Long-Term Follow-up of Standard and Small Diameter Implantable Cardioverter Leads

View through CrossRef
Background: Small-diameter implantable cardioverter-defibrillator (ICD) leads have been introduced into clinical practice to facilitate the implantation procedure. Despite their expected benefits, the reliability of these leads has proven to be questionable. The main purpose of our study is to investigate the impact of ICD lead diameter (≤8 F versus >8 F) on long-term lead durability. Methods: Overall, 206 consecutive patients implanted with a right ventricular ICD lead in the Electrophysiology and Cardiac Pacing Unit of our department from January 2008 to December 2013 were included in this analysis. ICD leads were defined, according to their diameter, as small (≤8 F) and standard (>8 F).The small-diameter leads (n=106) included Linox (Biotronik; n=58) and Durata (St. Jude Medical/Abbot; n=48). The standard-diameter ICD leads (n=100) consisted of Sprint Quattro (Medtronic; n=64) and Endotak (Boston Scientific; n=36). Results: After a median follow-up of 7.3 years, lead failure rate was significantly increased for small-diameter leads compared with standard-diameter leads (6.6% vs 1%; P=0.035). No difference in lead survival probability has been observed between Linox and Durata small-diameter leads (93% vs 92.7%; P=0.71). The majority of lead failures presented as noise (87.5%), without detectable abnormalities on fluoroscopic evaluation. Conclusion: Our single-centre study showed that both Linox and Durata small-diameter ICD are associated to be more susceptible to a greater risk of lead failure as compared to standard-diameter ICD leads. In this perspective, a comprehensive vigilance strategy including home monitoring is warranted for early detection of lead failure.
Title: Long-Term Follow-up of Standard and Small Diameter Implantable Cardioverter Leads
Description:
Background: Small-diameter implantable cardioverter-defibrillator (ICD) leads have been introduced into clinical practice to facilitate the implantation procedure.
Despite their expected benefits, the reliability of these leads has proven to be questionable.
The main purpose of our study is to investigate the impact of ICD lead diameter (≤8 F versus >8 F) on long-term lead durability.
Methods: Overall, 206 consecutive patients implanted with a right ventricular ICD lead in the Electrophysiology and Cardiac Pacing Unit of our department from January 2008 to December 2013 were included in this analysis.
ICD leads were defined, according to their diameter, as small (≤8 F) and standard (>8 F).
The small-diameter leads (n=106) included Linox (Biotronik; n=58) and Durata (St.
Jude Medical/Abbot; n=48).
The standard-diameter ICD leads (n=100) consisted of Sprint Quattro (Medtronic; n=64) and Endotak (Boston Scientific; n=36).
Results: After a median follow-up of 7.
3 years, lead failure rate was significantly increased for small-diameter leads compared with standard-diameter leads (6.
6% vs 1%; P=0.
035).
No difference in lead survival probability has been observed between Linox and Durata small-diameter leads (93% vs 92.
7%; P=0.
71).
The majority of lead failures presented as noise (87.
5%), without detectable abnormalities on fluoroscopic evaluation.
Conclusion: Our single-centre study showed that both Linox and Durata small-diameter ICD are associated to be more susceptible to a greater risk of lead failure as compared to standard-diameter ICD leads.
In this perspective, a comprehensive vigilance strategy including home monitoring is warranted for early detection of lead failure.

Related Results

IMPLANTABLE DEFIBRILLATOR LEAD EXTRACTION WITH OPTIMISED STANDARD EXTRACTION TECHNIQUES
IMPLANTABLE DEFIBRILLATOR LEAD EXTRACTION WITH OPTIMISED STANDARD EXTRACTION TECHNIQUES
Objectives Implantable cardioverter-defibrillator (ICD) leads might not be extracted in especially developing countries because of the high cost and lack of speci...
Does Patient Compliance Influence Wearable Cardioverter Defibrillator Effectiveness? A Single-Center Experience
Does Patient Compliance Influence Wearable Cardioverter Defibrillator Effectiveness? A Single-Center Experience
The study was designed to assess patient adherence to wearable cardioverter defibrillator as an indicator of device effectiveness. The patient training is not widely properly stand...
Implantable drug delivery systems: An overviews
Implantable drug delivery systems: An overviews
Implant are sterile solid mass that contains medicine, prepared by different ways like extrusion, moulding or contraction. The conventional routes of medicine administration has li...
Long-term care needs and hospitalization costs with long-term care insurance: a mixed-sectional study
Long-term care needs and hospitalization costs with long-term care insurance: a mixed-sectional study
BackgroundWith the rapid aging of the population, the health needs of the older adult have increased significantly, resulting in the frequent occurrence of the “social hospitalizat...
Effect of stent graft fixation types on pararenal aortic diameter and renal function after endovascular aortic repair
Effect of stent graft fixation types on pararenal aortic diameter and renal function after endovascular aortic repair
Background: This study aims to identify pararenal aortic diameter changes following endovascular aortic repair and to investigate the effects of suprarenal fixation on renal functi...
Unusual Presentation of Mixed Lymphatic Malformation: A Case Report with Literature Review
Unusual Presentation of Mixed Lymphatic Malformation: A Case Report with Literature Review
Abstract Introduction There is a scarcity of reports on mixed lymphatic malformation. This case highlights a child with an extensive mixed lymphatic malformation, disfiguring multi...
Extravascular implantable cardioverter and leadless pacemaker interactions
Extravascular implantable cardioverter and leadless pacemaker interactions
Abstract Background Traditional cardiac pacing and defibrillation devices rely on leads connected to a subcutaneous pulse genera...
Quality of long-term follow up and results of sleeve gastrectomy: 5-year prospective observational cohort study. 114 cases
Quality of long-term follow up and results of sleeve gastrectomy: 5-year prospective observational cohort study. 114 cases
Abstract Introduction Sleeve gastrectomy (SG) is the most used procedure worldwide for the treatment of morbid obesity. Despite sustained weight loss and improvement of com...

Back to Top