Comparison of haemoglobin concentration measurements using HemoCue-301 and Sysmex XN-Series 1500: a survey among anaemic Gambian infants aged 6-12 months

View through CrossRef
ABSTRACTBackground & AimsIn low-income countries, point-of-care photometers are used in the screening and management of anaemia in individuals, but also in the assessment of population iron status when evaluating efficacy of intervention studies or public health interventions. We aimed to evaluate the accuracy of a commonly used photometer, HemoCue-301, in determining haemoglobin concentration among anaemic children aged 6-12 months in a field setting in rural Africa.MethodsThis report concerns a secondary analysis of data from Gambian infants being screened for an ongoing randomized controlled trial. In those found to be anaemic by HemoCue-301, haemoglobin concentration was measured by Sysmex XN-1500, an automated haematology analyser that was used as a reference. Passing-Bablok regression analysis was used to estimate the regression constant (systematic deviation between two measurement methods that remain consistent across the range of measurements) and proportional bias (systematic deviation between two measurement methods that change in magnitude relative to the value being measured).ResultsAnalysis was based on 227 participants. There was strong evidence of absolute bias among moderately anaemic participants (haemoglobin concentration at 8.0g/dL) (absolute bias: 1.12g/dL; 95% CI: 0.91 to 1.37g/dL; proportional bias: 14.0%; 95% CI: 11.4% to 17.1%) in haemoglobin concentrations measured by HemoCue-301 compared to those measured by Sysmex XN-Series1500. Bias was marginal at haemoglobin concentration of 11.0g/dL (absolute bias: −0.08g/dL; 95% CI: −0.18 to 0.07g/dL; proportional bias: −7.3%; 95% CI: −6.5% to 0.6%).ConclusionHaemoglobin measurements by HemoCue-301 seem substantially biased in participants with haemoglobin less than 8.0g/dL.
Title: Comparison of haemoglobin concentration measurements using HemoCue-301 and Sysmex XN-Series 1500: a survey among anaemic Gambian infants aged 6-12 months
Description:
ABSTRACTBackground & AimsIn low-income countries, point-of-care photometers are used in the screening and management of anaemia in individuals, but also in the assessment of population iron status when evaluating efficacy of intervention studies or public health interventions.
We aimed to evaluate the accuracy of a commonly used photometer, HemoCue-301, in determining haemoglobin concentration among anaemic children aged 6-12 months in a field setting in rural Africa.
MethodsThis report concerns a secondary analysis of data from Gambian infants being screened for an ongoing randomized controlled trial.
In those found to be anaemic by HemoCue-301, haemoglobin concentration was measured by Sysmex XN-1500, an automated haematology analyser that was used as a reference.
Passing-Bablok regression analysis was used to estimate the regression constant (systematic deviation between two measurement methods that remain consistent across the range of measurements) and proportional bias (systematic deviation between two measurement methods that change in magnitude relative to the value being measured).
ResultsAnalysis was based on 227 participants.
There was strong evidence of absolute bias among moderately anaemic participants (haemoglobin concentration at 8.
0g/dL) (absolute bias: 1.
12g/dL; 95% CI: 0.
91 to 1.
37g/dL; proportional bias: 14.
0%; 95% CI: 11.
4% to 17.
1%) in haemoglobin concentrations measured by HemoCue-301 compared to those measured by Sysmex XN-Series1500.
Bias was marginal at haemoglobin concentration of 11.
0g/dL (absolute bias: −0.
08g/dL; 95% CI: −0.
18 to 0.
07g/dL; proportional bias: −7.
3%; 95% CI: −6.
5% to 0.
6%).
ConclusionHaemoglobin measurements by HemoCue-301 seem substantially biased in participants with haemoglobin less than 8.
0g/dL.

Related Results

Capillary versus venous haemoglobin determination in the assessment of healthy blood donors
Capillary versus venous haemoglobin determination in the assessment of healthy blood donors
Background and ObjectivesTo determine the accuracy of fingerstick haemoglobin assessment in blood donors, the performance of a portable haemoglobinometer (HemoCue Hb 201+) was pros...
Histopathological Changes of Placenta Associated with Maternal Anaemia in Northeast Ethiopia: A Comparative Study
Histopathological Changes of Placenta Associated with Maternal Anaemia in Northeast Ethiopia: A Comparative Study
BACKGROUND: Anaemia during pregnancy affects about half of all pregnant mothers in developing countries; it is the major causes of indirect maternal mortality. Anaemia can directly...
Serum Ferritin Status in Microcytic Hypochromic Anaemic Medical Students
Serum Ferritin Status in Microcytic Hypochromic Anaemic Medical Students
Background: Microcytic hypochromic anaemia is common in young adults , specially in females. It is important to know the serum ferritin status of the medical students who will be i...
Sickle Cell Disease as a Multi‐Factorial Condition
Sickle Cell Disease as a Multi‐Factorial Condition
Abstract The phenotype of sickle cell anaemia is heterogeneous. Although all patients have the identical sickle haemoglobin mutat...
Iron Deficiency Anaemia in Newborn sla Mice: a Genetic Defect of Placental Iron Transport
Iron Deficiency Anaemia in Newborn sla Mice: a Genetic Defect of Placental Iron Transport
Summary. . Newborn mice with X‐linked anaemia (gene symbol sla) have lower haemoglobin levels at birth than normal and carrier mice but there is considerable overlap. Serial obser...

Back to Top