e0565 Analysis of inappropriate therapy in the patients implantable cardioverter defibrillators
View through CrossRef
Objective
Observed the efficacy of the patients with implantable cardioverter defibrillators (ICD). Sought to identify the reasons of inappropriate shocks, to minimise the risk of inappropriate shocks.
Methods
Followed up nineteen patients with ICD. Stored ICD electrograms from all shocks episodes were analysed to distinguish appropriate and inappropriate shocks according to history of tachycardia. Any ICD therapy not delivered for VT or VF was deemed inappropriate shocks therapy. It can be considered as inappropriate that ICD firing for sinus tachycardia, supraventricular tachycardia, atrial fibrillation, atrial flutter, and nonsustained ventricular tachycaidia.
Result
There is no shock episode in ten patients (52.63%). Appropriate shocks occurred in 6 patients (31.58%) Inappropriate shocks occurred in three patients (15.79%). Inappropriate shock episodes constituted 4 of 11 total shock episodes. We presented three cases of inappropriate shocks due to atrial fibrillation, sinus tachycardia and high frequency noise.
Conclusions
Inappropriate shocks occurred commonly in our cases. Varied reason induced the inappropriate shocks. Programing the device parameter and choosing proper discriminators can minimised rate of inappropriate shocks.
Title: e0565 Analysis of inappropriate therapy in the patients implantable cardioverter defibrillators
Description:
Objective
Observed the efficacy of the patients with implantable cardioverter defibrillators (ICD).
Sought to identify the reasons of inappropriate shocks, to minimise the risk of inappropriate shocks.
Methods
Followed up nineteen patients with ICD.
Stored ICD electrograms from all shocks episodes were analysed to distinguish appropriate and inappropriate shocks according to history of tachycardia.
Any ICD therapy not delivered for VT or VF was deemed inappropriate shocks therapy.
It can be considered as inappropriate that ICD firing for sinus tachycardia, supraventricular tachycardia, atrial fibrillation, atrial flutter, and nonsustained ventricular tachycaidia.
Result
There is no shock episode in ten patients (52.
63%).
Appropriate shocks occurred in 6 patients (31.
58%) Inappropriate shocks occurred in three patients (15.
79%).
Inappropriate shock episodes constituted 4 of 11 total shock episodes.
We presented three cases of inappropriate shocks due to atrial fibrillation, sinus tachycardia and high frequency noise.
Conclusions
Inappropriate shocks occurred commonly in our cases.
Varied reason induced the inappropriate shocks.
Programing the device parameter and choosing proper discriminators can minimised rate of inappropriate shocks.
Related Results
The effects of age on quality of life in implantable cardioverter defibrillator recipients
The effects of age on quality of life in implantable cardioverter defibrillator recipients
Background. The implantable cardioverter defibrillator shows superiority over conventional pharmacological therapy. The implantable cardioverter defibrillator has been implanted w...
Urrets-Zavalia syndrome following implantable collamer lens (ICL) implantation: a case report and review of the literature
Urrets-Zavalia syndrome following implantable collamer lens (ICL) implantation: a case report and review of the literature
Abstract
Background
Urrets-Zavalia syndrome is a condition that arises after eye surgery, often linked to increased intraocular pressure followin...
Small Cell Lung Cancer and Tarlatamab: A Meta-Analysis of Clinical Trials
Small Cell Lung Cancer and Tarlatamab: A Meta-Analysis of Clinical Trials
Abstract
Introduction
Tarlatamab is a Delta-like ligand 3 (DLL3) -directed bispecific T-cell engager recently approved for use in patients with advanced small cell lung cancer (SCL...
Abstract 4373072: Evaluating the Outcomes and Complications of Implantable Cardioverter Defibrillator Therapy in Hypertrophic Cardiomyopathy
Abstract 4373072: Evaluating the Outcomes and Complications of Implantable Cardioverter Defibrillator Therapy in Hypertrophic Cardiomyopathy
Background:
Implantable cardioverter-defibrillators (ICDs) are critical for preventing sudden cardiac death in patients with hypertrophic cardiomyopathy (HCM). ...
Implantable antennas for biomedical applications: a systematic review
Implantable antennas for biomedical applications: a systematic review
AbstractThis review presents an in-depth examination of implantable antennas for various biomedical purposes. The development of implantable antennas, including their designs, mate...
Pembrolizumab and Sarcoma: A meta-analysis
Pembrolizumab and Sarcoma: A meta-analysis
Abstract
Introduction: Pembrolizumab is a monoclonal antibody that promotes antitumor immunity. This study presents a systematic review and meta-analysis of the efficacy and safety...
Abstract 4370453: Incidence and Predictors of Shock in Patients with Noncompaction Cardiomyopathy and Implantable Cardioverter-Defibrillators: A Multicenter Retrospective Study.
Abstract 4370453: Incidence and Predictors of Shock in Patients with Noncompaction Cardiomyopathy and Implantable Cardioverter-Defibrillators: A Multicenter Retrospective Study.
Background:
Noncompaction cardiomyopathy (NCCM) is a distinct myocardial disorder associated with a high risk of ventricular arrhythmias and sudden cardiac deat...
Abstract 11923: Ventricular Parasystole: An Under-Recognized Risk Factor for Ventricular Fibrillation in Patients With Implantable Cardioverter-Defibrillators
Abstract 11923: Ventricular Parasystole: An Under-Recognized Risk Factor for Ventricular Fibrillation in Patients With Implantable Cardioverter-Defibrillators
Introduction:
Ventricular parasystole is an uncommon characteristic of ventricular ectopic beats characterized by cyclical depolarization that is independent of the dom...