Development, Construction, and Evaluation of an Alternative Dosimetry Phantom for Computed Tomography
View through CrossRef
This article aims to present the development, construction, and evaluation of an alternative computed tomography dose index (CTDI) phantom. Epoxy resin was mixed with an iodine-based contrast agent to produce radiological characteristics resembling polymethyl methacrylate (PMMA) as a standard CTDI phantom. As a preliminary study, testing was carried out using computed tomography images (80 and 120 kVp) on 12 variations of epoxy-iodine resin mixtures to obtain relative electron density (ρe
) values and effective atomic numbers (Zeff
) of the samples. The alternative CTDI phantoms were then constructed with a resin-iodine mixture using iodine concentrations that yield on closest ρe
and Zeff
values to those of PMMA. The evaluation was carried out by comparing dose measurement results at various energies between the alternative phantom and the International Electrotechnical Commission-standard CTDI phantom. At a concentration of 0.46%, the epoxy resin has ρe
and Zeff
with a deviation against PMMA of 0.12% and 1.58%, respectively, so that composition was chosen for the alternative CTDI phantom construction. The average dose discrepancy values were 5% and 1%, respectively, for the head and body phantoms in the tested tube voltages of 80 kVp, 100 kVp, 120 kVp, and 135 kVp. The Student’s t-test result between the alternative and the standard phantoms also showed P < 0.05, indicating the comparability of the alternative CTDI phantom with the standard CTDI phantom.
Title: Development, Construction, and Evaluation of an Alternative Dosimetry Phantom for Computed Tomography
Description:
This article aims to present the development, construction, and evaluation of an alternative computed tomography dose index (CTDI) phantom.
Epoxy resin was mixed with an iodine-based contrast agent to produce radiological characteristics resembling polymethyl methacrylate (PMMA) as a standard CTDI phantom.
As a preliminary study, testing was carried out using computed tomography images (80 and 120 kVp) on 12 variations of epoxy-iodine resin mixtures to obtain relative electron density (ρe
) values and effective atomic numbers (Zeff
) of the samples.
The alternative CTDI phantoms were then constructed with a resin-iodine mixture using iodine concentrations that yield on closest ρe
and Zeff
values to those of PMMA.
The evaluation was carried out by comparing dose measurement results at various energies between the alternative phantom and the International Electrotechnical Commission-standard CTDI phantom.
At a concentration of 0.
46%, the epoxy resin has ρe
and Zeff
with a deviation against PMMA of 0.
12% and 1.
58%, respectively, so that composition was chosen for the alternative CTDI phantom construction.
The average dose discrepancy values were 5% and 1%, respectively, for the head and body phantoms in the tested tube voltages of 80 kVp, 100 kVp, 120 kVp, and 135 kVp.
The Student’s t-test result between the alternative and the standard phantoms also showed P < 0.
05, indicating the comparability of the alternative CTDI phantom with the standard CTDI phantom.
Related Results
Technologies for retrospective radiation dosimetry
Technologies for retrospective radiation dosimetry
Abstract
Radiation dosimetry is an important task for assessing the biological damages created in human being due to ionising radiation exposure. Ionising radiation ...
Rd-phantom and RD-Ext-phantom morphisms
Rd-phantom and RD-Ext-phantom morphisms
A morphism f of left R-modules is called an RD-phantom morphism if the
induced morphism Tor1(R=aR,f)=0 for any a ? R. Similarly, a morphism g
of left R-modules is said to be ...
Analysis of the relevance between computed tomography characterization and pathology of pulmonary ground-glass nodules with different pathology types
Analysis of the relevance between computed tomography characterization and pathology of pulmonary ground-glass nodules with different pathology types
Background: In this study, we aimed to analyze the relevance between computed tomography characterization and pathology of pulmonary ground-glass nodules with different pathology t...
Study of the Response and Calibration Procedures of LiF:Mg, Ti TLD-100 for the Dosimetry of Photon Beam
Study of the Response and Calibration Procedures of LiF:Mg, Ti TLD-100 for the Dosimetry of Photon Beam
To guarantee quality assurance and quality control of the received radiation dose and treatment level, Thermoluminescent Dosimeter (TLD) is frequently employed in personal control ...
Multicenter Repeatability Study of a Novel Quantitative Diffusion Kurtosis Imaging Phantom
Multicenter Repeatability Study of a Novel Quantitative Diffusion Kurtosis Imaging Phantom
Quantitative kurtosis phantoms are sought by multicenter clinical trials to establish accuracy and precision of quantitative imaging biomarkers on the basis of diffusion kurtosis i...
Dosimetric verification of radiotherapy treatment planning system at TMSS Cancer Center, Bogura, Bangladesh
Dosimetric verification of radiotherapy treatment planning system at TMSS Cancer Center, Bogura, Bangladesh
Authentication of TPS is the component of the QA program for RTPS and planning process. So, verification of dose calculation algorithm is essential to measure the efficiency of the...
Reactor Dosimetry and Nuclear Reactor Regulation
Reactor Dosimetry and Nuclear Reactor Regulation
An important function of reactor dosimetry is the verification of fluence values used in the calculation of pressure vessel embrittlement and its impact on (a) the fracture toughne...
Development of end-to-end test phantom for stereotactic radiosurgery using alanine dosimeter
Development of end-to-end test phantom for stereotactic radiosurgery using alanine dosimeter
Alanine dosimeters are generally used in high-dose dosimetry, while accurate and precise dose delivery verification procedures are important in stereotactic radiosurgery (SRS). Hen...