e0565 Analysis of inappropriate therapy in the patients implantable cardioverter defibrillators
View through CrossRef
Objective
Observed the efficacy of the patients with implantable cardioverter defibrillators (ICD). Sought to identify the reasons of inappropriate shocks, to minimise the risk of inappropriate shocks.
Methods
Followed up nineteen patients with ICD. Stored ICD electrograms from all shocks episodes were analysed to distinguish appropriate and inappropriate shocks according to history of tachycardia. Any ICD therapy not delivered for VT or VF was deemed inappropriate shocks therapy. It can be considered as inappropriate that ICD firing for sinus tachycardia, supraventricular tachycardia, atrial fibrillation, atrial flutter, and nonsustained ventricular tachycaidia.
Result
There is no shock episode in ten patients (52.63%). Appropriate shocks occurred in 6 patients (31.58%) Inappropriate shocks occurred in three patients (15.79%). Inappropriate shock episodes constituted 4 of 11 total shock episodes. We presented three cases of inappropriate shocks due to atrial fibrillation, sinus tachycardia and high frequency noise.
Conclusions
Inappropriate shocks occurred commonly in our cases. Varied reason induced the inappropriate shocks. Programing the device parameter and choosing proper discriminators can minimised rate of inappropriate shocks.
Title: e0565 Analysis of inappropriate therapy in the patients implantable cardioverter defibrillators
Description:
Objective
Observed the efficacy of the patients with implantable cardioverter defibrillators (ICD).
Sought to identify the reasons of inappropriate shocks, to minimise the risk of inappropriate shocks.
Methods
Followed up nineteen patients with ICD.
Stored ICD electrograms from all shocks episodes were analysed to distinguish appropriate and inappropriate shocks according to history of tachycardia.
Any ICD therapy not delivered for VT or VF was deemed inappropriate shocks therapy.
It can be considered as inappropriate that ICD firing for sinus tachycardia, supraventricular tachycardia, atrial fibrillation, atrial flutter, and nonsustained ventricular tachycaidia.
Result
There is no shock episode in ten patients (52.
63%).
Appropriate shocks occurred in 6 patients (31.
58%) Inappropriate shocks occurred in three patients (15.
79%).
Inappropriate shock episodes constituted 4 of 11 total shock episodes.
We presented three cases of inappropriate shocks due to atrial fibrillation, sinus tachycardia and high frequency noise.
Conclusions
Inappropriate shocks occurred commonly in our cases.
Varied reason induced the inappropriate shocks.
Programing the device parameter and choosing proper discriminators can minimised rate of inappropriate shocks.
Related Results
The Impact of IL28B Gene Polymorphisms on Drug Responses
The Impact of IL28B Gene Polymorphisms on Drug Responses
To achieve high therapeutic efficacy in the patient, information on pharmacokinetics, pharmacodynamics, and pharmacogenetics is required. With the development of science and techno...
Outcomes of implantable cardioverter-defibrillators in patients with a left ventricular assist device: a systematic review and meta-analysis
Outcomes of implantable cardioverter-defibrillators in patients with a left ventricular assist device: a systematic review and meta-analysis
Abstract
Introduction
Implantable Cardioverter-Defibrillators (ICDs) improve survival in advanced heart failure with reduced eje...
Shared decision making in implantable cardioverter-defibrillator patients
Shared decision making in implantable cardioverter-defibrillator patients
Abstract
Background
Implantable cardioverter-defibrillators (ICD) are effective in the prevention of sudden cardiac death and tr...
Edoxaban and Cancer-Associated Venous Thromboembolism: A Meta-analysis of Clinical Trials
Edoxaban and Cancer-Associated Venous Thromboembolism: A Meta-analysis of Clinical Trials
Abstract
Introduction
Cancer patients face a venous thromboembolism (VTE) risk that is up to 50 times higher compared to individuals without cancer. In 2010, direct oral anticoagul...
Carcinoma ex Pleomorphic Adenoma: A Case Series and Literature Review
Carcinoma ex Pleomorphic Adenoma: A Case Series and Literature Review
Abstract
Introduction
Carcinoma ex pleomorphic adenoma (CXPA) is a rare malignant salivary gland tumor that can lead to severe complications and carries a risk of distant metastasi...
Ventricular tachycardia initiation patterns on electrograms of implantable defibrillator devices
Ventricular tachycardia initiation patterns on electrograms of implantable defibrillator devices
Abstract
Introduction
Sudden cardiac death is an unexpected death from a cardiac origin attributed by arrhythmias such as ventri...
History of the Development of Automated External Defibrillators
History of the Development of Automated External Defibrillators
This chapter is structured as a historical overview of the history of the development of defibrillators and the most prominent personalities who contributed to the development of t...
International Breast Cancer Study Group (IBCSG)
International Breast Cancer Study Group (IBCSG)
This section provides current contact details and a summary of recent or ongoing clinical trials being coordinated by International Breast Cancer Study Group (IBCSG). Clinical tria...