(121) Validation of a Digital Rigidometer for the Evaluation of Erection Hardness During In-office Rigidity Assessment in Patients with Erectile Dysfunction
View through CrossRef
Abstract
Introduction
The evaluation of erectile function should include the measurement of axial rigidity, since it is one of the key factors in the pathophysiology, diagnosis and treatment of ED. This assessment is usually performed following in-office intracavernosal injections of vasoactive agents during Doppler ultrasound studies, curvature assessments or simply prior to initiation of intracavernosal pharmacotherapy. In this scenario, the Erectile Hardness Score (EHS) is the most common validated tool to evaluate penile rigidity. Although highly reproducible, EHS is still based on the examiner’s subjective perspective and might eventually lead to conflicting results.
Objective
The aim of the present study was to assess the diagnostic accuracy of a digital rigidometer in comparison to the EHS scale to allow for a more detailed rigidity assessment using an objective measure of penile axial load.
Methods
First we performed a controlled experiment using a digital dynamometer coupled to a 3D-printed tip to assess the maximal load applied to the Pfizer’s Erection Hardness Meter (Figure 1). Three urologists with expertise on the management of ED performed 2 different measures of axial rigidity on each shaft corresponding to the EHS scale in order to suggest a 4-grade classification range. Then, a sample of patients was analyzed during in-office erectile tests performed by 2 different urologists using redosing protocols of vasoactive agents from February to June 2022. Coefficients of variation were calculated using analysis of variance (ANOVA) and t-test was performed to compare the interval data. ROC curves were used to estimate the diagnostic accuracy of the digital rigidometer in comparison to the EHS. P values below 0.05 were considered significant.
Results
Mean axial load supported by each shaft of the Pfizer’s Erection Hardness Meter was: EHS 1: 0.51 ± 0.05 Kg; EHS 2: 0.75 ± 0.05 Kg; EHS 3: 1.2 ± 0.13 Kg; and EHS 4: 2.5 ± 0.51 Kg. The following coefficients of variations were found: EHS 1: 10.1%, EHS 2: 6.85%, EHS 3: 11.2%, and EHS 4: 20.4%. Distribution of different measures is shown in Figure 2. A total of 30 men with mean age of 54.5 ± 15.5 years were included in the analysis. The mean IIEF-EFD was 16.3 ± 5.1. During erection assessment 3 (10%) presented erection EHS 2, 7 (23.3%) EHS 3, and 20 (66.6%) EHS 4. The mean dose of trimix was 0,35 ml and 50% required sympathomimetic medication for erection reversion after the procedure. The digital rigidometer had an AUC of 0.91 (95% CI 0.78-1.00, p=0.0005) to discriminate EHS>3 (Figure 3). The best cuff value for that purpose is 1.2 Kg, yielding a sensitivity of 85.7% and specificity of 88.9%.
Conclusions
The digital rigidometer tested in the present study appears to provide reproducible measures of axial penile rigidity that might be useful in clinical practice. It has demonstrated interesting diagnostic properties for the in-office assessment of erectile rigidity in comparison to the EHS, in which the cutoff of 1.2 Kg is the one with the best diagnostic properties. An objective measurement of penile axial rigidity can contribute to both clinical and research environment.
Disclosure
No
Oxford University Press (OUP)
Title: (121) Validation of a Digital Rigidometer for the Evaluation of Erection Hardness During In-office Rigidity Assessment in Patients with Erectile Dysfunction
Description:
Abstract
Introduction
The evaluation of erectile function should include the measurement of axial rigidity, since it is one of the key factors in the pathophysiology, diagnosis and treatment of ED.
This assessment is usually performed following in-office intracavernosal injections of vasoactive agents during Doppler ultrasound studies, curvature assessments or simply prior to initiation of intracavernosal pharmacotherapy.
In this scenario, the Erectile Hardness Score (EHS) is the most common validated tool to evaluate penile rigidity.
Although highly reproducible, EHS is still based on the examiner’s subjective perspective and might eventually lead to conflicting results.
Objective
The aim of the present study was to assess the diagnostic accuracy of a digital rigidometer in comparison to the EHS scale to allow for a more detailed rigidity assessment using an objective measure of penile axial load.
Methods
First we performed a controlled experiment using a digital dynamometer coupled to a 3D-printed tip to assess the maximal load applied to the Pfizer’s Erection Hardness Meter (Figure 1).
Three urologists with expertise on the management of ED performed 2 different measures of axial rigidity on each shaft corresponding to the EHS scale in order to suggest a 4-grade classification range.
Then, a sample of patients was analyzed during in-office erectile tests performed by 2 different urologists using redosing protocols of vasoactive agents from February to June 2022.
Coefficients of variation were calculated using analysis of variance (ANOVA) and t-test was performed to compare the interval data.
ROC curves were used to estimate the diagnostic accuracy of the digital rigidometer in comparison to the EHS.
P values below 0.
05 were considered significant.
Results
Mean axial load supported by each shaft of the Pfizer’s Erection Hardness Meter was: EHS 1: 0.
51 ± 0.
05 Kg; EHS 2: 0.
75 ± 0.
05 Kg; EHS 3: 1.
2 ± 0.
13 Kg; and EHS 4: 2.
5 ± 0.
51 Kg.
The following coefficients of variations were found: EHS 1: 10.
1%, EHS 2: 6.
85%, EHS 3: 11.
2%, and EHS 4: 20.
4%.
Distribution of different measures is shown in Figure 2.
A total of 30 men with mean age of 54.
5 ± 15.
5 years were included in the analysis.
The mean IIEF-EFD was 16.
3 ± 5.
1.
During erection assessment 3 (10%) presented erection EHS 2, 7 (23.
3%) EHS 3, and 20 (66.
6%) EHS 4.
The mean dose of trimix was 0,35 ml and 50% required sympathomimetic medication for erection reversion after the procedure.
The digital rigidometer had an AUC of 0.
91 (95% CI 0.
78-1.
00, p=0.
0005) to discriminate EHS>3 (Figure 3).
The best cuff value for that purpose is 1.
2 Kg, yielding a sensitivity of 85.
7% and specificity of 88.
9%.
Conclusions
The digital rigidometer tested in the present study appears to provide reproducible measures of axial penile rigidity that might be useful in clinical practice.
It has demonstrated interesting diagnostic properties for the in-office assessment of erectile rigidity in comparison to the EHS, in which the cutoff of 1.
2 Kg is the one with the best diagnostic properties.
An objective measurement of penile axial rigidity can contribute to both clinical and research environment.
Disclosure
No.
Related Results
Current therapeutic strategies for erectile function recovery after radical prostatectomy – literature review and meta-analysis
Current therapeutic strategies for erectile function recovery after radical prostatectomy – literature review and meta-analysis
Radical prostatectomy is the most commonly performed treatment option for localised prostate cancer. In the last decades the surgical technique has been improved and modified in or...
Unraveling the intersection of sleep disorders and erectile dysfunction: Outcomes from two EPISONO editions
Unraveling the intersection of sleep disorders and erectile dysfunction: Outcomes from two EPISONO editions
AbstractBackgroundThere is growing interest in the relationship between sleep disorders and erectile dysfunction. We present the results from a 2015 follow‐up study in relation to ...
Erectile Dysfunction And Associated Factors Among Adult Hypertensive Patients Attending Outpatient Care At Public Hospitals in Harari Regional State,Eastern Ethiopia
Erectile Dysfunction And Associated Factors Among Adult Hypertensive Patients Attending Outpatient Care At Public Hospitals in Harari Regional State,Eastern Ethiopia
Abstract
Introduction. Sexual dysfunction is lack of sexual ideas or thoughts, diminished sexual interest or desire, and unresponsive sexual desire. Sexual dysfunctions is ...
Erectile Dysfunction and Associated Factors Among Adult Hypertensive Patients Attending Outpatient Care at Public Hospitals in Harari Regional State, Eastern Ethiopia
Erectile Dysfunction and Associated Factors Among Adult Hypertensive Patients Attending Outpatient Care at Public Hospitals in Harari Regional State, Eastern Ethiopia
<i>Introduction</i>. Sexual dysfunction is lack of sexual ideas or thoughts, diminished sexual interest or desire, and unresponsive sexual desire. Sexua...
Using Statistics and Data Mining Approaches to Analyze Male Sexual Behaviors and Use of Erectile Dysfunction Drugs Based on Large Questionnaire Data
Using Statistics and Data Mining Approaches to Analyze Male Sexual Behaviors and Use of Erectile Dysfunction Drugs Based on Large Questionnaire Data
The prevalence of erectile dysfunction (ED) has been extensively studied worldwide. Erectile dysfunction drugs has shown great efficacy in preventing male erectile dysfunction. In ...
Evaluation of Choroidal Vascularity Index, Retinal and Optic Nerve Changes in Erectile Dysfunction
Evaluation of Choroidal Vascularity Index, Retinal and Optic Nerve Changes in Erectile Dysfunction
Abstract
Objective: To investigate choroidal alterations through assessment of subfoveal choroidal thickness (SFCT) and choroidal vascularity index (CVI), optic nerve modif...
Comparison of Satisfaction With Penile Prosthesis Implantation in Patients With Radical Prostatectomy or Radical Cystoprostatectomy to the General Population
Comparison of Satisfaction With Penile Prosthesis Implantation in Patients With Radical Prostatectomy or Radical Cystoprostatectomy to the General Population
AbstractIntroductionPenile prosthesis implantation is a widely used treatment option for erectile dysfunction. Data is limited with regard to patient satisfaction with a penile pro...
Changes in the penile microcirculation in the treatment of mixed erectile dysfunction on the background of arterial hypentension in locomotive team workers
Changes in the penile microcirculation in the treatment of mixed erectile dysfunction on the background of arterial hypentension in locomotive team workers
BACKGROUND: Railway workers constantly experience stress in the workplace, associated with the need to maintain high attention and be ready for emergency action, which provokes a p...