A Comparative Evaluation of Performance of Sysmex XN 3000 and Horiba Yumizen H2500 Automated Complete Blood Count Analysers

View through CrossRef
AbstractModern automated laboratory haematology analysers use various methods to measure different haematological parameters. These parameters are useful in the diagnostic and clinical interpretation of patient symptoms. So, it is very important to compare the performance of different analysers measuring the same parameter. Hence, a comparison of complete blood counts analysed by Sysmex XN 3000 and Horiba Yumizen H2500 was performed. Total 296 EDTA anti-coagulated blood samples were processed in both the analysers in duplicate within 4 h of collection. The white blood cell count, red blood cell count, erythrocyte indices, differential leukocyte count, platelet count and platelet indices and reticulocyte count were compared. A good level of correlation and agreement between different parameters were obtained. A strong correlation was observed (r > 0.9) between Sysmex XN 3000 and Yumizen H2500 for WBC (0.997), RBC (0.997), Haemoglobin (0.999), haematocrit (0.974), MCV (0.902), MCH (0.99),, platelet count by impedance (0.989), mean platelet volume (0.954), plateletcrit (0.971), platelet distribution width (PDW) (0.916), neutrophils (0.997), lymphocytes (0.989), monocytes (0.943), and eosinophils (0.991) counts. A moderate correlation was observed for RDW-CV (0.75). The basophils count showed poor correlation (r < 0.5) possibly because of sample selection with mostly low basophils count. An acceptable bias was observed for most of the parameters like WBC, RBC, Haemoglobin, Haematocrit, platelet counts, neutrophils, lymphocytes, eosinophils and monocytes. The studied instruments ensured satisfactory interchangeability except for few parameters, thus facilitate substitution of one analyser by another without affecting the clinical decision making.
Title: A Comparative Evaluation of Performance of Sysmex XN 3000 and Horiba Yumizen H2500 Automated Complete Blood Count Analysers
Description:
AbstractModern automated laboratory haematology analysers use various methods to measure different haematological parameters.
These parameters are useful in the diagnostic and clinical interpretation of patient symptoms.
So, it is very important to compare the performance of different analysers measuring the same parameter.
Hence, a comparison of complete blood counts analysed by Sysmex XN 3000 and Horiba Yumizen H2500 was performed.
Total 296 EDTA anti-coagulated blood samples were processed in both the analysers in duplicate within 4 h of collection.
The white blood cell count, red blood cell count, erythrocyte indices, differential leukocyte count, platelet count and platelet indices and reticulocyte count were compared.
A good level of correlation and agreement between different parameters were obtained.
A strong correlation was observed (r > 0.
9) between Sysmex XN 3000 and Yumizen H2500 for WBC (0.
997), RBC (0.
997), Haemoglobin (0.
999), haematocrit (0.
974), MCV (0.
902), MCH (0.
99),, platelet count by impedance (0.
989), mean platelet volume (0.
954), plateletcrit (0.
971), platelet distribution width (PDW) (0.
916), neutrophils (0.
997), lymphocytes (0.
989), monocytes (0.
943), and eosinophils (0.
991) counts.
A moderate correlation was observed for RDW-CV (0.
75).
The basophils count showed poor correlation (r < 0.
5) possibly because of sample selection with mostly low basophils count.
An acceptable bias was observed for most of the parameters like WBC, RBC, Haemoglobin, Haematocrit, platelet counts, neutrophils, lymphocytes, eosinophils and monocytes.
The studied instruments ensured satisfactory interchangeability except for few parameters, thus facilitate substitution of one analyser by another without affecting the clinical decision making.

Related Results

The Effect of Sample Storage on Impedance and Optical Platelet Counts
The Effect of Sample Storage on Impedance and Optical Platelet Counts
Abstract Introduction: Processing blood samples for full blood count may be delayed for a multitude of reasons. There is little published information on the effect o...
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Primerjalna književnost na prelomu tisočletja
Primerjalna književnost na prelomu tisočletja
In a comprehensive and at times critical manner, this volume seeks to shed light on the development of events in Western (i.e., European and North American) comparative literature ...
Tracing Hematological Shifts in Pregnancy: How Anemia and Thrombocytopenia Evolve Across Trimesters
Tracing Hematological Shifts in Pregnancy: How Anemia and Thrombocytopenia Evolve Across Trimesters
Abstract Introduction Given pregnancy's significant impact on hematological parameters, monitoring these changes across trimesters is crucial. This study aims to evaluate hematolog...
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Abstract The rapid growth of open access publishing (OAP) has significantly improved the accessibility and dissemination of scientific knowledge. However, this expansion has also c...

Back to Top