Accuracy of Consumer Wearable Heart Rate Measurement During an Ecologically Valid 24-Hour Period: Intraindividual Validation Study (Preprint)
View through CrossRef
BACKGROUND
Wrist-worn smart watches and fitness monitors (ie, wearables) have become widely adopted by consumers and are gaining increased attention from researchers for their potential contribution to naturalistic digital measurement of health in a scalable, mobile, and unobtrusive way. Various studies have examined the accuracy of these devices in controlled laboratory settings (eg, treadmill and stationary bike); however, no studies have investigated the heart rate accuracy of wearables during a continuous and ecologically valid 24-hour period of actual consumer device use conditions.
OBJECTIVE
The aim of this study was to determine the heart rate accuracy of 2 popular wearable devices, the Apple Watch 3 and Fitbit Charge 2, as compared with the gold standard reference method, an ambulatory electrocardiogram (ECG), during consumer device use conditions in an individual. Data were collected across 5 daily conditions, including sitting, walking, running, activities of daily living (ADL; eg, chores, brushing teeth), and sleeping.
METHODS
One participant, (first author; 29-year-old Caucasian male) completed a 24-hour ecologically valid protocol by wearing 2 popular wrist wearable devices (Apple Watch 3 and Fitbit Charge 2). In addition, an ambulatory ECG (Vrije Universiteit Ambulatory Monitoring System) was used as the gold standard reference method, which resulted in the collection of 102,740 individual heartbeats. A single-subject design was used to keep all variables constant except for wearable devices while providing a rapid response design to provide initial assessment of wearable accuracy for allowing the research cycle to keep pace with technological advancements. Accuracy of these devices compared with the gold standard ECG was assessed using mean error, mean absolute error, and mean absolute percent error. These data were supplemented with Bland-Altman analyses and concordance class correlation to assess agreement between devices.
RESULTS
The Apple Watch 3 and Fitbit Charge 2 were generally highly accurate across the 24-hour condition. Specifically, the Apple Watch 3 had a mean difference of −1.80 beats per minute (bpm), a mean absolute error percent of 5.86%, and a mean agreement of 95% when compared with the ECG across 24 hours. The Fitbit Charge 2 had a mean difference of −3.47 bpm, a mean absolute error of 5.96%, and a mean agreement of 91% when compared with the ECG across 24 hours. These findings varied by condition.
CONCLUSIONS
The Apple Watch 3 and the Fitbit Charge 2 provided acceptable heart rate accuracy (<±10%) across the 24 hour and during each activity, except for the Apple Watch 3 during the daily activities condition. Overall, these findings provide preliminary support that these devices appear to be useful for implementing ambulatory measurement of cardiac activity in research studies, especially those where the specific advantages of these methods (eg, scalability, low participant burden) are particularly suited to the population or research question.
Title: Accuracy of Consumer Wearable Heart Rate Measurement During an Ecologically Valid 24-Hour Period: Intraindividual Validation Study (Preprint)
Description:
BACKGROUND
Wrist-worn smart watches and fitness monitors (ie, wearables) have become widely adopted by consumers and are gaining increased attention from researchers for their potential contribution to naturalistic digital measurement of health in a scalable, mobile, and unobtrusive way.
Various studies have examined the accuracy of these devices in controlled laboratory settings (eg, treadmill and stationary bike); however, no studies have investigated the heart rate accuracy of wearables during a continuous and ecologically valid 24-hour period of actual consumer device use conditions.
OBJECTIVE
The aim of this study was to determine the heart rate accuracy of 2 popular wearable devices, the Apple Watch 3 and Fitbit Charge 2, as compared with the gold standard reference method, an ambulatory electrocardiogram (ECG), during consumer device use conditions in an individual.
Data were collected across 5 daily conditions, including sitting, walking, running, activities of daily living (ADL; eg, chores, brushing teeth), and sleeping.
METHODS
One participant, (first author; 29-year-old Caucasian male) completed a 24-hour ecologically valid protocol by wearing 2 popular wrist wearable devices (Apple Watch 3 and Fitbit Charge 2).
In addition, an ambulatory ECG (Vrije Universiteit Ambulatory Monitoring System) was used as the gold standard reference method, which resulted in the collection of 102,740 individual heartbeats.
A single-subject design was used to keep all variables constant except for wearable devices while providing a rapid response design to provide initial assessment of wearable accuracy for allowing the research cycle to keep pace with technological advancements.
Accuracy of these devices compared with the gold standard ECG was assessed using mean error, mean absolute error, and mean absolute percent error.
These data were supplemented with Bland-Altman analyses and concordance class correlation to assess agreement between devices.
RESULTS
The Apple Watch 3 and Fitbit Charge 2 were generally highly accurate across the 24-hour condition.
Specifically, the Apple Watch 3 had a mean difference of −1.
80 beats per minute (bpm), a mean absolute error percent of 5.
86%, and a mean agreement of 95% when compared with the ECG across 24 hours.
The Fitbit Charge 2 had a mean difference of −3.
47 bpm, a mean absolute error of 5.
96%, and a mean agreement of 91% when compared with the ECG across 24 hours.
These findings varied by condition.
CONCLUSIONS
The Apple Watch 3 and the Fitbit Charge 2 provided acceptable heart rate accuracy (<±10%) across the 24 hour and during each activity, except for the Apple Watch 3 during the daily activities condition.
Overall, these findings provide preliminary support that these devices appear to be useful for implementing ambulatory measurement of cardiac activity in research studies, especially those where the specific advantages of these methods (eg, scalability, low participant burden) are particularly suited to the population or research question.
Related Results
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Aim/Purpose: The aim of this article is to make a case of the role of validation in doctoral education. The purpose is to detail findings from three studies which explore PhD stude...
Materialism and Environmental Knowledge as a Mediator for Relationships between Religiosity and Ethical Consumption
Materialism and Environmental Knowledge as a Mediator for Relationships between Religiosity and Ethical Consumption
ABSTRACTOn a global and regional scale, Indonesia has one of the least environmentally sustainable economies in the Asia-Pacific region. Consumption is one of the key factors contr...
Analisis Produktivitas Alat Berat Pada Pekerjaan Peningkatan Jalan Batas Kota Pegaf – Batas Kabupaten Manokwari Selatan
Analisis Produktivitas Alat Berat Pada Pekerjaan Peningkatan Jalan Batas Kota Pegaf – Batas Kabupaten Manokwari Selatan
Roads in Manokwari City still need a lot to be improved, especially on the Pegaf city boundary road - South Manokwari district, because this road connects the border between the Ci...
Heart failure monitoring using the wearable cardioverter defibrillator in patients with newly diagnosed heart failure
Heart failure monitoring using the wearable cardioverter defibrillator in patients with newly diagnosed heart failure
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
...
Dynamic Scheduled Access Medium Access Control for Emerging Wearable Applications
Dynamic Scheduled Access Medium Access Control for Emerging Wearable Applications
Context and MotivationWearable technology is emerging as one of the key enablers for the internet of everything (IoE). The technology is getting mature by every day with more appli...
50 Biological Variability of Soluble ST2 in Stable Chronic Heart Failure
50 Biological Variability of Soluble ST2 in Stable Chronic Heart Failure
Purpose
The evidence for serially monitoring patients with the B-Type Natriuretic Peptides (BNP/NT-proBNP) to detect deterioration in heart failure symptoms has s...
Wearable Sensor Application for Integrated Early Warning and Health Surveillance
Wearable Sensor Application for Integrated Early Warning and Health Surveillance
Objective: The Wearable Sensor Application developed by Pacific Northwest National Laboratory (PNNL) provides an early warning system for stressors to individual and group health u...
AN ALTERNATIVE WAY OF RESOLVING DISPUTE WITH CONSUMER PARTICIPATION: EXPERIENCE FOR UKRAINE
AN ALTERNATIVE WAY OF RESOLVING DISPUTE WITH CONSUMER PARTICIPATION: EXPERIENCE FOR UKRAINE
Background. The Government-endorsed Concept of State Policy for Consumer Protection by 2020 recognizes the need to reform a consumer protection system that has lost its control and...