Differences of blood pressure values between Korotkoff and oscillometric device

View through CrossRef
AbstractSubjectUsing intro-aortic BP as reference,to compare the accuracy of a Korotkoff sound electronic device (Hanvon FY730) and an oscillometric device (Omron HBP-1320T3) in BP measurement,MethodsBP was measured in catheterization coronary angiography. A crossover design was adopted for the using sequence of the two kinds of devices. The average of two BP readings were recorded as the final SBP, DBP and mean arterial pressure (MAP). Non-invasive BP measurement was initiated when the intro-aortic BP curve was stabilized, and the intro-aortic BP curve was simultaneously recorded for 15–20 seconds. BP differences (ΔBP) between non-invasive and its relative intro-aortic BP values were calculated as Oscillometric BP – intro-aortic BP difference (Osc-Ao) or Korotkoff sound method – intro-aortic BP difference (Kor-Ao).ResultsUsing intro-aortic BP as the reference, the Korotkoff sound device provided closer readings compared to the oscillometric device.The Kor-Ao SBP was 19.5 ± 17.6 mmHg, but Osc-Ao SBP was 24.8 ± 20.2 mmHg (P < 0.001). Similarly, the MAP difference was smaller for the Korotkoff device (10.5 ± 10.9 mmHg vs. 13.7 ± 11.3 mmHg, P = 0.007). Although the DBP difference was also smaller for the Korotkoff device (6.1 ± 12.3 mmHg vs. 8.1 ± 11.5 mmHg), but no statistical significance (P = 0.147).ConclusionUsing intro-aortic BP reading as reference, the Korotkoff sound device provides more accurate brachial SBP, DBP and MAP values against the oscillometric device.
Cold Spring Harbor Laboratory
Title: Differences of blood pressure values between Korotkoff and oscillometric device
Description:
AbstractSubjectUsing intro-aortic BP as reference,to compare the accuracy of a Korotkoff sound electronic device (Hanvon FY730) and an oscillometric device (Omron HBP-1320T3) in BP measurement,MethodsBP was measured in catheterization coronary angiography.
A crossover design was adopted for the using sequence of the two kinds of devices.
The average of two BP readings were recorded as the final SBP, DBP and mean arterial pressure (MAP).
Non-invasive BP measurement was initiated when the intro-aortic BP curve was stabilized, and the intro-aortic BP curve was simultaneously recorded for 15–20 seconds.
BP differences (ΔBP) between non-invasive and its relative intro-aortic BP values were calculated as Oscillometric BP – intro-aortic BP difference (Osc-Ao) or Korotkoff sound method – intro-aortic BP difference (Kor-Ao).
ResultsUsing intro-aortic BP as the reference, the Korotkoff sound device provided closer readings compared to the oscillometric device.
The Kor-Ao SBP was 19.
5 ± 17.
6 mmHg, but Osc-Ao SBP was 24.
8 ± 20.
2 mmHg (P < 0.
001).
Similarly, the MAP difference was smaller for the Korotkoff device (10.
5 ± 10.
9 mmHg vs.
13.
7 ± 11.
3 mmHg, P = 0.
007).
Although the DBP difference was also smaller for the Korotkoff device (6.
1 ± 12.
3 mmHg vs.
8.
1 ± 11.
5 mmHg), but no statistical significance (P = 0.
147).
ConclusionUsing intro-aortic BP reading as reference, the Korotkoff sound device provides more accurate brachial SBP, DBP and MAP values against the oscillometric device.

Related Results

[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Online Micro–Controller Non–Invasive Blood Pressure Monitoring System (E–BPM)
Online Micro–Controller Non–Invasive Blood Pressure Monitoring System (E–BPM)
Pengukuran tekanan darah telahpun merupakan sebahagian daripada pemeriksaan klinikal pada zaman perubatan moden ini. Dua daripada kaedah yang sering diaplikasi dalam mengukur tekan...
Algoritma Interpolasi Nearest-Neighbor untuk Pendeteksian Sampul Pulsa Oscilometri Menggunakan Mikrokontroler Berbiaya Rendah
Algoritma Interpolasi Nearest-Neighbor untuk Pendeteksian Sampul Pulsa Oscilometri Menggunakan Mikrokontroler Berbiaya Rendah
Non-invasive blood pressure measurement devices are widely available in the marketplace. Most of these devices use the oscillometric principle that store and analyze oscillometric ...
Determinants of Cerebrovascular Reserve in Patients with Significant Carotid Stenosis
Determinants of Cerebrovascular Reserve in Patients with Significant Carotid Stenosis
Abstract Introduction In patients with 70% to 99% diameter carotid artery stenosis cerebral blood flow reserve may be protectiv...
Blood Cross Matching Without Anti-Human Globulin (AHG) and Bovine Serum: A New Interest for an Old Idea
Blood Cross Matching Without Anti-Human Globulin (AHG) and Bovine Serum: A New Interest for an Old Idea
Abstract  Introduction Transfusion medicine promotes the safety of blood transfusions by rigorously testing to eliminate risks of infection and hemolytic. The efficacy (to correct ...
Extensions of Pressure Build-Up Analysis Methods
Extensions of Pressure Build-Up Analysis Methods
RUSSELL, D.G.,* MEMBER AIME, SHELL DEVELOPMENT CO., HOUSTON, TEX. Abstract Two techniques have been developed with which the app...

Back to Top