Agreement between paired blood gas values in samples transported either by a pneumatic system or by human courier
View through CrossRef
Abstract
Background: Rapid accurate assessment of metabolic derangements is crucial in the critically ill. We evaluated if arterial blood gas (ABG) samples transported through a pneumatic tube system (PTS) agreed with values transported by a human courier.
Methods: In this prospective study of 50-paired ABG samples, the couriered reference ABG was compared with those transported by PTS. Agreement was summarised by the mean difference with 95% limits of agreement (LOA) and Lin's concordance correlation (pc).
Results: The mean (±SD) time from sampling to analysis was 35.7±23.2 (courier) and 38.6±22.1 (PTS) minutes. Agreement was good between courier and PTS for pH, PaCO2, bicarbonate, oxygen saturation and PaO2 values (pc>0.97). Although the mean difference in PaO2 values between PTS and courier was small (–0.9 mm Hg) and the agreement was good, individual differences were clinically significant (95% LOA –40.8 to 39.0). For PaO2 <160 mm Hg, analysis of PTS samples yielded erroneously high PaO2 values and vice versa for PaO2>160 mm Hg compared to manual courier. This suggested exaggerated oxygen movement between the blood sample and air in the PTS.
Conclusions: In this study, analysis of samples transported through the PTS resulted in clinically unacceptable PaO2 values. Delay in transport and analysis of ABG samples should be avoided and samples transported manually if they cannot be assessed on-site.
Title: Agreement between paired blood gas values in samples transported either by a pneumatic system or by human courier
Description:
Abstract
Background: Rapid accurate assessment of metabolic derangements is crucial in the critically ill.
We evaluated if arterial blood gas (ABG) samples transported through a pneumatic tube system (PTS) agreed with values transported by a human courier.
Methods: In this prospective study of 50-paired ABG samples, the couriered reference ABG was compared with those transported by PTS.
Agreement was summarised by the mean difference with 95% limits of agreement (LOA) and Lin's concordance correlation (pc).
Results: The mean (±SD) time from sampling to analysis was 35.
7±23.
2 (courier) and 38.
6±22.
1 (PTS) minutes.
Agreement was good between courier and PTS for pH, PaCO2, bicarbonate, oxygen saturation and PaO2 values (pc>0.
97).
Although the mean difference in PaO2 values between PTS and courier was small (–0.
9 mm Hg) and the agreement was good, individual differences were clinically significant (95% LOA –40.
8 to 39.
0).
For PaO2 <160 mm Hg, analysis of PTS samples yielded erroneously high PaO2 values and vice versa for PaO2>160 mm Hg compared to manual courier.
This suggested exaggerated oxygen movement between the blood sample and air in the PTS.
Conclusions: In this study, analysis of samples transported through the PTS resulted in clinically unacceptable PaO2 values.
Delay in transport and analysis of ABG samples should be avoided and samples transported manually if they cannot be assessed on-site.
Related Results
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Abstract
Introduction
The exact manner in which large language models (LLMs) will be integrated into pathology is not yet fully comprehended. This study examines the accuracy, bene...
Efficiency of Courier Service Companies in Malaysia Using Data Envelopment Analysis (DEA) Models
Efficiency of Courier Service Companies in Malaysia Using Data Envelopment Analysis (DEA) Models
Many organisations in the service industry encounter the issue of unequal efficiency results. Many large corporations, such as banks, hotels and courier companies, are concerned...
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Impact of Common Anticoagulants on Complete Blood Count Parameters Among Humans
Impact of Common Anticoagulants on Complete Blood Count Parameters Among Humans
Abstract
Introduction
Among the most frequently used anticoagulants in hematological testing are tetra-acetic acid (EDTA), sodium citrate, and sodium heparin. However, there is a n...
[RETRACTED] Guardian Blood Balance Australia- Reviews - Guardian Botanicals Blood Balance [AU] SCAM ALERT! Read Real Critical Reports.. Price in Australia v1
[RETRACTED] Guardian Blood Balance Australia- Reviews - Guardian Botanicals Blood Balance [AU] SCAM ALERT! Read Real Critical Reports.. Price in Australia v1
[RETRACTED]Guardian Blood Balance Australia Reviews - Diabetes and blood sugar are some of the common problems that are attacking so many adult individuals nowadays. Obesity is t...
Development of android-based apps for courier service management
Development of android-based apps for courier service management
E-commerce has grown exponentially over the years. The growth has been characterized by strong consumer demands and the increasing number of various products available online. This...
Comparisons of Pore Structure for Unconventional Tight Gas, Coalbed Methane and Shale Gas Reservoirs
Comparisons of Pore Structure for Unconventional Tight Gas, Coalbed Methane and Shale Gas Reservoirs
Extended abstract
Tight sands gas, coalbed methane and shale gas are three kinds of typical unconventional natural gas. With the decrease of conventional oil and gas...
Critical Gas Saturation During Depressurisation and its Importance in the Brent Field
Critical Gas Saturation During Depressurisation and its Importance in the Brent Field
Critical Gas Saturation During Depressurisation and its Importance in the Brent Field.
Abstract
After some 20 years of pressure ...