Differences of blood pressure values between Korotkoff and oscillometric device

View through CrossRef
Abstract Subject Using intro-aortic BP as reference,to compare the accuracy of a Korotkoff sound electronic device (Hanvon FY730) and an oscillometric device (Omron HBP-1320T3) in BP measurement, Methods BP was measured in catheterization coronary angiography. A crossover design was adopted for the using sequence of the two kinds of devices. The average of two BP readings were recorded as the final SBP, DBP and mean arterial pressure (MAP). Non-invasive BP measurement was initiated when the intro-aortic BP curve was stabilized, and the intro-aortic BP curve was simultaneously recorded for 15–20 seconds. BP differences (ΔBP) between non-invasive and its relative intro-aortic BP values were calculated as Oscillometric BP – intro-aortic BP difference (Osc-Ao) or Korotkoff sound method – intro-aortic BP difference (Kor-Ao). Results Using intro-aortic BP as the reference, the Korotkoff sound device provided closer readings compared to the oscillometric device.The Kor-Ao SBP was 19.5 ± 17.6 mmHg, but Osc-Ao SBP was 24.8 ± 20.2 mmHg (P < 0.001). Similarly, the MAP difference was smaller for the Korotkoff device (10.5 ± 10.9 mmHg vs. 13.7 ± 11.3 mmHg, P = 0.007). Although the DBP difference was also smaller for the Korotkoff device (6.1 ± 12.3 mmHg vs. 8.1 ± 11.5 mmHg), but no statistical significance (P = 0.147). Conclusion Using intro-aortic BP reading as reference, the Korotkoff sound device provides more accurate brachial SBP, DBP and MAP values against the oscillometric device.
Title: Differences of blood pressure values between Korotkoff and oscillometric device
Description:
Abstract Subject Using intro-aortic BP as reference,to compare the accuracy of a Korotkoff sound electronic device (Hanvon FY730) and an oscillometric device (Omron HBP-1320T3) in BP measurement, Methods BP was measured in catheterization coronary angiography.
A crossover design was adopted for the using sequence of the two kinds of devices.
The average of two BP readings were recorded as the final SBP, DBP and mean arterial pressure (MAP).
Non-invasive BP measurement was initiated when the intro-aortic BP curve was stabilized, and the intro-aortic BP curve was simultaneously recorded for 15–20 seconds.
BP differences (ΔBP) between non-invasive and its relative intro-aortic BP values were calculated as Oscillometric BP – intro-aortic BP difference (Osc-Ao) or Korotkoff sound method – intro-aortic BP difference (Kor-Ao).
Results Using intro-aortic BP as the reference, the Korotkoff sound device provided closer readings compared to the oscillometric device.
The Kor-Ao SBP was 19.
5 ± 17.
6 mmHg, but Osc-Ao SBP was 24.
8 ± 20.
2 mmHg (P < 0.
001).
Similarly, the MAP difference was smaller for the Korotkoff device (10.
5 ± 10.
9 mmHg vs.
13.
7 ± 11.
3 mmHg, P = 0.
007).
Although the DBP difference was also smaller for the Korotkoff device (6.
1 ± 12.
3 mmHg vs.
8.
1 ± 11.
5 mmHg), but no statistical significance (P = 0.
147).
Conclusion Using intro-aortic BP reading as reference, the Korotkoff sound device provides more accurate brachial SBP, DBP and MAP values against the oscillometric device.

Related Results

[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Influence of short-term blood pressure variability on blood pressure determinations.
Influence of short-term blood pressure variability on blood pressure determinations.
To evaluate the effect of blood pressure variability on Riva Rocci Korotkoff blood pressure determinations, we studied the intra-arterial pressure during Riva Rocci Korotkoff deter...
Online Micro–Controller Non–Invasive Blood Pressure Monitoring System (E–BPM)
Online Micro–Controller Non–Invasive Blood Pressure Monitoring System (E–BPM)
Pengukuran tekanan darah telahpun merupakan sebahagian daripada pemeriksaan klinikal pada zaman perubatan moden ini. Dua daripada kaedah yang sering diaplikasi dalam mengukur tekan...
Measuring Blood Pressure
Measuring Blood Pressure
Abstract Blood pressure measurements are one part of a circulatory assessment (Docherty and McCallum 2009). Treatments for raised or low blood pressure may be initia...
Reconstruction of Brachial Artery Pressure From Noninvasive Finger Pressure Measurements
Reconstruction of Brachial Artery Pressure From Noninvasive Finger Pressure Measurements
Background Pulse wave distortions, mainly caused by reflections, and pressure gradients, caused by flow in the resistive vascular tree, may cause differences between fi...
Ambulatory blood pressure of adults in Ohasama, Japan.
Ambulatory blood pressure of adults in Ohasama, Japan.
We performed a cross-sectional study in a small town in northern Japan to evaluate the distribution, reference values, and daily variation in ambulatory blood pressure. A total of ...
Determinants of Cerebrovascular Reserve in Patients with Significant Carotid Stenosis
Determinants of Cerebrovascular Reserve in Patients with Significant Carotid Stenosis
Abstract Introduction In patients with 70% to 99% diameter carotid artery stenosis cerebral blood flow reserve may be protectiv...

Back to Top