Javascript must be enabled to continue!
A-401 Performance Evaluation of POCT Equipment for Blood Gas Measurement and Metabolic Parameters
View through CrossRef
Abstract
Background
According to the Clinical & Laboratory Standards Institute (CLSI) guideline, Point of Care technology (POCT) refers to devices used to provide clinical parameters at the site of a patient care or its proximity, allowing faster test results and improving medical response. However, the improvement in clinical outcomes depends on how healthcare professionals utilize this advantage and the performance of the chosen system. With a vast and growing scope, POCT devices enable the analysis of a wide range of analytes, including cardiac markers, coagulation markers, glucose, blood gases, chemical and metabolic parameters, among others. The Stat Profile Prime® analyzer, which exhibits POCT characteristics, offers benefits to patient care, especially during emergency periods and in restricted situations related to patient movement for conducting tests. In this context, the analytical verification of the device is important to clarify the process, the risks associated with the use of the equipment as well as suitability for the intended use. Moreover, performance evaluation ensures the quality and safety of the results obtained from the tests carried out. This study aimed to evaluate the performance of the POCT Stat Profile Prime equipment to measure blood gases and metabolic parameters.
Methods
Verification tests were conducted between October and December 2024 based on the CLSI-POCT09-A protocol. Precision study was assayed using three commercial control levels, used to monitor the performance of the Stat Profile Prime (Prime Auto QC Cartridge CCS), in quintuplicate over 5 days, for the quantitative determination of the parameters pH, PCO2, PO2, Htc, Na+, K+, Cl-, iCa, Gli (glucose), and Lac (lactate). The acceptance criteria for this analysis included the coefficient of variation (CV) or standard deviation (SD) according to the manufacturer*s specifications. For the comparative study, two sample groups were used: 42 heparinized whole blood samples and 34 serum samples for iCa. Both sample groups were tested on the Gem 3500 (Werfen) and Stat Profile Prime (Nova Biomedical) instruments. The allowable total error and correlation (r = 0.95) were considered approval criteria for the major of tested analytes and Kappa index (= 75%) for Htc. The linearity study was determined using four levels of the linearity standards (Nova Linearity Levels set), which were processed in three replicates, with the expected correlation of r = 0.99.
Results
The precision tests performed with 3 control levels showed a confidence interval greater than 95% and SD and CV close to the values specified by the manufacturer, indicating satisfactory performance for all evaluated parameters. In the comparative study, the tested parameters showed a correlation = 0.95, displaying 95% of the results within the total allowable error (TEa), and a Kappa index of 79.3% for Htc. According to quality specifications, the parameters from linearity study demonstrated a correlation = 0.99.
Conclusion
The Stat Profile Prime device demonstrated reliable performance and verification data in accordance with quality standards, ensuring safety and efficiency for clinical use. Furthermore, as a POCT, the equipment provides prompt and accurate results, which is crucial in a hospital environment, improving patient care.
Oxford University Press (OUP)
Title: A-401 Performance Evaluation of POCT Equipment for Blood Gas Measurement and Metabolic Parameters
Description:
Abstract
Background
According to the Clinical & Laboratory Standards Institute (CLSI) guideline, Point of Care technology (POCT) refers to devices used to provide clinical parameters at the site of a patient care or its proximity, allowing faster test results and improving medical response.
However, the improvement in clinical outcomes depends on how healthcare professionals utilize this advantage and the performance of the chosen system.
With a vast and growing scope, POCT devices enable the analysis of a wide range of analytes, including cardiac markers, coagulation markers, glucose, blood gases, chemical and metabolic parameters, among others.
The Stat Profile Prime® analyzer, which exhibits POCT characteristics, offers benefits to patient care, especially during emergency periods and in restricted situations related to patient movement for conducting tests.
In this context, the analytical verification of the device is important to clarify the process, the risks associated with the use of the equipment as well as suitability for the intended use.
Moreover, performance evaluation ensures the quality and safety of the results obtained from the tests carried out.
This study aimed to evaluate the performance of the POCT Stat Profile Prime equipment to measure blood gases and metabolic parameters.
Methods
Verification tests were conducted between October and December 2024 based on the CLSI-POCT09-A protocol.
Precision study was assayed using three commercial control levels, used to monitor the performance of the Stat Profile Prime (Prime Auto QC Cartridge CCS), in quintuplicate over 5 days, for the quantitative determination of the parameters pH, PCO2, PO2, Htc, Na+, K+, Cl-, iCa, Gli (glucose), and Lac (lactate).
The acceptance criteria for this analysis included the coefficient of variation (CV) or standard deviation (SD) according to the manufacturer*s specifications.
For the comparative study, two sample groups were used: 42 heparinized whole blood samples and 34 serum samples for iCa.
Both sample groups were tested on the Gem 3500 (Werfen) and Stat Profile Prime (Nova Biomedical) instruments.
The allowable total error and correlation (r = 0.
95) were considered approval criteria for the major of tested analytes and Kappa index (= 75%) for Htc.
The linearity study was determined using four levels of the linearity standards (Nova Linearity Levels set), which were processed in three replicates, with the expected correlation of r = 0.
99.
Results
The precision tests performed with 3 control levels showed a confidence interval greater than 95% and SD and CV close to the values specified by the manufacturer, indicating satisfactory performance for all evaluated parameters.
In the comparative study, the tested parameters showed a correlation = 0.
95, displaying 95% of the results within the total allowable error (TEa), and a Kappa index of 79.
3% for Htc.
According to quality specifications, the parameters from linearity study demonstrated a correlation = 0.
99.
Conclusion
The Stat Profile Prime device demonstrated reliable performance and verification data in accordance with quality standards, ensuring safety and efficiency for clinical use.
Furthermore, as a POCT, the equipment provides prompt and accurate results, which is crucial in a hospital environment, improving patient care.
Related Results
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
The Epidemiology of Tuberculosis among Internal Displaced Persons in EL-salam Camp, South Darfur State - Sudan 2022.
The Epidemiology of Tuberculosis among Internal Displaced Persons in EL-salam Camp, South Darfur State - Sudan 2022.
Abstract
Background: Sudan is a large country. It has a diverse economy, a large population, and a history of civil wars and conflicts. The poverty rate is high. With an es...
A combined laboratory and field evaluation of the Cholestech LDX and CardioChek PA point-of-care testing lipid and glucose analysers
A combined laboratory and field evaluation of the Cholestech LDX and CardioChek PA point-of-care testing lipid and glucose analysers
Background Combined lipid and glucose point-of-care testing (POCT) devices could facilitate widespread population screening for cardiovascular disease (CVD) and diabetes as part of...
Unconventional Reservoirs: Basic Petrophysical Concepts for Shale Gas
Unconventional Reservoirs: Basic Petrophysical Concepts for Shale Gas
Abstract
Unconventional reservoirs have burst with considerable force in oil and gas production worldwide. Shale Gas is one of them, with intense activity taking pla...
Comparisons of Pore Structure for Unconventional Tight Gas, Coalbed Methane and Shale Gas Reservoirs
Comparisons of Pore Structure for Unconventional Tight Gas, Coalbed Methane and Shale Gas Reservoirs
Extended abstract
Tight sands gas, coalbed methane and shale gas are three kinds of typical unconventional natural gas. With the decrease of conventional oil and gas...
Point-of-Care Blood Testing in Severe Mental Illness: A Mixed-Methods Evaluation
Point-of-Care Blood Testing in Severe Mental Illness: A Mixed-Methods Evaluation
AimsThere is a significant mortality gap between the general population and people with SMI. This is especially prominent in those with psychotic disorders, underpinned by an incre...
Blood Cross Matching Without Anti-Human Globulin (AHG) and Bovine Serum: A New Interest for an Old Idea
Blood Cross Matching Without Anti-Human Globulin (AHG) and Bovine Serum: A New Interest for an Old Idea
Abstract
Introduction
Transfusion medicine promotes the safety of blood transfusions by rigorously testing to eliminate risks of infection and hemolytic. The efficacy (to correct ...
Critical Gas Saturation During Depressurisation and its Importance in the Brent Field
Critical Gas Saturation During Depressurisation and its Importance in the Brent Field
Critical Gas Saturation During Depressurisation and its Importance in the Brent Field.
Abstract
After some 20 years of pressure ...

