Technical Note: Film‐based measurement of gold nanoparticle dose enhancement for 192Ir

View through CrossRef
PurposeThe purpose of this work is to introduce a simple yet accurate technique to measure the dose enhancement factor (DEF) of a citrate‐capped gold nanoparticle (GNP) solution using EBT3 film in an 192Ir setup.MethodsDose enhancement factor is the ratio of absorbed dose in a solution compared to absorbed dose in water, assuming identical irradiation parameters. Citrate‐capped GNPs were synthesized. An acrylic apparatus was constructed such that the EBT3 film was placed in charged particle equilibrium within the GNP solution with 0.28%, 0.56%, and 0.77% gold by mass. Sets of 12 dose measurements were collected for each GNP concentration as well as for water. The expected value of DEF was also calculated with the effective mass absorption coefficient of the GNP solution and water for an 192Ir spectrum. Furthermore, Burlin cavity correction factors were calculated and experimentally verified. Experimental verification of the cavity correction was performed by measuring DEF using stacks of 1, 3, and 5 sheets of film and extrapolating the DEF to 0 sheets of film.ResultsExperimental cavity corrections agreed with those calculated with the Burlin cavity formalism. The calculated DEF was 1.013, 1.027, and 1.037 for the 0.28%, 0.56%, and 0.77% gold by mass GNP solutions, respectively. The corresponding uncorrected DEF measurement values were 1.013 ± 0.006, 1.024 ± 0.010, and 1.032 ± 0.006, respectively. When applying the Burlin cavity formalism, the final corrected DEF measurement values were 1.016 ± 0.006, 1.029 ± 0.010, and 1.039 ± 0.006, respectively.ConclusionsThe experimental cavity correction results agreed with the theoretical Burlin calculations, which allowed for the Burlin corrections to be performed for all GNP concentrations and measured DEF values. The adjusted DEF values agreed with the theoretical calculations. Thus, these results indicate that a Burlin cavity calculation can be applied to correct film‐based DEF measurements for 192Ir.
Title: Technical Note: Film‐based measurement of gold nanoparticle dose enhancement for 192Ir
Description:
PurposeThe purpose of this work is to introduce a simple yet accurate technique to measure the dose enhancement factor (DEF) of a citrate‐capped gold nanoparticle (GNP) solution using EBT3 film in an 192Ir setup.
MethodsDose enhancement factor is the ratio of absorbed dose in a solution compared to absorbed dose in water, assuming identical irradiation parameters.
Citrate‐capped GNPs were synthesized.
An acrylic apparatus was constructed such that the EBT3 film was placed in charged particle equilibrium within the GNP solution with 0.
28%, 0.
56%, and 0.
77% gold by mass.
Sets of 12 dose measurements were collected for each GNP concentration as well as for water.
The expected value of DEF was also calculated with the effective mass absorption coefficient of the GNP solution and water for an 192Ir spectrum.
Furthermore, Burlin cavity correction factors were calculated and experimentally verified.
Experimental verification of the cavity correction was performed by measuring DEF using stacks of 1, 3, and 5 sheets of film and extrapolating the DEF to 0 sheets of film.
ResultsExperimental cavity corrections agreed with those calculated with the Burlin cavity formalism.
The calculated DEF was 1.
013, 1.
027, and 1.
037 for the 0.
28%, 0.
56%, and 0.
77% gold by mass GNP solutions, respectively.
The corresponding uncorrected DEF measurement values were 1.
013 ± 0.
006, 1.
024 ± 0.
010, and 1.
032 ± 0.
006, respectively.
When applying the Burlin cavity formalism, the final corrected DEF measurement values were 1.
016 ± 0.
006, 1.
029 ± 0.
010, and 1.
039 ± 0.
006, respectively.
ConclusionsThe experimental cavity correction results agreed with the theoretical Burlin calculations, which allowed for the Burlin corrections to be performed for all GNP concentrations and measured DEF values.
The adjusted DEF values agreed with the theoretical calculations.
Thus, these results indicate that a Burlin cavity calculation can be applied to correct film‐based DEF measurements for 192Ir.

Related Results

Reclaiming the Wasteland: Samson and Delilah and the Historical Perception and Construction of Indigenous Knowledges in Australian Cinema
Reclaiming the Wasteland: Samson and Delilah and the Historical Perception and Construction of Indigenous Knowledges in Australian Cinema
It was always based on a teenage love story between the two kids. One is a sniffer and one is not. It was designed for Central Australia because we do write these kids off there. N...
[RETRACTED] Rhino XL Male Enhancement v1
[RETRACTED] Rhino XL Male Enhancement v1
[RETRACTED]Rhino XL Reviews, NY USA: Studies show that testosterone levels in males decrease constantly with growing age. There are also many other problems that males face due ...
[RETRACTED] Gro-X Male Enhancement | Safely Grow Your Size, Sex Drive v1
[RETRACTED] Gro-X Male Enhancement | Safely Grow Your Size, Sex Drive v1
[RETRACTED]Gro-X Male Enhancement Reviews - Is It Worth the Money? Scam or Legit? Gro-X Male Enhancement Male health is very important, especially for a couple. Low sperm count an...
Alternative Entrances: Phillip Noyce and Sydney’s Counterculture
Alternative Entrances: Phillip Noyce and Sydney’s Counterculture
Phillip Noyce is one of Australia’s most prominent film makers—a successful feature film director with both iconic Australian narratives and many a Hollywood blockbuster under his ...
Ary Scheffer, een Nederlandse Fransman
Ary Scheffer, een Nederlandse Fransman
AbstractAry Scheffer (1795-1858) is so generally included in the French School (Note 2)- unsurprisingly, since his career was confined almost entirely to Paris - that the fact that...
Cinema as a form of composition
Cinema as a form of composition
Technique and creativity Having been called upon to provide a contribution to a publication dedicated to “Techne”, I feel it is fitting to start from the theme of technique, ...
Chemical dosimetry in the near zone of brachytherapy sources
Chemical dosimetry in the near zone of brachytherapy sources
New treatment methods and technical advances in the field of radiation oncology have challenged Medical Physicists to design sophisticated dosimeters to measure the dose in the nea...
[RETRACTED] Omni Male Enhancement Review v1
[RETRACTED] Omni Male Enhancement Review v1
[RETRACTED][video width="852" height="480" mp4="https://lexcliq.com/wp-content/uploads/2022/04/Untitled-video-‐-Made-with-Clipchamp.mp4"][/video] Omni Male Enhancement Reviews Omn...

Back to Top