GW24-e1773 Application of a novel pacing guide wire in cardiac resynchronisation therapy
View through CrossRef
Objectives
The implantation of left ventricular (LV) lead was the most complicated process in the cardiac resynchronisation therapy (CRT), the aim of this study was to investigate the availability of a novel Visionwire guide wire in the implantation of LV lead in CRT.
Methods
Five heart failure patients selected for CRT were involved in the study. The pacing threshold, R-wave sensing, phrenic nerve stimulation at local coronary sinus (CS) branch were measured by the Visionwire guide wire and LV lead separately.
Results
The pacing parameters were analysed through the Visionwire guide wire and LV lead separately at the 16 CS branches in 5 patients. There were no significant difference between the pacing parameters measured by the two methods. Significant correlation was found between LV pacing threshold assessed by the Visionwire guide wire and LV lead (r = 0.90, P < 0.01). Correlation for R wave sensing was also significant (r = 0.67, P < 0.01). The pacing sites accompanied with phrenic nerve stimulation while pacing at 10 V/0.42 ms were similar by the two methods. Visionwire guide wire could be used for local potential electrical mapping. The procedure time for pacing test in a single coronary sinus branch by Visionwire guide wire was much less than that by LV lead. (12.4 ± 7.5min vs 18.3 ± 12.2min, P < 0.01).
Conclusions
Visionwire guide wire facilitated transvenous LV lead implantation by prediction of pacing parameter and locate the target CS branch quickly. The electrophysiological mapping function of Visionwire guide wire as a unipolar lead might have further potential usefulness.
Title: GW24-e1773 Application of a novel pacing guide wire in cardiac resynchronisation therapy
Description:
Objectives
The implantation of left ventricular (LV) lead was the most complicated process in the cardiac resynchronisation therapy (CRT), the aim of this study was to investigate the availability of a novel Visionwire guide wire in the implantation of LV lead in CRT.
Methods
Five heart failure patients selected for CRT were involved in the study.
The pacing threshold, R-wave sensing, phrenic nerve stimulation at local coronary sinus (CS) branch were measured by the Visionwire guide wire and LV lead separately.
Results
The pacing parameters were analysed through the Visionwire guide wire and LV lead separately at the 16 CS branches in 5 patients.
There were no significant difference between the pacing parameters measured by the two methods.
Significant correlation was found between LV pacing threshold assessed by the Visionwire guide wire and LV lead (r = 0.
90, P < 0.
01).
Correlation for R wave sensing was also significant (r = 0.
67, P < 0.
01).
The pacing sites accompanied with phrenic nerve stimulation while pacing at 10 V/0.
42 ms were similar by the two methods.
Visionwire guide wire could be used for local potential electrical mapping.
The procedure time for pacing test in a single coronary sinus branch by Visionwire guide wire was much less than that by LV lead.
(12.
4 ± 7.
5min vs 18.
3 ± 12.
2min, P < 0.
01).
Conclusions
Visionwire guide wire facilitated transvenous LV lead implantation by prediction of pacing parameter and locate the target CS branch quickly.
The electrophysiological mapping function of Visionwire guide wire as a unipolar lead might have further potential usefulness.
Related Results
ASSA13-02-22 Application of a Novel Pacing Guide Wire in Cardiac Resynchronization Therapy
ASSA13-02-22 Application of a Novel Pacing Guide Wire in Cardiac Resynchronization Therapy
Background
The implantation of left ventricular (LV) lead was the most complicated process in the cardiac resynchronization therapy (CRT), the aim of this study w...
RESYNCHRONISATION THERAPY IN PATIENTS WITH CHRONIC HEART FAILURE (10 CASES REPORTED PLUS)
RESYNCHRONISATION THERAPY IN PATIENTS WITH CHRONIC HEART FAILURE (10 CASES REPORTED PLUS)
Objectives
To evaluate the clinical effect of cardiac resynchronisation therapy on systolic function in patients with New York Heart Association class III or IV c...
Long‐Term Follow‐up Impact of Dual‐Chamber Pacing on Patients with Hypertrophic Obstructive Cardiomyopathy
Long‐Term Follow‐up Impact of Dual‐Chamber Pacing on Patients with Hypertrophic Obstructive Cardiomyopathy
BackgroundPacing has been proposed as a treatment for patients with hypertrophic obstructive cardiomyopathy (HOCM), but there are few studies with long‐term follow‐up. We evaluated...
Hemodynamic Effects of Artificial Pacing in Complete Heart Block Complicating Acute Myocardial Infarction
Hemodynamic Effects of Artificial Pacing in Complete Heart Block Complicating Acute Myocardial Infarction
The hemodynamic changes associated with complete heart block complicating acute myocardial infarction and the effects of artificial pacing at various rates on these measurements we...
CLINICAL APPLICATION OF EMERGENCY BEDSIDE TEMPORARY CARDIAC PACING WITHOUT X-RAY FLUOROSCOPY
CLINICAL APPLICATION OF EMERGENCY BEDSIDE TEMPORARY CARDIAC PACING WITHOUT X-RAY FLUOROSCOPY
Objectives
To investigate the clinical value and safety of emergency bedside temporary cardiac pacing without X-ray fluoroscopy.
...
Numerical simulation of wire temperature field for prediction of wire transfer stability in laser hot wire welding
Numerical simulation of wire temperature field for prediction of wire transfer stability in laser hot wire welding
With preheating wire by resistance heat, laser hot wire welding improves process stability and wire deposition efficiency, which gives broad potential applications in surfacing and...
CARDIAC RESYNCHRONISATION THERAPY EFFECTS COMPARE OF THE SURGERY EPICARDIAL LEAD VERSUS CORONARY SINUS LEAD PLACEMENT
CARDIAC RESYNCHRONISATION THERAPY EFFECTS COMPARE OF THE SURGERY EPICARDIAL LEAD VERSUS CORONARY SINUS LEAD PLACEMENT
Objectives
Cardiac resynchronisation therapy (CRT) with biventricular pacing has demonstrated cardiac function improvement. Two strategies (coronary sinus vs epic...
Right ventricular pacing is not harmful as perceived previously
Right ventricular pacing is not harmful as perceived previously
Abstract
Background: Pacing from the apex of the RV is not optimal. This study aimed to assess and compare RV and LV pacing of LV function and dyssynchrony.
Thirty-six pati...