A protected inert‐gas sample manipulation and transfer environment for cryo electron microscopy and analytics
View through CrossRef
The preservation of the native sample state during manipulation and exchange between sample preparation and imaging devices is a prerequisite for artefact and contamination free structure investigation. In particular during cryo‐transfer and (re‐)mounting of frozen hydrated samples, which often in electron microscopy (EM), have a very large surface area compared to its volume, can be very challenging when samples have to be exchanged or remounted several times. Some of the main challenges are: 1. preventing the sample from warming up to the de‐vitrification temperature of water (approximately −137°C), 2. Preventing the sample from being covered with ice or other contaminants condensing from the surrounding atmosphere and 3. Handling these samples without chemical or mechanical alteration of their native state.
Currently the handling and transfer of samples under vacuum and cryo conditions is solved by systems provided from different suppliers. However the transfer of frozen (vitrified) samples under environmental pressure (e.g. from a high‐pressure‐machine to a low temperature coater or cryo‐EM) is still a challenge and a satisfactory practical solution is missing.
Here, we present an ambient pressure inert‐gas sample mounting and exchange chamber (Glove‐Box) (Fig. 1 &B) that addresses all the needs for structure and composition preservation during the manipulation, mounting and transfer of mainly frozen or hydration/oxygen sensitive samples. Once mounted these samples can be transferred and exchanged between different cryo‐preparation and imaging systems (Fig. 1 E) under controlled inert‐gas and temperature conditions. This controlled inert‐gas “Glove‐Box” includes a LN2 immersed sample mounting area with various interfaces to different inert‐gas/cryo transfer systems for RT or low temperature TEM or FIB/SEM applications, Furthermore, monitoring of the environmental parameters (Fig. 1 C) during transfer (humidity, oxygen, carbon dioxide), on‐screen imaging to support sample manipulation (Fig. 1 D), heating of tools and recording of all experimental parameters for documentation are included into this dedicated LN2 and inert‐gas “Glove‐Box”. The interface to other sample carrier devices is open and flexible and can be adapted upon individual laboratory needs. Finally such a “Glove‐Box” can play an essential role in a dedicated modern cryo‐EM & analytical environment as sketched in Fig. 1 E.
Title: A protected inert‐gas sample manipulation and transfer environment for cryo electron microscopy and analytics
Description:
The preservation of the native sample state during manipulation and exchange between sample preparation and imaging devices is a prerequisite for artefact and contamination free structure investigation.
In particular during cryo‐transfer and (re‐)mounting of frozen hydrated samples, which often in electron microscopy (EM), have a very large surface area compared to its volume, can be very challenging when samples have to be exchanged or remounted several times.
Some of the main challenges are: 1.
preventing the sample from warming up to the de‐vitrification temperature of water (approximately −137°C), 2.
Preventing the sample from being covered with ice or other contaminants condensing from the surrounding atmosphere and 3.
Handling these samples without chemical or mechanical alteration of their native state.
Currently the handling and transfer of samples under vacuum and cryo conditions is solved by systems provided from different suppliers.
However the transfer of frozen (vitrified) samples under environmental pressure (e.
g.
from a high‐pressure‐machine to a low temperature coater or cryo‐EM) is still a challenge and a satisfactory practical solution is missing.
Here, we present an ambient pressure inert‐gas sample mounting and exchange chamber (Glove‐Box) (Fig.
1 &B) that addresses all the needs for structure and composition preservation during the manipulation, mounting and transfer of mainly frozen or hydration/oxygen sensitive samples.
Once mounted these samples can be transferred and exchanged between different cryo‐preparation and imaging systems (Fig.
1 E) under controlled inert‐gas and temperature conditions.
This controlled inert‐gas “Glove‐Box” includes a LN2 immersed sample mounting area with various interfaces to different inert‐gas/cryo transfer systems for RT or low temperature TEM or FIB/SEM applications, Furthermore, monitoring of the environmental parameters (Fig.
1 C) during transfer (humidity, oxygen, carbon dioxide), on‐screen imaging to support sample manipulation (Fig.
1 D), heating of tools and recording of all experimental parameters for documentation are included into this dedicated LN2 and inert‐gas “Glove‐Box”.
The interface to other sample carrier devices is open and flexible and can be adapted upon individual laboratory needs.
Finally such a “Glove‐Box” can play an essential role in a dedicated modern cryo‐EM & analytical environment as sketched in Fig.
1 E.
Related Results
GRl-95-0466 Inert Base Gas Field Experiment
GRl-95-0466 Inert Base Gas Field Experiment
IGTs research efforts in the laboratory and over a decade of field experience of inert gas (nitrogen) application in French underground gas storage fields clearly indicate that rep...
Low BTU Gas Monetisation Technology Commercial Evaluation
Low BTU Gas Monetisation Technology Commercial Evaluation
Abstract
Natural gas is normally composed of three types of group components; i.e., hydrocarbons, inert and sour components.
The natural gas hyd...
A comprehensive foundation model for cryo-EM image processing
A comprehensive foundation model for cryo-EM image processing
AbstractCryogenic electron microscopy (cryo-EM) has become a premier technique for high-resolution structural determination of biological macromolecules1–4. However, its widespread...
People Analytics
People Analytics
People analytics refers to the systematic and scientific process of applying quantitative or qualitative data analysis methods to derive insights that shape and inform employee-rel...
Measuring whole body inert gas wash-out
Measuring whole body inert gas wash-out
Introduction: Quantifying inert gas wash-out is crucial to understanding the pathophysiology of decompression sickness. In this study, we developed a portable closed-circuit device...
Awareness of Cryo Electro-Tomography among Dental Students
Awareness of Cryo Electro-Tomography among Dental Students
CryoElectronomography (CryoET) is indeed an imaging method used to create high resolution (~1-4 nm) three-dimensional viewpoints of specimen, usually physiological macromolecules a...
Comparisons of Pore Structure for Unconventional Tight Gas, Coalbed Methane and Shale Gas Reservoirs
Comparisons of Pore Structure for Unconventional Tight Gas, Coalbed Methane and Shale Gas Reservoirs
Extended abstract
Tight sands gas, coalbed methane and shale gas are three kinds of typical unconventional natural gas. With the decrease of conventional oil and gas...
Optimization of adhesion for high throughput cryo-electron tomography of vitreous sections
Optimization of adhesion for high throughput cryo-electron tomography of vitreous sections
ABSTRACT
Cellular cryo electron tomography explores tissue and cells in their unstained flash-frozen native state, revealing in situ the structure of macromolecules tog...