Study of High-sulfur Natural Gas Field Water Treatment
Abstract High-sulfide gas field water with more than 100mg/L hydrogen sulfide account for 54.5% of gas wells in the Southern and Eastern gas fields of Sichuan, Ch...
Contrasting Geochemistry of Apatite from Peridotites and Sulfide Ores of the Jinchuan Ni-Cu Sulfide Deposit, NW China
Contrasting Geochemistry of Apatite from Peridotites and Sulfide Ores of the Jinchuan Ni-Cu Sulfide Deposit, NW China
Abstract Apatite is present within both the hosting lherzolite and sulfide ore at the Jinchuan magmatic Ni-Cu sulfide deposit of northwest China. Apatite grains with...
Elucidating hydrogen-solid interactions using computational modeling
Elucidating hydrogen-solid interactions using computational modeling
Hydrogen has significant chemical utility, both as a synthetic reagent and as an energy carrier. As the world moves away from fossil fuels being the predominant energy carrier, the...
Sulfate/sulfide removal from wastewater by lab-scale microbial fuel cell
Sulfate/sulfide removal from wastewater by lab-scale microbial fuel cell
Sulfate/sulfide-containing wastewater is a widespread environmental contaminant resulting from human activities. These pollutants have negative impact on natural ecosystems and hum...
Provocative Tests in Diagnosis of Thoracic Outlet Syndrome: A Narrative Review
Provocative Tests in Diagnosis of Thoracic Outlet Syndrome: A Narrative Review
Abstract Thoracic outlet syndrome (TOS) is a group of conditions caused by the compression of the neurovascular bundle within the thoracic outlet. It is classified into three main ...

Back to Top