Search engine for discovering works of Art, research articles, and books related to Art and Culture
ShareThis
Javascript must be enabled to continue!

Efficacy of the Flow Re-direction Endoluminal Device for cerebral aneurysms and causes of failed deployment

View through CrossRef
Abstract Purpose The Flow Re-direction Endoluminal Device (FRED) has recently become available for flow diversion in Japan. We have encountered cases that failed to deploy the FRED. In this study, we report our initial experience with the FRED for cerebral aneurysms and clarify the causes of failed FRED deployment. Methods A retrospective data analysis was performed to identify patients treated with the FRED between June 2020 and March 2021. Follow-up digital subtraction angiography was performed at 3 and 6 months and assessed using the O’Kelly-Marotta (OKM) grading scale. Results Thirty-nine aneurysms in 36 patients (average age: 54.4 years) were treated with the FRED. The average sizes of the dome and neck were 9.9 mm and 5.2 mm, respectively. In nine patients, additional coiling was performed. In one patient (2.6%), proximal vessel injury caused direct carotid-cavernous fistula during deployment. Ischaemic complications were encountered in one patient (2.6%) with transient symptoms. Angiographic follow-up at 6 months revealed OKM grade C or D in 86.6% of patients. FRED deployment was successful in 35 (92.1%) procedures. In the failure group, the differences between the FRED and the minimum vessel diameter (P = 0.04) and the rate of the parent vessel having an S-shaped curve (P = 0.04) were greater than those in the success group. Conclusions Flow diversion using the FRED is effective and safe for treating cerebral aneurysms. The use of the FRED for patients with an S-shaped curve in the parent vessel and oversizing of more than 2 mm should be considered carefully.
Title: Efficacy of the Flow Re-direction Endoluminal Device for cerebral aneurysms and causes of failed deployment
Description:
Abstract Purpose The Flow Re-direction Endoluminal Device (FRED) has recently become available for flow diversion in Japan.
We have encountered cases that failed to deploy the FRED.
In this study, we report our initial experience with the FRED for cerebral aneurysms and clarify the causes of failed FRED deployment.
Methods A retrospective data analysis was performed to identify patients treated with the FRED between June 2020 and March 2021.
Follow-up digital subtraction angiography was performed at 3 and 6 months and assessed using the O’Kelly-Marotta (OKM) grading scale.
Results Thirty-nine aneurysms in 36 patients (average age: 54.
4 years) were treated with the FRED.
The average sizes of the dome and neck were 9.
9 mm and 5.
2 mm, respectively.
In nine patients, additional coiling was performed.
In one patient (2.
6%), proximal vessel injury caused direct carotid-cavernous fistula during deployment.
Ischaemic complications were encountered in one patient (2.
6%) with transient symptoms.
Angiographic follow-up at 6 months revealed OKM grade C or D in 86.
6% of patients.
FRED deployment was successful in 35 (92.
1%) procedures.
In the failure group, the differences between the FRED and the minimum vessel diameter (P = 0.
04) and the rate of the parent vessel having an S-shaped curve (P = 0.
04) were greater than those in the success group.
Conclusions Flow diversion using the FRED is effective and safe for treating cerebral aneurysms.
The use of the FRED for patients with an S-shaped curve in the parent vessel and oversizing of more than 2 mm should be considered carefully.

Related Results

Determinants of Cerebrovascular Reserve in Patients with Significant Carotid Stenosis
Determinants of Cerebrovascular Reserve in Patients with Significant Carotid Stenosis
Abstract Introduction In patients with 70% to 99% diameter carotid artery stenosis cerebral blood flow reserve may be protectiv...
Outcome evaluation of flow diverter stent placement in treatment of ruptured and unruptured cerebral aneurysms
Outcome evaluation of flow diverter stent placement in treatment of ruptured and unruptured cerebral aneurysms
Background: Cerebral aneurysms affect about 5% of the population and are associated with high mortality rates and severe complications for families and society. Complex aneurysm ty...
Two possible hemodynamic mechanisms underlying the growth of cerebral aneurysms depending on their size: The NHO CFD ABO study
Two possible hemodynamic mechanisms underlying the growth of cerebral aneurysms depending on their size: The NHO CFD ABO study
Cerebral aneurysm rupture has a poor prognosis, and growing aneurysms are prone to rupture. We therefore conducted a prospective observational study to clarify hemodynamics inducin...
Flow Diverter
Flow Diverter
The flow diverter is a relatively new but revolutionary technology in the realm of Endo vascular management of complex cerebral aneurysms. The traditional approach in the past has ...
Efficacy of Flow Diverters in Managing Multiple Aneurysms: A Clinical Case Report
Efficacy of Flow Diverters in Managing Multiple Aneurysms: A Clinical Case Report
Brain aneurysms, also known as cerebral aneurysms, are the growths of the parent artery. Based on their shape, aneurysms can be categorized as saccular or non-saccular. Several fac...
Upper Extremity Aneurysms
Upper Extremity Aneurysms
Aneurysms of the upper extremity arteries can be divided into those that occur in the central great vessels and those that occur in the arteries of the upper extremity. Aneurysms o...
New Insights into the Pathophysiology of Coronary Artery Aneurysms
New Insights into the Pathophysiology of Coronary Artery Aneurysms
Coronary aneurysms are typically defined as sections of a coronary artery where the diameter is more than 1.5 times that of an adjacent normal segment. In rare circumstances, these...

Back to Top