Long-Term Follow-up of Standard and Small Diameter Implantable Cardioverter Leads
View through CrossRef
Background: Small-diameter implantable cardioverter-defibrillator (ICD) leads have been introduced into clinical practice to facilitate the implantation procedure. Despite their expected benefits, the reliability of these leads has proven to be questionable. The main purpose of our study is to investigate the impact of ICD lead diameter (≤8 F versus >8 F) on long-term lead durability.
Methods: Overall, 206 consecutive patients implanted with a right ventricular ICD lead in the Electrophysiology and Cardiac Pacing Unit of our department from January 2008 to December 2013 were included in this analysis. ICD leads were defined, according to their diameter, as small (≤8 F) and standard (>8 F).The small-diameter leads (n=106) included Linox (Biotronik; n=58) and Durata (St. Jude Medical/Abbot; n=48). The standard-diameter ICD leads (n=100) consisted of Sprint Quattro (Medtronic; n=64) and Endotak (Boston Scientific; n=36).
Results: After a median follow-up of 7.3 years, lead failure rate was significantly increased for small-diameter leads compared with standard-diameter leads (6.6% vs 1%; P=0.035). No difference in lead survival probability has been observed between Linox and Durata small-diameter leads (93% vs 92.7%; P=0.71). The majority of lead failures presented as noise (87.5%), without detectable abnormalities on fluoroscopic evaluation.
Conclusion: Our single-centre study showed that both Linox and Durata small-diameter ICD are associated to be more susceptible to a greater risk of lead failure as compared to standard-diameter ICD leads. In this perspective, a comprehensive vigilance strategy including home monitoring is warranted for early detection of lead failure.
Mapsci Digital Publisher OPC Pvt. Ltd.
Title: Long-Term Follow-up of Standard and Small Diameter Implantable Cardioverter Leads
Description:
Background: Small-diameter implantable cardioverter-defibrillator (ICD) leads have been introduced into clinical practice to facilitate the implantation procedure.
Despite their expected benefits, the reliability of these leads has proven to be questionable.
The main purpose of our study is to investigate the impact of ICD lead diameter (≤8 F versus >8 F) on long-term lead durability.
Methods: Overall, 206 consecutive patients implanted with a right ventricular ICD lead in the Electrophysiology and Cardiac Pacing Unit of our department from January 2008 to December 2013 were included in this analysis.
ICD leads were defined, according to their diameter, as small (≤8 F) and standard (>8 F).
The small-diameter leads (n=106) included Linox (Biotronik; n=58) and Durata (St.
Jude Medical/Abbot; n=48).
The standard-diameter ICD leads (n=100) consisted of Sprint Quattro (Medtronic; n=64) and Endotak (Boston Scientific; n=36).
Results: After a median follow-up of 7.
3 years, lead failure rate was significantly increased for small-diameter leads compared with standard-diameter leads (6.
6% vs 1%; P=0.
035).
No difference in lead survival probability has been observed between Linox and Durata small-diameter leads (93% vs 92.
7%; P=0.
71).
The majority of lead failures presented as noise (87.
5%), without detectable abnormalities on fluoroscopic evaluation.
Conclusion: Our single-centre study showed that both Linox and Durata small-diameter ICD are associated to be more susceptible to a greater risk of lead failure as compared to standard-diameter ICD leads.
In this perspective, a comprehensive vigilance strategy including home monitoring is warranted for early detection of lead failure.
Related Results
Implantable antennas for biomedical applications: a systematic review
Implantable antennas for biomedical applications: a systematic review
AbstractThis review presents an in-depth examination of implantable antennas for various biomedical purposes. The development of implantable antennas, including their designs, mate...
IMPLANTABLE DEFIBRILLATOR LEAD EXTRACTION WITH OPTIMISED STANDARD EXTRACTION TECHNIQUES
IMPLANTABLE DEFIBRILLATOR LEAD EXTRACTION WITH OPTIMISED STANDARD EXTRACTION TECHNIQUES
Objectives
Implantable cardioverter-defibrillator (ICD) leads might not be extracted in especially developing countries because of the high cost and lack of speci...
Aortic knob diameter in chest radiographs of healthy adults in Uganda.
Aortic knob diameter in chest radiographs of healthy adults in Uganda.
Abstract
Background
The burden of cardiovascular diseases, such as aortic and degenerative diseases grows in the aging population. Chest radiograph still plays an importan...
Does Patient Compliance Influence Wearable Cardioverter Defibrillator Effectiveness? A Single-Center Experience
Does Patient Compliance Influence Wearable Cardioverter Defibrillator Effectiveness? A Single-Center Experience
The study was designed to assess patient adherence to wearable cardioverter defibrillator as an indicator of device effectiveness. The patient training is not widely properly stand...
Implantable drug delivery systems: An overviews
Implantable drug delivery systems: An overviews
Implant are sterile solid mass that contains medicine, prepared by different ways like extrusion, moulding or contraction. The conventional routes of medicine administration has li...
Analisis performansi kolektor surya pelat datar dengan diameter lubang sirip berbeda sebagai impinging jet
Analisis performansi kolektor surya pelat datar dengan diameter lubang sirip berbeda sebagai impinging jet
Solar collector is a device that is able to absorb and transfer heat from solar energy to the working fluid. Solar collectors in general, use a mass flow rate of air that is parall...
Long-term care needs and hospitalization costs with long-term care insurance: a mixed-sectional study
Long-term care needs and hospitalization costs with long-term care insurance: a mixed-sectional study
BackgroundWith the rapid aging of the population, the health needs of the older adult have increased significantly, resulting in the frequent occurrence of the “social hospitalizat...
Mid-long term results after elective Impella-supported, complex high-risk procedures of percutaneous coronary intervention
Mid-long term results after elective Impella-supported, complex high-risk procedures of percutaneous coronary intervention
Abstract
Background
Mechanical circulatory support with percutaneous left ventricular assist device (pLVAD) in complex hi...