Fracture of a polyurethane cardiac catheter in the aortic arch: A complication related to polymer aging
View through CrossRef
AbstractAn undated Cordis Ducor polyurethane coronary catheter fractured and separated in the aortic arch during a cardiac catheterization, and was removed with an intravascular retriever. The recovered specimen appeared brittle, prompting an investigation of the effects of aging on the tensile strength of 91 polyurethane catheters manufactured by Cordis and the United States Catheter and Instrument Company (USCI). Cordis catheters have stable tensile strength for five years from manufacture, but then deteriorate substantially by seven years, particularly at the bond between the catheter tip and the stainless steel‐reinforced catheter body. The deterioration is associated with the microscopic appearance of deep cracks in the catheter lumen wall. USCI catheters showed a modest but significant loss of tensile strength by three years after manufacture at the catheter bond. It would seem prudent to place a three‐year expiration date on USCI polyurethane catheters, similar to those already on Cordis catheters.
Title: Fracture of a polyurethane cardiac catheter in the aortic arch: A complication related to polymer aging
Description:
AbstractAn undated Cordis Ducor polyurethane coronary catheter fractured and separated in the aortic arch during a cardiac catheterization, and was removed with an intravascular retriever.
The recovered specimen appeared brittle, prompting an investigation of the effects of aging on the tensile strength of 91 polyurethane catheters manufactured by Cordis and the United States Catheter and Instrument Company (USCI).
Cordis catheters have stable tensile strength for five years from manufacture, but then deteriorate substantially by seven years, particularly at the bond between the catheter tip and the stainless steel‐reinforced catheter body.
The deterioration is associated with the microscopic appearance of deep cracks in the catheter lumen wall.
USCI catheters showed a modest but significant loss of tensile strength by three years after manufacture at the catheter bond.
It would seem prudent to place a three‐year expiration date on USCI polyurethane catheters, similar to those already on Cordis catheters.
Related Results
Selective aortic arch perfusion: a first-in-human observational cadaveric study
Selective aortic arch perfusion: a first-in-human observational cadaveric study
Abstract
Background
Selective aortic arch perfusion (SAAP) is a novel endovascular technique that combines thoracic aortic occlusion with extracorpo...
Selective Aortic Arch Perfusion: A First-in-Man Report of a Novel Endovascular Resuscitation Modality
Selective Aortic Arch Perfusion: A First-in-Man Report of a Novel Endovascular Resuscitation Modality
Abstract
Background
Selective aortic arch perfusion (SAAP) is a novel endovascular technique that combines thoracic aortic occlusion with extracorporeal perfusion of the b...
Fracture Modelling Using Seismic Based Fracture Intensity Volume, a Case Study in Middle East
Fracture Modelling Using Seismic Based Fracture Intensity Volume, a Case Study in Middle East
Abstract
In this paper, a case study in a fractured carbonate reservoir is presented to demonstrate the approach of fracture modeling using fracture intensity vol...
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract
Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
Peripherally Inserted Central Catheters in Children: A Prospective Single-Center Analysis of Associated Complications
Peripherally Inserted Central Catheters in Children: A Prospective Single-Center Analysis of Associated Complications
Highlights
Overall, 72% (n = 63) of PICCs remained in situ until intended therapy was completed.
In 24...
Surgical Management of Norwood Procedure for Atypical Aortic Arch
Surgical Management of Norwood Procedure for Atypical Aortic Arch
Background: Postoperative restenosis of the aortic arch after the Norwood procedure is still an important complication that significantly affects surgical outcomes. The rarity of t...
Cases of Ultrasound-Diagnosed Right Aortic Arch with Right Arterial Duct and the Treatment
Cases of Ultrasound-Diagnosed Right Aortic Arch with Right Arterial Duct and the Treatment
This paper aims to discuss the value of ultrasound to diagnose right aortic arch with right arterial duct. A retrospective analysis of fetal echocardiography characteristics of 10 ...
Successful Aging
Successful Aging
The emerging concept of successful aging is based on evidence that in healthy individual when they get aged, there are considerable variations in physiological functions alteratio...