e0565 Analysis of inappropriate therapy in the patients implantable cardioverter defibrillators

View through CrossRef
Objective Observed the efficacy of the patients with implantable cardioverter defibrillators (ICD). Sought to identify the reasons of inappropriate shocks, to minimise the risk of inappropriate shocks. Methods Followed up nineteen patients with ICD. Stored ICD electrograms from all shocks episodes were analysed to distinguish appropriate and inappropriate shocks according to history of tachycardia. Any ICD therapy not delivered for VT or VF was deemed inappropriate shocks therapy. It can be considered as inappropriate that ICD firing for sinus tachycardia, supraventricular tachycardia, atrial fibrillation, atrial flutter, and nonsustained ventricular tachycaidia. Result There is no shock episode in ten patients (52.63%). Appropriate shocks occurred in 6 patients (31.58%) Inappropriate shocks occurred in three patients (15.79%). Inappropriate shock episodes constituted 4 of 11 total shock episodes. We presented three cases of inappropriate shocks due to atrial fibrillation, sinus tachycardia and high frequency noise. Conclusions Inappropriate shocks occurred commonly in our cases. Varied reason induced the inappropriate shocks. Programing the device parameter and choosing proper discriminators can minimised rate of inappropriate shocks.
Title: e0565 Analysis of inappropriate therapy in the patients implantable cardioverter defibrillators
Description:
Objective Observed the efficacy of the patients with implantable cardioverter defibrillators (ICD).
Sought to identify the reasons of inappropriate shocks, to minimise the risk of inappropriate shocks.
Methods Followed up nineteen patients with ICD.
Stored ICD electrograms from all shocks episodes were analysed to distinguish appropriate and inappropriate shocks according to history of tachycardia.
Any ICD therapy not delivered for VT or VF was deemed inappropriate shocks therapy.
It can be considered as inappropriate that ICD firing for sinus tachycardia, supraventricular tachycardia, atrial fibrillation, atrial flutter, and nonsustained ventricular tachycaidia.
Result There is no shock episode in ten patients (52.
63%).
Appropriate shocks occurred in 6 patients (31.
58%) Inappropriate shocks occurred in three patients (15.
79%).
Inappropriate shock episodes constituted 4 of 11 total shock episodes.
We presented three cases of inappropriate shocks due to atrial fibrillation, sinus tachycardia and high frequency noise.
Conclusions Inappropriate shocks occurred commonly in our cases.
Varied reason induced the inappropriate shocks.
Programing the device parameter and choosing proper discriminators can minimised rate of inappropriate shocks.

Related Results

Pembrolizumab and Sarcoma: A meta-analysis
Pembrolizumab and Sarcoma: A meta-analysis
Abstract Introduction: Pembrolizumab is a monoclonal antibody that promotes antitumor immunity. This study presents a systematic review and meta-analysis of the efficacy and safety...
Implantable antennas for biomedical applications: a systematic review
Implantable antennas for biomedical applications: a systematic review
AbstractThis review presents an in-depth examination of implantable antennas for various biomedical purposes. The development of implantable antennas, including their designs, mate...
The Impact of IL28B Gene Polymorphisms on Drug Responses
The Impact of IL28B Gene Polymorphisms on Drug Responses
To achieve high therapeutic efficacy in the patient, information on pharmacokinetics, pharmacodynamics, and pharmacogenetics is required. With the development of science and techno...

Back to Top